Literature DB >> 15940438

Biocompatibility screening in cardiovascular implants.

M Sigler1, T Paul, R G Grabitz.   

Abstract

BACKGROUND: Interest in information on biocompatibility of implants is increasing. The purpose of this paper is to discuss methods and results of pathological biocompatibility screening of explanted cardiovascular implants.
METHODS: Use of standard histology after embedding in paraffin is limited since metallic implants have to be removed during workup with disruption of the specimen. Alternatively, tissue blocks containing an implant can be embedded in methylmethacrylate or hydroxyethylmethacrylate and processed by sectioning with a diamond cutter and grinding, thus leaving the implant in situ and saving the tissue/implant interface for detection of local inflammatory reactions. Another important aspect of evaluation is the progress of thrombus organisation after initial fibrin clotting on the metal surface or in the inner part of occlusion devices. New methacrylate resins and embedding techniques allow for specific immunohistochemical staining of the specimen thus enabling characterisation of tissues surrounding the implant. Information on endothelialisation of the vascular surface of the implant can be obtained by means of immunohistochemistry or by scanning electron microscopy.
RESULTS: Illustrating the use of these technologies, we demonstrate findings in tissue specimens from animal studies with different types of devices (i.e. stents, occlusion devices). We present corresponding findings in human specimens with implants that were removed during corrective surgery for congenital heart defects. Early endothelialisation of the vascular surface was seen after implantation in all types of devices. Cells within occlusion devices could be characterised histologically and immunohistochemically as fibromuscular cells as seen in intimal hyperplasia after stent implantation. Inflammatory implant-host reactions ranged from mild to moderate (medical grade stainless steel, nitinol) to severe (polytetrafluoroethylene [PTFE]).
CONCLUSIONS: With an optimal work-up of cardiovascular implants, ingrowth and endothelialisation as well as inflammatory reactions in the surrounding tissue can be assessed. This information allows evaluation of individual tissue reactions to the implant and may serve as valuable basis for optimisation of biocompatibility by implant modification.

Entities:  

Mesh:

Substances:

Year:  2005        PMID: 15940438     DOI: 10.1007/s00392-005-0231-4

Source DB:  PubMed          Journal:  Z Kardiol        ISSN: 0300-5860


  37 in total

Review 1.  Endothelial cell function and thrombosis.

Authors:  J D Pearson
Journal:  Baillieres Best Pract Res Clin Haematol       Date:  1999-09

Review 2.  Coronary in-stent restenosis - predictors, treatment and prevention.

Authors:  R Hoffmann; G S Mintz
Journal:  Eur Heart J       Date:  2000-11       Impact factor: 29.983

3.  Clinical experience with autologous endothelial cell-seeded polytetrafluoroethylene coronary artery bypass grafts.

Authors:  H R Laube; J Duwe; W Rutsch; W Konertz
Journal:  J Thorac Cardiovasc Surg       Date:  2000-07       Impact factor: 5.209

Review 4.  Gene transfer strategies to inhibit neointima formation.

Authors:  Martin F Crook; Levent M Akyürek
Journal:  Trends Cardiovasc Med       Date:  2003-04       Impact factor: 6.677

5.  Intimal lesion formation in rat carotid arteries after endothelial denudation in absence of medial injury.

Authors:  J Fingerle; Y P Au; A W Clowes; M A Reidy
Journal:  Arteriosclerosis       Date:  1990 Nov-Dec

6.  [Reopening of a persistent left superior vena cava in the early postoperative period following bidirectional cavopulmonary anastomosis--treatment by coil embolization].

Authors:  I Michel-Behnke; H Akintürk; D Schranz
Journal:  Z Kardiol       Date:  1999-08

7.  The anti-aggregating properties of vascular endothelium: interactions between prostacyclin and nitric oxide.

Authors:  M W Radomski; R M Palmer; S Moncada
Journal:  Br J Pharmacol       Date:  1987-11       Impact factor: 8.739

Review 8.  Current anticoagulation options in percutaneous intervention: designing patient-specific strategies.

