Nature and determinants of skin "burns" after transthoracic cardioversion.

L A Pagan-Carlo1, M S Stone, R E Kerber.   

Abstract

Skin biopsies obtained 24 hours after elective cardioversion of 30 patients showed variable epidermal necrosis and upper dermal perivascular inflammation, most noticeably in patients receiving high individual peak (> or = 300 J) and cumulative (> or = 350 J) shock energies. Thus, damped sine wave shocks cause skin injury--first degree burns--the severity of which is a function of peak and cumulative shock energy.

Entities:  

Mesh:

Year:  1997        PMID: 9068538     DOI: 10.1016/s0002-9149(96)00845-4

Source DB:  PubMed          Journal:  Am J Cardiol        ISSN: 0002-9149            Impact factor:   2.778


  6 in total

1.  Impedance-gradient electrode reduces skin irritation induced by transthoracic defibrillation.

Authors:  P F Meyer; P D Gadsby; D Van Sickle; W E Schoenlein; K S Foster; G P Graber
Journal:  Med Biol Eng Comput       Date:  2005-03       Impact factor: 2.602

2.  Reduction of edge effect on disk electrodes by optimized current waveform.

Authors:  Boshuo Wang; Artin Petrossians; James D Weiland
Journal:  IEEE Trans Biomed Eng       Date:  2014-08       Impact factor: 4.538

3.  Breast calcifications following electrical defibrillation: An unusual mammographic appearance.

Authors:  Steven M Westphal; Manish Jani; Sunil Badve
Journal:  Radiol Case Rep       Date:  2015-11-06

4.  Investigating the efficacy of chest pressure for direct current cardioversion in atrial fibrillation: a randomised control trial protocol (Pressure-AF).

Authors:  David Ferreira; Philo Mikhail; Michael McGee; Andrew Boyle; Aaron Sverdlov; Maged William; Nicholas Jackson; Malcolm Barlow; James Leitch; Nicholas Collins; Thomas Ford; Bradley Wilsmore
Journal:  Open Heart       Date:  2021-09

5.  Hands-On defibrillation-the end of "i'm clear, you're clear, we're all clear"?

Authors:  Richard E Kerber
Journal:  J Am Heart Assoc       Date:  2012-10-25       Impact factor: 5.501

6.  Estimation of current density distribution under electrodes for external defibrillation.

Authors:  Vessela Tz Krasteva; Sava P Papazov
Journal:  Biomed Eng Online       Date:  2002-12-16       Impact factor: 2.819

  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.