Analysis of a retrieved polyethylene total disc replacement component.

Steven M Kurtz1, John Peloza, Ryan Siskey, Marta L Villarraga.   

Abstract

BACKGROUND CONTEXT: Although total disc replacements have been performed in Europe since the 1980s, this type of surgery is still new in the United States. The clinical performance of polyethylene in total disc replacements is still not well understood.
PURPOSE: To describe the wear, surface damage, oxidation and mechanical properties in an explanted polyethylene total disc replacement component. STUDY DESIGN/
SETTING: Case report, analysis of retrieved implant. PATIENT SAMPLE: Case report. OUTCOME MEASURES: Analysis of wear, oxidation and mechanical properties in the retrieved total disc replacement.
METHODS: A 49-year-old female patient was implanted at L5-S1 with an SB Charite total disc prosthesis (DePuy Spine, Raynham, MA). After 1.6 years, the patient underwent a posterior, instrumented fusion because of intractable low back, left buttock and radicular left leg pain. Preoperative diagnostics revealed loosening at the bone implant interface at L5 and S1, anterior migration of the L5 base plate and severe degeneration of the L5-S1 facet joints. The retrieved polyethylene core showed evidence of damage around the periphery or rim. Transverse, subsurface cracks in the polyethylene, which initiated near the rim and penetrated into the interior of the component, were imaged using thin-film optical microscopy and micro-computed-tomography imaging. Analysis using Fourier transform infrared spectroscopy (American Society for Testing and Materials [ASTM] F2102) documented low levels of oxidation within 1 mm of the articulating surface. Miniature specimen mechanical testing (ASTM F2183), conducted near the surface where the oxidation was greatest, demonstrated that the mechanical properties were not substantially degraded.
CONCLUSION: In this case, the anterior revision surgery was difficult and potentially life-threatening. The revision strategy of an instrumented posterior fusion to salvage a failed SB Charite disc replacement may be unpredictable and, in this case, ultimately unsuccessful. Despite the small size of the retrieved polyethylene core, ASTM standard test techniques developed for analysis of retrieved hip and knee replacements were readily adapted for the total disc prosthesis.

Entities:  

Mesh:

Substances:

Year:  2005        PMID: 15863092     DOI: 10.1016/j.spinee.2004.11.011

Source DB:  PubMed          Journal:  Spine J        ISSN: 1529-9430            Impact factor:   4.166


  19 in total

1.  Preferential superior surface motion in wear simulations of the Charité total disc replacement.

Authors:  Curtis M Goreham-Voss; Rachel Vicars; Richard M Hall; Thomas D Brown
Journal:  Eur Spine J       Date:  2010-06-26       Impact factor: 3.134

2.  Influence of the loading frequency on the wear rate of a polyethylene-on-metal lumbar intervertebral disc replacement.

Authors:  Annette Kettler; Michael Bushelow; Hans-Joachim Wilke
Journal:  Eur Spine J       Date:  2010-10-10       Impact factor: 3.134

Review 3.  Design concepts in lumbar total disc arthroplasty.

Authors:  Fabio Galbusera; Chiara M Bellini; Thomas Zweig; Stephen Ferguson; Manuela T Raimondi; Claudio Lamartina; Marco Brayda-Bruno; Maurizio Fornari
Journal:  Eur Spine J       Date:  2008-10-23       Impact factor: 3.134

Review 4.  Strategies towards injectable, load-bearing materials for the intervertebral disc: a review and outlook.

Authors:  Cecilia Persson; Svante Berg
Journal:  J Mater Sci Mater Med       Date:  2012-09-29       Impact factor: 3.896

5.  Circumferential dynamic stabilization of the lumbar spine: a biomechanical analysis.

Authors:  Wolfram Käfer; Balkan Cakir; Stefan Midderhoff; Heiko Reichel; Hans-Joachim Wilke
Journal:  Eur Spine J       Date:  2014-04-11       Impact factor: 3.134

6.  Are periprosthetic tissue reactions observed after revision of total disc replacement comparable to the reactions observed after total hip or knee revision surgery?

Authors:  Ilona M Punt; Shennah Austen; Jack P M Cleutjens; Steven M Kurtz; René H M ten Broeke; Lodewijk W van Rhijn; Paul C Willems; André van Ooij
Journal:  Spine (Phila Pa 1976)       Date:  2012-01-15       Impact factor: 3.468

Review 7.  Total disc replacement in the lumbar spine: a systematic review of the literature.

Authors:  Brian J C Freeman; James Davenport
Journal:  Eur Spine J       Date:  2006-07-22       Impact factor: 3.134

8.  What is the correlation of in vivo wear and damage patterns with in vitro TDR motion response?

Authors:  Steven M Kurtz; Avinash Patwardhan; Daniel MacDonald; Lauren Ciccarelli; André van Ooij; Mark Lorenz; Michael Zindrick; Patrick O'Leary; Jorge Isaza; Raymond Ross
Journal:  Spine (Phila Pa 1976)       Date:  2008-03-01       Impact factor: 3.468

9.  Endosomal damage and TLR2 mediated inflammasome activation by alkane particles in the generation of aseptic osteolysis.

Authors:  Radhashree Maitra; Cristina C Clement; Brian Scharf; Giovanna M Crisi; Sriram Chitta; Daniel Paget; P Edward Purdue; Neil Cobelli; Laura Santambrogio
Journal:  Mol Immunol       Date:  2009-10-04       Impact factor: 4.407

10.  Spine imaging after lumbar disc replacement: pitfalls and current recommendations.

Authors:  Yohan Robinson; Bengt Sandén
Journal:  Patient Saf Surg       Date:  2009-07-20
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.