Authors:  Matthew H French; David P Faxon
Journal:  Rev Cardiovasc Med       Date:  2002       Impact factor: 2.930

9.  Evidence for circulating bone marrow-derived endothelial cells.

Authors:  Q Shi; S Rafii; M H Wu; E S Wijelath; C Yu; A Ishida; Y Fujita; S Kothari; R Mohle; L R Sauvage; M A Moore; R F Storb; W P Hammond
Journal:  Blood       Date:  1998-07-15       Impact factor: 22.113

10.  Similarity in behavior of polytetrafluoroethylene (ePTFE) prostheses implanted into different interfaces.

Authors:  J M Bellón; J Buján; L A Contreras; A Hernando; F Jurado
Journal:  J Biomed Mater Res       Date:  1996-05
View more
  11 in total

Review 1.  Annular management during aortic valve repair: a systematic review.

Authors:  Takashi Kunihara
Journal:  Gen Thorac Cardiovasc Surg       Date:  2015-11-25

2.  Histopathological workup of an Amplatzer atrial septal defect occluder after surgical removal.

Authors:  Matthias Sigler; Christian Jux; Peter Ewert
Journal:  Pediatr Cardiol       Date:  2006-10-09       Impact factor: 1.655

3.  Optical coherence tomography provides images similar to histology and allows the performance of extensive measurements of drug-eluting metal stents in animal ureters.

Authors:  P Kallidonis; G C Kagadis; P Kitrou; A Tsamandas; I Kyriazis; I Georgiopoulos; D Karnabatidis; S Tsantis; D Liourdi; A Al-Aown; E Liatsikos
Journal:  Lasers Med Sci       Date:  2014-03-04       Impact factor: 3.161

4.  Secondary residual shunt after atrial septal defect closure with an amplatzer occluder: surgical removal and evaluation of device biocompatibility after 7 years.

Authors:  Weihua Zhu; Henning Neubauer
Journal:  Pediatr Cardiol       Date:  2010-08-01       Impact factor: 1.655

5.  Biocompatibility of septal defect closure devices.

Authors:  Matthias Sigler; Christian Jux
Journal:  Heart       Date:  2006-10-11       Impact factor: 5.994

6.  Comparison of Two Percutaneous Atrial Septal Defect Occluders for Device Healing and Nickel Release in a Chronic Porcine Model.

Authors:  Zakaria Jalal; Younes Boudjemline; Xavier Iriart; Pierre-Emmanuel Seguela; Samantha Delmond; Virgine Loyer; David Gonthier; Olivier Bernus; Marlène Durand; Laurence Bordenave; Jean-Benoit Thambo
Journal:  J Interv Cardiol       Date:  2020-06-20       Impact factor: 2.279

Review 7.  Role of animal models for percutaneous atrial septal defect closure.

Authors:  Zakaria Jalal; Pierre-Emmanuel Seguela; Alban-Elouen Baruteau; David Benoist; Olivier Bernus; Olivier Villemain; Younes Boudjemline; Xavier Iriart; Jean-Benoit Thambo
Journal:  J Thorac Dis       Date:  2018-09       Impact factor: 2.895

Review 8.  Biomaterials and host versus graft response: a short review.

Authors:  Tomaz Velnar; Gorazd Bunc; Robert Klobucar; Lidija Gradisnik
Journal:  Bosn J Basic Med Sci       Date:  2016-02-19       Impact factor: 3.363

9.  Transvenous closure of patent foramen ovale: preliminary results with a new self-expanding nitinol wire mesh in a Swine model.

Authors:  F Krizanic; M Sigler; H R Figulla
Journal:  Cardiol Res Pract       Date:  2009-09-09       Impact factor: 1.866

10.  Stent implantation into the tracheo-bronchial system in rabbits: histopathologic sequelae in bare metal vs. drug-eluting stents.

Authors:  Matthias Sigler; Julia Klötzer; Thomas Quentin; Thomas Paul; Oliver Möller
Journal:  Mol Cell Pediatr       Date:  2015-11-05
View more