A new non-invasive continuous cardiac output trend solely utilizing routine cardiovascular monitors.

Hironori Ishihara1, Hirobumi Okawa, Ken Tanabe, Toshihito Tsubo, Yoshihiro Sugo, Takeshi Akiyama, Sunao Takeda.   

Abstract

OBJECTIVE: Three of the us developed a new non-invasive continuous cardiac output (CCO) measurement method utilizing routine clinical monitors based on the pulse-contour analysis combined with pulse wave transit time (PWTT). Using pulmonary artery catheter (CCOpa), we compared this estimated CCO (esCO) with the thermodilution CCO early after cardiac surgery, and tested whether the esCO method has potential of being an alternative measure of CCO.
METHODS: Thirty-six patients without continued arrhythmias were studied. esCO was computed using electrocardiogram (ECG) monitor, arterial pressure monitor and pulse-oximetry system. Both sets of data (esCO and CCOpa), by averaging the results of the preceding 10 min, were compared at 30-min intervals throughout the 15.8 +/- 3.3 h (S.D.) of study. Bland-Altman plots and correlation analysis were used for statistical comparison.
RESULTS: A total of 981 paired sets of data (89.9%) among 1093 measurements were compared in the absence of displacement of either pulse-oximetry or ECG probes and/or inaccurate detection of R wave. The difference between esCO and CCOpa results was -0.06 +/- 0.82 L/min (S.D.), and there was a linear correlation between them (r = 0.80, p < 0.0001). The difference between them was 0.00 +/- 0.48 L/min at the first 1 h, which remained unchanged throughout 20 h after the start of measurement.
CONCLUSIONS: The results demonstrate that esCO has a close correlation with the CCOpa, even though the two methods are not interchangeable. The results suggest that esCO method has potential of being an alternative non-invasive cardiac output trend, unless there are apparent arrhythmias.

Entities:  

Mesh:

Year:  2004        PMID: 15957621     DOI: 10.1007/s10877-005-2452-5

Source DB:  PubMed          Journal:  J Clin Monit Comput        ISSN: 1387-1307            Impact factor:   2.502


  15 in total

1.  Continuous and intermittent cardiac output measurement: pulmonary artery catheter versus aortic transpulmonary technique.

Authors:  G Della Rocca; M G Costa; L Pompei; C Coccia; P Pietropaoli
Journal:  Br J Anaesth       Date:  2002-03       Impact factor: 9.166

2.  The relationship between modified pulse wave transit time and cardiovascular changes in isoflurane anesthetized dogs.

Authors:  R Ochiai; J Takeda; H Hosaka; Y Sugo; R Tanaka; T Soma
Journal:  J Clin Monit Comput       Date:  1999-12       Impact factor: 2.502

3.  Continuous cardiac output catheters: delay in in vitro response time after controlled flow changes.

Authors:  M Aranda; F G Mihm; S Garrett; M N Mihm; R G Pearl
Journal:  Anesthesiology       Date:  1998-12       Impact factor: 7.892

4.  Delayed time response of the continuous cardiac output pulmonary artery catheter.

Authors:  L C Siegel; M M Hennessy; R G Pearl
Journal:  Anesth Analg       Date:  1996-12       Impact factor: 5.108

5.  Reproducibility of thermodilution cardiac output determination in critically ill patients: comparison between bolus and continuous method.

Authors:  Y Le Tulzo; M Belghith; P Seguin; J Dall'Ava; M Monchi; R Thomas; J F Dhainaut
Journal:  J Clin Monit       Date:  1996-09

6.  Evaluation of the accuracy and response time of STAT-mode continuous cardiac output.

Authors:  M A Lazor; E T Pierce; G D Stanley; J L Cass; E F Halpern; R H Bode
Journal:  J Cardiothorac Vasc Anesth       Date:  1997-06       Impact factor: 2.628

7.  Continuous cardiac output measurements do not agree with conventional bolus thermodilution cardiac output determination.

