Literature DB >> 15875117

Interference by cellular telephone with an implantable loop recorder.

Alexandre Trigano1, Olivier Blandeau, Samuel Levy.   

Abstract

We report the observation of interference between an implanted loop recorder and a cellular telephone when the latter was placed over the subcutaneous pocket. The artifact produced during the ringing of the telephone was automatically recorded and stored in the device memory. A similar interference was documented during ex vivo testing.

Mesh:

Year:  2005        PMID: 15875117     DOI: 10.1007/s10840-005-0303-4

Source DB:  PubMed          Journal:  J Interv Card Electrophysiol        ISSN: 1383-875X            Impact factor:   1.900


  6 in total

1.  Electromagnetic interference of an implantable loop recorder by commonly encountered electronic devices.

Authors:  C C de Cock; H J Spruijt; L M van Campen; A W Plu; C A Visser
Journal:  Pacing Clin Electrophysiol       Date:  2000-10       Impact factor: 1.976

2.  Electromagnetic interference of pacemakers by mobile phones.

Authors:  W Irnich; L Batz; R Müller; R Tobisch
Journal:  Pacing Clin Electrophysiol       Date:  1996-10       Impact factor: 1.976

3.  Intermittent pacemaker dysfunction caused by digital mobile telephones.

Authors:  B Naegeli; S Osswald; M Deola; F Burkart
Journal:  J Am Coll Cardiol       Date:  1996-05       Impact factor: 24.094

4.  Interference with cardiac pacemakers by cellular telephones.

Authors:  D L Hayes; P J Wang; D W Reynolds; M Estes; J L Griffith; R A Steffens; G L Carlo; G K Findlay; C M Johnson
Journal:  N Engl J Med       Date:  1997-05-22       Impact factor: 91.245

5.  Artefact mimicking tachycardia during magnetic resonance imaging in a patient with an implantable loop recorder.

Authors:  J R Gimbel; B L Wilkoff
Journal:  Heart       Date:  2003-03       Impact factor: 5.994

6.  Arrhythmia detection by patient and auto-activation in implantable loop recorders.

Authors:  Ernest Ng; Peter J Stafford; G André Ng
Journal:  J Interv Card Electrophysiol       Date:  2004-04       Impact factor: 1.900

  6 in total
  1 in total

1.  Inappropriate asystole detection in early postoperative phase after loop recorder implantation.

Authors:  Miriam Bortnik; Eraldo Occhetta; Andrea Magnani; Anna Degiovanni; Paolo Marino
Journal:  ISRN Cardiol       Date:  2011-04-17
  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.