BACKGROUND AND
PURPOSE: The purpose of this study was to examine the efficacy and safety of the 6F Closer S device (Perclose; Redwood City, CA) versus manual compression to close arteriotomy sites after neurointerventional procedures in both virgin vessels and those previously treated with the device.
METHODS: This single-center, multiple-operator, controlled, prospective study included 475 procedures (337 patients) with the device and 79 procedures (79 patients) with manual compression. A substantial number of patients receiving anticoagulation and/or antiplatelet medications were included. Efficacy and safety were evaluated on the basis of the rate of hemostatic success and the incidence of clinically significant complications. The literature was reviewed by using MEDLINE.
RESULTS: Overall success rates were 95% in the device group versus 96% in the manual-compression group (P = 0.78), and clinically relevant complication rates were 0.6% versus 2.5%, respectively (P = 0.15). Success rates significantly declined in vessels previously treated with the device three or more times. However, complication rates did not significantly change. Literature review yielded 12 articles reporting complication rates of 3.2-35% for the device and 2.3%-33.3% for manual compression.
CONCLUSION: The device was safe and effective for closing arteriotomy sites in patients undergoing neurointerventional procedures, including those receiving anticoagulation/antiplatelet therapy or those previously treated with the device one or two times.

Entities:  

Mesh:

Year:  2005        PMID: 15956509      PMCID: PMC8149059     

Source DB:  PubMed          Journal:  AJNR Am J Neuroradiol        ISSN: 0195-6108            Impact factor:   3.825


  19 in total

1.  Management of arterial puncture site after catheterization procedures: evaluating a suture-mediated closure device.

Authors:  U Gerckens; N Cattelaens; E G Lampe; E Grube
Journal:  Am J Cardiol       Date:  1999-06-15       Impact factor: 2.778

2.  Suture-mediated closure of the femoral access site after cardiac catheterization: results of the suture to ambulate aNd discharge (STAND I and STAND II) trials.

Authors:  D S Baim; W D Knopf; T Hinohara; D E Schwarten; R A Schatz; C A Pinkerton; D E Cutlip; M Fitzpatrick; K K Ho; R E Kuntz
Journal:  Am J Cardiol       Date:  2000-04-01       Impact factor: 2.778

3.  A randomised controlled trial of Prostar Plus for haemostasis in patients after coronary angioplasty.

Authors:  T Noguchi; S Miyazaki; S Yasuda; T Baba; H Sumida; I Morii; S Daikoku; Y Goto; H Nonogi
Journal:  Eur J Vasc Endovasc Surg       Date:  2000-05       Impact factor: 7.069

4.  Platelet glycoprotein IIb/IIIa receptor blockade and low-dose heparin during percutaneous coronary revascularization.

Authors: 
Journal:  N Engl J Med       Date:  1997-06-12       Impact factor: 91.245

5.  Safety and efficacy of suture-mediated closure after percutaneous coronary interventions.

Authors:  M R Rinder; P K Tamirisa; M Taniuchi; H I Kurz; K Mumm; J M Lasala
Journal:  Catheter Cardiovasc Interv       Date:  2001-10       Impact factor: 2.692

6.  Predictors of groin complications after balloon and new-device coronary intervention.

Authors:  R Waksman; S B King; J S Douglas; Y Shen; H Ewing; L Mueller; Z M Ghazzal; W S Weintraub
Journal:  Am J Cardiol       Date:  1995-05-01       Impact factor: 2.778

7.  Safety and efficacy of the Perclose suture-mediated closure device after diagnostic and interventional catheterizations in a large consecutive population.

Authors:  Zalman M Kahn; Mukesh Kumar; Gerald Hollander; Robert Frankel
Journal:  Catheter Cardiovasc Interv       Date:  2002-01       Impact factor: 2.692

8.  A new device for percutaneous suture-mediated closure of arterial puncture sites using exteriorized needles: initial experience.

Authors:  U Hahn; A Betsch; J Wiskirchen; C König; G Tepe; A Baumbach; S H Duda
Journal:  J Invasive Cardiol       Date:  2001-06       Impact factor: 2.022

Review 9.  Arterial puncture closing devices compared with standard manual compression after cardiac catheterization: systematic review and meta-analysis.

Authors:  Maria Koreny; Eva Riedmüller; Mariam Nikfardjam; Peter Siostrzonek; Marcus Müllner
Journal:  JAMA       Date:  2004-01-21       Impact factor: 56.272

10.  Safety and efficacy of the Perclose suture-mediated closure device following carotid artery stenting under clopidogrel platelet blockade.

Authors:  Niels Zorger; Thomas Finkenzeller; Markus Lenhart; Okka Hamer; Christian Paetzel; Inghita Borisch; Ingolf Toepel; Stefan Feuerbach; Johann Link
Journal:  Eur Radiol       Date:  2003-11-11       Impact factor: 5.315

View more
  3 in total

1.  StarClose vascular closure device: safety and efficacy of deployment and reaccess in a neurointerventional radiology service.

Authors:  R A McTaggart; D Raghavan; R A Haas; M V Jayaraman
Journal:  AJNR Am J Neuroradiol       Date:  2010-01-21       Impact factor: 3.825

2.  A retrospective review on feasibility and safety of a new pneumatic compression device for femoral arteriotomy hemostasis.

Authors:  Minho Kang; Sung Wook Shin
Journal:  Korean J Radiol       Date:  2011-12-23       Impact factor: 3.500

3.  Prospective comparison of angio-seal versus manual compression for hemostasis after neurointerventional procedures under systemic heparinization.

Authors:  H-F Wong; C-W Lee; Y-L Chen; Y-M Wu; H-H Weng; Y-H Wang; H-M Liu
Journal:  AJNR Am J Neuroradiol       Date:  2012-08-02       Impact factor: 3.825

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.