Literature DB >> 18703942

Effect of improper scan alignment on retinal nerve fiber layer thickness measurements using Stratus optical coherence tomograph.

Gianmarco Vizzeri1, Christopher Bowd, Felipe A Medeiros, Robert N Weinreb, Linda M Zangwill.   

Abstract

OBJECTIVE: Misalignment of the Stratus optical coherence tomograph scan circle placed by the operator around the optic nerve head (ONH) during each retinal nerve fiber layer (RNFL) examination can affect the instrument reproducibility and its theoretical ability to detect true structural changes in the RNFL thickness over time. We evaluated the effect of scan circle placement on RNFL measurements.
DESIGN: Observational clinical study.
METHODS: Sixteen eyes of 8 normal participants were examined using the Stratus optical coherence tomograph Fast RNFL thickness acquisition protocol (software version 4.0.7; Carl Zeiss Meditec, Dublin, CA). Four consecutive images were taken by the same operator with the circular scan centered on the optic nerve head. Four images each with the scan displaced superiorly, inferiorly, temporally, and nasally were also acquired. Differences in average and sectoral RNFL thicknesses were determined. For the centered scans, the coefficients of variation (CV) and the intraclass correlation coefficient for the average RNFL thickness measured were calculated.
RESULTS: When the average RNFL thickness of the centered scans was compared with the average RNFL thickness of the displaced scans individually using analysis of variance with post-hoc analysis, no difference was found between the average RNFL thickness of the nasally (105.2 microm), superiorly (106.2 microm), or inferiorly (104.1 microm) displaced scans and the centered scans (106.4 microm). However, a significant difference (analysis of variance with Dunnett's test: F=8.82, P<0.0001) was found between temporally displaced scans (115.8 microm) and centered scans. Significant differences in sectoral RNFL thickness measurements were found between centered and each displaced scan. The coefficient of variation for average RNFL thickness was 1.75% and intraclass correlation coefficient was 0.95.
CONCLUSIONS: In normal eyes, average RNFL thickness measurements are robust and similar with significant superior, inferior, and nasal scan displacement, but average RNFL thickness is greater when scans are displaced temporally. Parapapillary scan misalignment produces significant changes in RNFL assessment characterized by an increase in measured RNFL thickness in the quadrant in which the scan is closer to the disc, and a significant decrease in RNFL thickness in the quadrant in which the scan is displaced further from the optic disc.

Entities:  

Mesh:

Year:  2008        PMID: 18703942      PMCID: PMC3465954          DOI: 10.1097/IJG.0b013e31815c3aeb

Source DB:  PubMed          Journal:  J Glaucoma        ISSN: 1057-0829            Impact factor:   2.503


  21 in total

1.  Thickness of the peripapillary retina in healthy subjects with different degrees of ametropia.

Authors:  E Garcia-Valenzuela; M Mori; D P Edward; M Shahidi
Journal:  Ophthalmology       Date:  2000-07       Impact factor: 12.079

2.  Diagnostic capability of optical coherence tomography in evaluating the degree of glaucomatous retinal nerve fiber damage.

Authors:  Ramanjit Sihota; Parul Sony; Viney Gupta; Tanuj Dada; Rajvir Singh
Journal:  Invest Ophthalmol Vis Sci       Date:  2006-05       Impact factor: 4.799

3.  Reproducibility of nerve fiber thickness, macular thickness, and optic nerve head measurements using StratusOCT.

Authors:  Lelia A Paunescu; Joel S Schuman; Lori Lyn Price; Paul C Stark; Siobahn Beaton; Hiroshi Ishikawa; Gadi Wollstein; James G Fujimoto
Journal:  Invest Ophthalmol Vis Sci       Date:  2004-06       Impact factor: 4.799

4.  Combining structural and functional testing for detection of glaucoma.

Authors:  Neha N Shah; Christopher Bowd; Felipe A Medeiros; Robert N Weinreb; Pamela A Sample; Esther M Hoffmann; Linda M Zangwill
Journal:  Ophthalmology       Date:  2006-09       Impact factor: 12.079

5.  Influence of pupil size and cataract on retinal nerve fiber layer thickness measurements by Stratus OCT.

Authors:  Giacomo Savini; Maurizio Zanini; Piero Barboni
Journal:  J Glaucoma       Date:  2006-08       Impact factor: 2.503

6.  Structure-function relationships using confocal scanning laser ophthalmoscopy, optical coherence tomography, and scanning laser polarimetry.

Authors:  Christopher Bowd; Linda M Zangwill; Felipe A Medeiros; Ivan M Tavares; Esther M Hoffmann; Rupert R Bourne; Pamela A Sample; Robert N Weinreb
Journal:  Invest Ophthalmol Vis Sci       Date:  2006-07       Impact factor: 4.799

7.  Optical coherence tomography for clinical detection and monitoring of glaucoma?

Authors:  Christopher Bowd
Journal:  Br J Ophthalmol       Date:  2007-07       Impact factor: 4.638

8.  Comparison between GDx VCC scanning laser polarimetry and Stratus OCT optical coherence tomography in the diagnosis of chronic glaucoma.

