Digital slot-scan charge-coupled device radiography versus AMBER and Bucky screen-film radiography: comparison of image quality in a phantom study.

Wouter J H Veldkamp1, Lucia J M Kroft, Bart J A Mertens, Jacob Geleijns.   

Abstract

PURPOSE: To evaluate the image quality and performance of a chest digital radiography system and to compare this with the image quality and performance of advanced multiple-beam equalization radiography (AMBER) and Bucky screen-film radiography systems.
MATERIALS AND METHODS: The chest digital radiography system is a digital charge-coupled device (CCD) chest imaging unit that uses slot-scan technology. A contrast-detail test object was used in combination with a phantom that simulates the primary and scatter transmission for the lungs and mediastinum. Twelve phantom images were obtained with each modality (ie, CCD digital radiography and AMBER and Bucky screen-film radiography) and were judged by six observers. CCD digital radiography soft-copy reading was compared with AMBER hard-copy reading. To measure image quality, contrast-detail curves were constructed from the observer data. The Wilcoxon signed rank test was used for statistical analysis.
RESULTS: For the lung configuration, contrast-detail curves showed lower threshold depth for hard-copy images obtained with CCD digital radiography than for those obtained with Bucky screen-film radiography. For hard-copy images, the difference between contrast-detail curves for CCD digital radiography and those for Bucky screen-film radiography was statistically significant (P < .006). No significant difference was found between CCD digital radiography and AMBER for hard-copy images obtained in either the lung or mediastinum configuration. For the lung configuration, a lower threshold depth was observed for CCD digital radiography soft-copy reading than for AMBER hard-copy reading, with significantly different contrast-detail curves for CCD digital radiography soft copy and AMBER hard copy (P < .006). No significant difference was found between either system for the mediastinum configuration.
CONCLUSION: Contrast-detail curves indicate that image quality for the CCD chest system provides a digital alternative to AMBER and Bucky screen-film radiography. Copyright RSNA, 2005.

Entities:  

Mesh:

Year:  2005        PMID: 15845787     DOI: 10.1148/radiol.2353031919

Source DB:  PubMed          Journal:  Radiology        ISSN: 0033-8419            Impact factor:   11.105


  7 in total

1.  Contrast-detail evaluation and dose assessment of eight digital chest radiography systems in clinical practice.

Authors:  Wouter J H Veldkamp; Lucia J M Kroft; Mireille V Boot; Bart J A Mertens; Jacob Geleijns
Journal:  Eur Radiol       Date:  2005-08-31       Impact factor: 5.315

2.  Detection of simulated microcalcifications in a phantom with digital mammography: effect of pixel size.

Authors:  Sankararaman Suryanarayanan; Andrew Karellas; Srinivasan Vedantham; Ioannis Sechopoulos; Carl J D'Orsi
Journal:  Radiology       Date:  2007-05-23       Impact factor: 11.105

3.  Rejection and redistribution of scattered radiation in scan equalization digital radiography (SEDR): simulation with spot images.

Authors:  Xinming Liu; Chris C Shaw
Journal:  Med Phys       Date:  2007-07       Impact factor: 4.071

4.  A technique for simulating the effect of dose reduction on image quality in digital chest radiography.

Authors:  Wouter J H Veldkamp; Lucia J M Kroft; Jan Pieter A van Delft; Jacob Geleijns
Journal:  J Digit Imaging       Date:  2008-02-08       Impact factor: 4.056

5.  Scatter rejection and low-contrast performance of a slot-scan digital chest radiography system with electronic aft-collimation: a chest phantom study.

Authors:  Xinming Liu; Chris C Shaw; Chao-Jen Lai; Mustafa C Altunbas; Lingyun Chen; Tao Han; Tianpeng Wang
Journal:  Med Phys       Date:  2008-06       Impact factor: 4.071

6.  Comparison of scatter rejection and low-contrast performance of scan equalization digital radiography (SEDR), slot-scan digital radiography, and full-field digital radiography systems for chest phantom imaging.

Authors:  Xinming Liu; Chris C Shaw; Chao-Jen Lai; Tianpeng Wang
Journal:  Med Phys       Date:  2011-01       Impact factor: 4.071

7.  Experimental realization of fluence field modulated CT using digital beam attenuation.

Authors:  T P Szczykutowicz; C A Mistretta
Journal:  Phys Med Biol       Date:  2014-02-20       Impact factor: 3.609

  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.