Measurement of scatter fractions in clinical bedside radiography.

C E Floyd1, J A Baker, J Y Lo, C E Ravin.   

Abstract

The authors present measurements of scatter fraction (SF), the ratio of scattered to total imaged photons, from clinical bedside radiographs of 102 patients. These measurements were obtained by using a new posterior beam-stop technique that does not alter the diagnostic image but that simultaneously provides SF measurements at 224 locations in the image. The SF values in the lung were found to be consistent with previous measurements, while the SF values in the mediastinal and retrocardiac areas were larger than previously reported. SFs in diseased lung were significantly larger than SFs in normal lung. The range of SF values was large for all anatomic locations. For applications in which accurate scatter estimation is required, this wide range of values suggests that SFs should be measured in each individual image.

Entities:  

Mesh:

Year:  1992        PMID: 1584947     DOI: 10.1148/radiology.183.3.1584947

Source DB:  PubMed          Journal:  Radiology        ISSN: 0033-8419            Impact factor:   11.105


  7 in total

1.  A method to produce and validate a digitally reconstructed radiograph-based computer simulation for optimisation of chest radiographs acquired with a computed radiography imaging system.

Authors:  C S Moore; G P Liney; A W Beavis; J R Saunderson
Journal:  Br J Radiol       Date:  2011-10       Impact factor: 3.039

2.  Rejection and redistribution of scattered radiation in scan equalization digital radiography (SEDR): simulation with spot images.

Authors:  Xinming Liu; Chris C Shaw
Journal:  Med Phys       Date:  2007-07       Impact factor: 4.071

3.  Scatter rejection and low-contrast performance of a slot-scan digital chest radiography system with electronic aft-collimation: a chest phantom study.

Authors:  Xinming Liu; Chris C Shaw; Chao-Jen Lai; Mustafa C Altunbas; Lingyun Chen; Tao Han; Tianpeng Wang
Journal:  Med Phys       Date:  2008-06       Impact factor: 4.071

4.  Effective DQE (eDQE) and dose to optimize radiographic technical parameters: a survey of pediatric chest X-ray examinations in Korea.

Authors:  Hye-Suk Park; Ye-Seul Kim; Ok-Seob Park; Sang-Tae Kim; Chang-Woo Jeon; Hee-Joung Kim
Journal:  Radiol Med       Date:  2013-12-12       Impact factor: 3.469

5.  Comparison of scatter rejection and low-contrast performance of scan equalization digital radiography (SEDR), slot-scan digital radiography, and full-field digital radiography systems for chest phantom imaging.

Authors:  Xinming Liu; Chris C Shaw; Chao-Jen Lai; Tianpeng Wang
Journal:  Med Phys       Date:  2011-01       Impact factor: 4.071

6.  An alternate line erasure and readout (ALER) method for implementing slot-scan imaging technique with a flat-panel detector--initial experiences.

Authors:  Xinming Liu; Chris C Shaw; Mustafa C Altunbas; Tianpeng Wang
Journal:  IEEE Trans Med Imaging       Date:  2006-04       Impact factor: 10.048

7.  The Potential Role of Grid-Like Software in Bedside Chest Radiography in Improving Image Quality and Dose Reduction: An Observer Preference Study.

Authors:  Su Yeon Ahn; Kum Ju Chae; Jin Mo Goo
Journal:  Korean J Radiol       Date:  2018-04-06       Impact factor: 3.500

  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.