Authors:  C Zöllner; A E Goetz; M Weis; K Mörstedt; B Pichler; P Lamm; E Kilger; M Haller
Journal:  Can J Anaesth       Date:  2001-12       Impact factor: 5.063

8.  Comparison of pulmonary arterial thermodilution and arterial pulse contour analysis: evaluation of a new algorithm.

Authors:  Thomas W Felbinger; Daniel A Reuter; Holger K Eltzschig; Karl Moerstedt; Oliver Goedje; Alwin E Goetz
Journal:  J Clin Anesth       Date:  2002-06       Impact factor: 9.452

9.  [Continuous measurement of cardiac output with pulse contour analysis].

Authors:  M Irlbeck; H Forst; J Briegel; M Haller; K Peter
Journal:  Anaesthesist       Date:  1995-07       Impact factor: 1.041

10.  Effect of the degree of tricuspid regurgitation on cardiac output measurements by thermodilution.

Authors:  Martin Balik; Jan Pachl; Jan Hendl; Balik Martin; Pachl Jan; Hendl Jan
Journal:  Intensive Care Med       Date:  2002-06-15       Impact factor: 17.440

View more
  9 in total

1.  Comparison of the ability of two continuous cardiac output monitors to measure trends in cardiac output: estimated continuous cardiac output measured by modified pulse wave transit time and an arterial pulse contour-based cardiac output device.

Authors:  Takashi Terada; Ayano Oiwa; Yumi Maemura; Samuna Robert; Sayaka Kessoku; Ryoichi Ochiai
Journal:  J Clin Monit Comput       Date:  2015-09-14       Impact factor: 2.502

Review 2.  Year in review in journal of clinical monitoring and computing 2014: cardiovascular and hemodynamic monitoring.

Authors:  Karim Bendjelid; Steffen Rex; Thomas Scheeren; Lester Critchley
Journal:  J Clin Monit Comput       Date:  2015-03-05       Impact factor: 2.502

3.  Evaluation of the estimated continuous cardiac output monitoring system in adults and children undergoing kidney transplant surgery: a pilot study.

Authors:  Takashi Terada; Yumi Maemura; Akiko Yoshida; Rika Muto; Ryoichi Ochiai
Journal:  J Clin Monit Comput       Date:  2013-08-21       Impact factor: 2.502

4.  The ability of a new continuous cardiac output monitor to measure trends in cardiac output following implementation of a patient information calibration and an automated exclusion algorithm.

Authors:  Hironori Ishihara; Yoshihiro Sugo; Masato Tsutsui; Takashige Yamada; Tetsufumi Sato; Toshimasa Akazawa; Nobukazu Sato; Koichi Yamashita; Junzo Takeda
Journal:  J Clin Monit Comput       Date:  2012-08-02       Impact factor: 2.502

5.  Agreement between continuous and intermittent pulmonary artery thermodilution for cardiac output measurement in perioperative and intensive care medicine: a systematic review and meta-analysis.

Authors:  Karim Kouz; Frederic Michard; Alina Bergholz; Christina Vokuhl; Luisa Briesenick; Phillip Hoppe; Moritz Flick; Gerhard Schön; Bernd Saugel
Journal:  Crit Care       Date:  2021-03-29       Impact factor: 9.097

6.  Estimation of cardiac output and peripheral resistance using square-wave-approximated aortic flow signal.

Authors:  Nima Fazeli; Jin-Oh Hahn
Journal:  Front Physiol       Date:  2012-07-25       Impact factor: 4.566

7.  Pulse Wave Transit Time Measurements of Cardiac Output in Septic Shock Patients: A Comparison of the Estimated Continuous Cardiac Output System with Transthoracic Echocardiography.

Authors:  Marc Feissel; Ludwig Serge Aho; Stefan Georgiev; Romain Tapponnier; Julio Badie; Rémi Bruyère; Jean-Pierre Quenot
Journal:  PLoS One       Date:  2015-06-30       Impact factor: 3.240

8.  Pulse-wave transit time with ventilator-induced variation for the prediction of fluid responsiveness.

Authors:  Koichi Yamashita
Journal:  Acute Med Surg       Date:  2020-01-27

9.  Relationship between stroke volume and pulse pressure during blood volume perturbation: a mathematical analysis.

Authors:  Ramin Bighamian; Jin-Oh Hahn
Journal:  Biomed Res Int       Date:  2014-05-20       Impact factor: 3.411

  9 in total