Scan tracking coordinates for improved centering of Stratus OCT scan pattern.

Gianmarco Vizzeri1, Christopher Bowd, Felipe A Medeiros, Robert N Weinreb, Linda M Zangwill.   

Abstract

OBJECTIVE: To describe and evaluate a technique to optimize scan centering during the Stratus optical coherence tomography (OCT) image acquisition process using currently available scan tracking coordinates.
DESIGN: Observational clinical study. PARTICIPANTS: Twelve eyes of six normal subjects were examined using the Fast retinal nerve fiber layer (RNFL) thickness and the Fast Optic Disc acquisition protocols.
METHODS: At visit 1, 3 consecutive measurements (trials) were taken by 2 different operators with the scan subjectively centered on the optic disc for the Fast RNFL thickness protocol and Fast Optic Disc protocol. At visit 2, 3 consecutive measurements were taken by positioning the scan using scan tracking coordinates. The scan coordinates were recorded twice by each operator and the limits of agreement and Bland-Altman plots were used to estimate agreement. The within subjects standard deviation (Sw) and the coefficient of variation (CV) were calculated for RNFL and optic disc parameters for each operator separately and differences by scan positioning method were evaluated using a 3-way (trial x operator x visit) analysis of variance for repeated measures.
RESULTS: The Sw and CV for the RNFL thickness parameters were generally higher when the scan was subjectively centered on the disc compared to when using the newly described coordinate system (eg, for operator 2, temporal sector Sw was 1.60+/-0.78 and 4.09+/-0.99 and CV was 2.2% and 5.7% with and without coordinate use, respectively). For the Fast RNFL protocol, the use of scan tracking coordinates resulted in significantly less variability than subjective placement of the scan circle using the landmark feature (currently recommended technique) in the temporal sectors only. No significant difference was found for any of the optic disc parameters. Bland-Altman plots showed good agreement within each operator for calculating scan coordinates suggesting this technique is reproducible.
CONCLUSIONS: Reproducibility of RNFL thickness measurements generally improves with the use of scan tracking coordinates, particularly in the temporal sector. However, small changes in the position of the scan do not significantly affect the reproducibility of optic disc parameters.

Entities:  

Mesh:

Year:  2009        PMID: 19142141      PMCID: PMC3408212          DOI: 10.1097/IJG.0b013e31816b3063

Source DB:  PubMed          Journal:  J Glaucoma        ISSN: 1057-0829            Impact factor:   2.503


  27 in total

1.  Increasing sampling density improves reproducibility of optical coherence tomography measurements.

Authors:  R Gurses-Ozden; H Ishikawa; S T Hoh; J M Liebmann; A Mistlberger; D S Greenfield; H L Dou; R Ritch
Journal:  J Glaucoma       Date:  1999-08       Impact factor: 2.503

Review 2.  Measuring agreement in method comparison studies.

Authors:  J M Bland; D G Altman
Journal:  Stat Methods Med Res       Date:  1999-06       Impact factor: 3.021

Review 3.  On making multiple comparisons in clinical and experimental pharmacology and physiology.

Authors:  J Ludbrook
Journal:  Clin Exp Pharmacol Physiol       Date:  1991-06       Impact factor: 2.557

4.  Optical coherence tomography.

Authors:  D Huang; E A Swanson; C P Lin; J S Schuman; W G Stinson; W Chang; M R Hee; T Flotte; K Gregory; C A Puliafito
Journal:  Science       Date:  1991-11-22       Impact factor: 47.728

5.  Evaluation of retinal nerve fiber layer, optic nerve head, and macular thickness measurements for glaucoma detection using optical coherence tomography.

Authors:  Felipe A Medeiros; Linda M Zangwill; Christopher Bowd; Roberto M Vessani; Remo Susanna; Robert N Weinreb
Journal:  Am J Ophthalmol       Date:  2005-01       Impact factor: 5.258

6.  Quantification of nerve fiber layer thickness in normal and glaucomatous eyes using optical coherence tomography.

Authors:  J S Schuman; M R Hee; C A Puliafito; C Wong; T Pedut-Kloizman; C P Lin; E Hertzmark; J A Izatt; E A Swanson; J G Fujimoto
Journal:  Arch Ophthalmol       Date:  1995-05

7.  Sensitivity and specificity of the StratusOCT for perimetric glaucoma.

Authors:  Donald L Budenz; Anika Michael; Robert T Chang; John McSoley; Joanne Katz
Journal:  Ophthalmology       Date:  2005-01       Impact factor: 12.079

8.  Comparison of the GDx VCC scanning laser polarimeter, HRT II confocal scanning laser ophthalmoscope, and stratus OCT optical coherence tomograph for the detection of glaucoma.

Authors:  Felipe A Medeiros; Linda M Zangwill; Christopher Bowd; Robert N Weinreb
Journal:  Arch Ophthalmol       Date:  2004-06

9.  Effect of improper scan alignment on retinal nerve fiber layer thickness measurements using Stratus optical coherence tomograph.

Authors:  Gianmarco Vizzeri; Christopher Bowd; Felipe A Medeiros; Robert N Weinreb; Linda M Zangwill
Journal:  J Glaucoma       Date:  2008-08       Impact factor: 2.503

10.  Quantitative assessment of structural damage in eyes with localized visual field abnormalities.

Authors:  Harmohina Bagga; David S Greenfield
Journal:  Am J Ophthalmol       Date:  2004-05       Impact factor: 5.258

View more
  2 in total

1.  Detection of glaucoma progression with stratus OCT retinal nerve fiber layer, optic nerve head, and macular thickness measurements.

Authors:  Felipe A Medeiros; Linda M Zangwill; Luciana M Alencar; Christopher Bowd; Pamela A Sample; Remo Susanna; Robert N Weinreb
Journal:  Invest Ophthalmol Vis Sci       Date:  2009-10-08       Impact factor: 4.799

2.  Agreement between spectral-domain and time-domain OCT for measuring RNFL thickness.

Authors:  G Vizzeri; R N Weinreb; A O Gonzalez-Garcia; C Bowd; F A Medeiros; P A Sample; L M Zangwill
Journal:  Br J Ophthalmol       Date:  2009-03-19       Impact factor: 4.638

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.