Authors:  Paolo Brusini; Maria L Salvetat; Marco Zeppieri; Claudia Tosoni; Lucia Parisi; Mirella Felletti
Journal:  Acta Ophthalmol Scand       Date:  2006-10

9.  Quantitative assessment of structural damage in eyes with localized visual field abnormalities.

Authors:  Harmohina Bagga; David S Greenfield
Journal:  Am J Ophthalmol       Date:  2004-05       Impact factor: 5.258

Review 10.  Retinal nerve fiber layer analysis in the diagnosis of glaucoma.

Authors:  Linda M Zangwill; Christopher Bowd
Journal:  Curr Opin Ophthalmol       Date:  2006-04       Impact factor: 3.761

View more
  29 in total

1.  Effect of image quality on tissue thickness measurements obtained with spectral domain-optical coherence tomography.

Authors:  Madhusudhanan Balasubramanian; Christopher Bowd; Gianmarco Vizzeri; Robert N Weinreb; Linda M Zangwill
Journal:  Opt Express       Date:  2009-03-02       Impact factor: 3.894

2.  Influence of anterior segment power on the scan path and RNFL thickness using SD-OCT.

Authors:  Nimesh B Patel; Brenda Garcia; Ronald S Harwerth
Journal:  Invest Ophthalmol Vis Sci       Date:  2012-08-24       Impact factor: 4.799

3.  Diagnostic Capability of Peripapillary Three-dimensional Retinal Nerve Fiber Layer Volume for Glaucoma Using Optical Coherence Tomography Volume Scans.

Authors:  Ziad Khoueir; Firas Jassim; Linda Yi-Chieh Poon; Edem Tsikata; Geulah S Ben-David; Yingna Liu; Eric Shieh; Ramon Lee; Rong Guo; Georgia Papadogeorgou; Boy Braaf; Huseyin Simavli; Christian Que; Benjamin J Vakoc; Brett E Bouma; Johannes F de Boer; Teresa C Chen
Journal:  Am J Ophthalmol       Date:  2017-08-12       Impact factor: 5.258

4.  Agreement between retinal nerve fiber layer measures from Spectralis and Cirrus spectral domain OCT.

Authors:  Nimesh B Patel; Joe L Wheat; Aldon Rodriguez; Victoria Tran; Ronald S Harwerth
Journal:  Optom Vis Sci       Date:  2012-05       Impact factor: 1.973

Review 5.  Three dimensional optical coherence tomography imaging: advantages and advances.

Authors:  Michelle L Gabriele; Gadi Wollstein; Hiroshi Ishikawa; Juan Xu; Jongsick Kim; Larry Kagemann; Lindsey S Folio; Joel S Schuman
Journal:  Prog Retin Eye Res       Date:  2010-06-11       Impact factor: 21.198

6.  Detection of progressive retinal nerve fiber layer thickness loss with optical coherence tomography using 4 criteria for functional progression.

Authors:  Dilraj S Grewal; Mitra Sehi; James D Paauw; David S Greenfield
Journal:  J Glaucoma       Date:  2012 Apr-May       Impact factor: 2.503

7.  Patient characteristics associated with artifacts in Spectralis optical coherence tomography imaging of the retinal nerve fiber layer in glaucoma.

Authors:  Yingna Liu; Huseyin Simavli; Christian John Que; Jennifer L Rizzo; Edem Tsikata; Rie Maurer; Teresa C Chen
Journal:  Am J Ophthalmol       Date:  2014-12-12       Impact factor: 5.258

8.  Scan tracking coordinates for improved centering of Stratus OCT scan pattern.

Authors:  Gianmarco Vizzeri; Christopher Bowd; Felipe A Medeiros; Robert N Weinreb; Linda M Zangwill
Journal:  J Glaucoma       Date:  2009-01       Impact factor: 2.503

9.  Age-associated changes in the retinal nerve fiber layer and optic nerve head.

Authors:  Nimesh B Patel; Mimi Lim; Avni Gajjar; Kelsey B Evans; Ronald S Harwerth
Journal:  Invest Ophthalmol Vis Sci       Date:  2014-07-22       Impact factor: 4.799

10.  Effect of signal strength and improper alignment on the variability of stratus optical coherence tomography retinal nerve fiber layer thickness measurements.

Authors:  Gianmarco Vizzeri; Christopher Bowd; Felipe A Medeiros; Robert N Weinreb; Linda M Zangwill
Journal:  Am J Ophthalmol       Date:  2009-05-09       Impact factor: 5.258

View more