Impedance-gradient electrode reduces skin irritation induced by transthoracic defibrillation.

P F Meyer1, P D Gadsby, D Van Sickle, W E Schoenlein, K S Foster, G P Graber.   

Abstract

A new type of disposable external defibrillation electrode has been developed to reduce the skin irritation commonly associated with defibrillation and synchronised cardioversion. This design employs an impedance gradient to reduce the proportion of current delivered to the electrode periphery. The temperature distribution under the new electrode was compared with that of four other types of commercially available electrodes after repeated high-energy biphasic defibrillation discharges to domestic swine. Skin temperature distributions were acquired using non-invasive thermography. Measurements of the maximum temperature rise at each electrode site, taken 3.6s after the fifth defibrillation discharge, demonstrated that the new impedance-gradient electrode produced 50-60% less skin heating than two of the three uniform-impedance electrode designs. Histological examination of erythematous sites excised 24 h after defibrillation quantified the associated skin damage using a scoring protocol developed for this study. In contrast to previous studies, histological examinations demonstrated second-degree skin burns following defibrillation. The new electrode design, however, induced 44-46% less skin damage than two of the traditional uniform-impedance electrodes.

Entities:  

Mesh:

Year:  2005        PMID: 15865132     DOI: 10.1007/bf02345959

Source DB:  PubMed          Journal:  Med Biol Eng Comput        ISSN: 0140-0118            Impact factor:   2.602


  12 in total

1.  Electrical current distribution under transthoracic defibrillation and pacing electrodes.

Authors:  S Papazov; Z Kostov; I Daskalov
Journal:  J Med Eng Technol       Date:  2002 Jan-Feb

2.  Effects of thermal properties and geometrical dimensions on skin burn injuries.

Authors:  S C Jiang; N Ma; H J Li; X X Zhang
Journal:  Burns       Date:  2002-12       Impact factor: 2.744

3.  In vitro measurement and characterization of current density profiles produced by non-recessed, simple recessed, and radially varying recessed stimulating electrodes.

Authors:  M F Suesserman; F A Spelman; J T Rubinstein
Journal:  IEEE Trans Biomed Eng       Date:  1991-05       Impact factor: 4.538

4.  High perimeter impedance defibrillation electrodes reduce skin burns in transthoracic cardioversion.

Authors:  L A Garcia; L A Pagan-Carlo; M S Stone; R E Kerber
Journal:  Am J Cardiol       Date:  1998-11-01       Impact factor: 2.778

5.  Skin-temperature distribution under a new type of defibrillating electrode.

Authors:  J D Bourland; L A Geddes; G P Graber; K S Foster; W E Schoenlein
Journal:  Biomed Instrum Technol       Date:  1998 Mar-Apr

6.  Nature and determinants of skin "burns" after transthoracic cardioversion.

Authors:  L A Pagan-Carlo; M S Stone; R E Kerber
Journal:  Am J Cardiol       Date:  1997-03-01       Impact factor: 2.778

Review 7.  Current density profiles of surface mounted and recessed electrodes for neural prostheses.

Authors:  J T Rubinstein; F A Spelman; M Soma; M F Suesserman
Journal:  IEEE Trans Biomed Eng       Date:  1987-11       Impact factor: 4.538

8.  Analysis and control of the current distribution under circular dispersive electrodes.

Authors:  J D Wiley; J G Webster
Journal:  IEEE Trans Biomed Eng       Date:  1982-05       Impact factor: 4.538

9.  Biphasic versus monophasic shock waveform for conversion of atrial fibrillation: the results of an international randomized, double-blind multicenter trial.

Authors:  Richard L Page; Richard E Kerber; James K Russell; Tom Trouton; Johan Waktare; Donna Gallik; Jeff E Olgin; Philippe Ricard; Gavin W Dalzell; Ramakota Reddy; Ralph Lazzara; Kerry Lee; Mark Carlson; Blair Halperin; Gust H Bardy
Journal:  J Am Coll Cardiol       Date:  2002-06-19       Impact factor: 24.094

10.  Estimation of current density distribution under electrodes for external defibrillation.

Authors:  Vessela Tz Krasteva; Sava P Papazov
Journal:  Biomed Eng Online       Date:  2002-12-16       Impact factor: 2.819

View more
  3 in total

1.  Reduction of edge effect on disk electrodes by optimized current waveform.

Authors:  Boshuo Wang; Artin Petrossians; James D Weiland
Journal:  IEEE Trans Biomed Eng       Date:  2014-08       Impact factor: 4.538

2.  Focused RF hyperthermia using magnetic fluids.

Authors:  T Onur Tasci; Ibrahim Vargel; Anil Arat; Elif Guzel; Petek Korkusuz; Ergin Atalar
Journal:  Med Phys       Date:  2009-05       Impact factor: 4.071

3.  A miniaturized solid salt reverse electrodialysis battery: a durable and fully ionic power source.

Authors:  Song Yi Yeon; Jeongse Yun; Sun-Heui Yoon; Dahye Lee; Woohyuk Jang; Seok Hee Han; Chung Mu Kang; Taek Dong Chung