Custom measurement of retinal nerve fiber layer thickness using STRATUS OCT in normal eyes.

P Carpineto1, M Ciancaglini, A Aharrh-Gnama, D Cirone, L Mastropasqua.   

Abstract

PURPOSE: To evaluate variability of retinal nerve fiber layer (RNFL) thickness measurements in normal eyes and their correlation with optic disc diameter by using two different scan options of the ultimate commercial optical coherence tomography (OCT) unit (STRATUS OCT, Carl Zeiss Meditec, Inc., Dublin, CA).
METHODS: In this observational case series and instrument validation study 30 eyes of 30 normal subjects were enrolled. Each eye underwent optic disc vertical diameter measurement by means of both stereoscopic photography and planimetry and OCT; RNFL thickness measurements were performed using OCT. Three repetitions of two series of scans were performed. Each eye was scanned at two different options (RNFL thickness 3.4 and Nerve Head Circle). For each option descriptive statistics, analysis of variance, intraclass correlation coefficients (ICCs), and coefficients of variation (COVs) were calculated. To verify the correlation between the two methods of optic disc diameter assessment and to study the influence of optic disc diameter on RNFL measurement using the two different OCT options, Pearson's correlation coefficients were calculated.
RESULTS: Optic disc diameter length ranged from 1.47 to 2.04 mm (mean 1.709 mm, SD +/- 0.147) with stereoscopic photographs, and from 1.47 to 2.02 mm (mean 1.703 mm, SD +/- 0.143) with OCT (Pearson correlation coefficient 0.999, p<0.001). Mean RNFL thickness was 89.29 mm (SD +/- 10.80 mm) using the RNFL thickness 3.4 scanning option and 89.88 mm (SD +/- 1.72 mm) using the Nerve Head Circle protocol (Pearson correlation coefficient 0.065, p=0.734). The intersubject variance is higher using the RNFL thickness 3.4 option than using the NHC protocol (sum of square: 1,014,760 vs. 25,741) (p<0.001); the intrasubject variance is very similar in the two groups (2,372 vs 2,360) (p=NS). The ICC is 99.89% when using the RNFL thickness 3.4 option, 95.62% with the NHC protocol (p=NS). COVs were 12.10% and 1.91% by using RNFL thickness 3.4 and Nerve Head Circle option, respectively. Pearson's correlation coefficient was 0.988 (p<0.001) when comparing optic disc diameter and RNFL thickness by using the RNFL thickness 3.4 option and -0.016 (p=0.932) when comparing optic disc diameter and RNFL thickness by using the Nerve Head Circle option.
CONCLUSIONS: These results suggest that both scan options give good RNFL thickness measurement reproducibility; the use of the Nerve Head Circle option leads to less interindividual variability and can minimize the effect of differences in optic disc diameter on RNFL thickness measurements in normal subjects.

Mesh:

Year:  2005        PMID: 15945005     DOI: 10.1177/112067210501500308

Source DB:  PubMed          Journal:  Eur J Ophthalmol        ISSN: 1120-6721            Impact factor:   2.597


  12 in total

1.  Correlation between optic disc area and retinal nerve fiber layer thickness: a study on scanning laser polarimetry with variable corneal compensation.

Authors:  Stefano Da Pozzo; Pierluigi Iacono; Luca Michelone; Marco Paoloni; Giuseppe Ravalico
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2006-11-17       Impact factor: 3.117

2.  Sources of longitudinal variability in optical coherence tomography nerve-fibre layer measurements.

Authors:  L Kagemann; T Mumcuoglu; G Wollstein; R Bilonick; H Ishikawa; K A Townsend; M Gabriele; J G Fujimoto; J S Schuman
Journal:  Br J Ophthalmol       Date:  2008-06       Impact factor: 4.638

3.  Symmetry between the right and left eyes of the normal retinal nerve fiber layer measured with optical coherence tomography (an AOS thesis).

Authors:  Donald L Budenz
Journal:  Trans Am Ophthalmol Soc       Date:  2008

4.  Determinants of normal retinal nerve fiber layer thickness measured by Stratus OCT.

Authors:  Donald L Budenz; Douglas R Anderson; Rohit Varma; Joel Schuman; Louis Cantor; Jonathan Savell; David S Greenfield; Vincent Michael Patella; Harry A Quigley; James Tielsch
Journal:  Ophthalmology       Date:  2007-01-08       Impact factor: 12.079

5.  Does optic nerve head size variation affect circumpapillary retinal nerve fiber layer thickness measurement by optical coherence tomography?

Authors:  David Huang; Vikas Chopra; Ake Tzu-Hui Lu; Ou Tan; Brian Francis; Rohit Varma
Journal:  Invest Ophthalmol Vis Sci       Date:  2012-07-27       Impact factor: 4.799

6.  Scan tracking coordinates for improved centering of Stratus OCT scan pattern.

Authors:  Gianmarco Vizzeri; Christopher Bowd; Felipe A Medeiros; Robert N Weinreb; Linda M Zangwill
Journal:  J Glaucoma       Date:  2009-01       Impact factor: 2.503

7.  Effect of improper scan alignment on retinal nerve fiber layer thickness measurements using Stratus optical coherence tomograph.

Authors:  Gianmarco Vizzeri; Christopher Bowd; Felipe A Medeiros; Robert N Weinreb; Linda M Zangwill
Journal:  J Glaucoma       Date:  2008-08       Impact factor: 2.503

8.  Effect of signal strength and improper alignment on the variability of stratus optical coherence tomography retinal nerve fiber layer thickness measurements.

Authors:  Gianmarco Vizzeri; Christopher Bowd; Felipe A Medeiros; Robert N Weinreb; Linda M Zangwill
Journal:  Am J Ophthalmol       Date:  2009-05-09       Impact factor: 5.258

9.  The effect of various factors on variability of retinal nerve fiber layer thickness measurements using optical coherence tomography.

Authors:  Dong Ju Youm; Hyunjoong Kim; Seong Hee Shim; Hyo Ju Jang; Joon Mo Kim; Ki Ho Park; Chul Young Choi; Jung Gon Cho
Journal:  Korean J Ophthalmol       Date:  2012-03-22

10.  Size of the Optic Nerve Head and Its Relationship with the Thickness of the Macular Ganglion Cell Complex and Peripapillary Retinal Nerve Fiber Layer in Patients with Primary Open Angle Glaucoma.

Authors:  Nobuko Enomoto; Ayako Anraku; Kyoko Ishida; Asuka Takeyama; Fumihiko Yagi; Goji Tomita
Journal:  J Ophthalmol       Date:  2015-08-03       Impact factor: 1.909

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.