Literature DB >> 15814942

Changes in the intraaneurysmal pressure due to HydroCoil embolization.

Gádor Cantón1, David I Levy, Juan C Lasheras.   

Abstract

BACKGROUND AND
PURPOSE: The purpose of this study was to measure the changes in the intraaneurysmal fluid pressure and parent vessel flow characteristics resulting from packing the aneurysmal sac with hydrogel-coated coils.
METHODS: Platinum coils coated with an expansible hydrogel were used to embolize a silicone model of the basilar tip aneurysm. The intraaneurysmal fluid pressure was measured with a NeuroCare Camino Pressure Micro-Probe at various packing ratios. A programmable pulsatile pump was used to input a physiologically relevant pulsatile flow in the parent artery.
RESULTS: The intraaneurysmal fluid pressure did not increase when packing the aneurysm with hydrogel-coated platinum coils, even with a coil density up to 93%.
CONCLUSION: Packing the aneurysm with hydrogel-coated coils at a density up to 93% did not increase the intraaneurysmal fluid pressure.

Entities:  

Mesh:

Substances:

Year:  2005        PMID: 15814942      PMCID: PMC7977101     

Source DB:  PubMed          Journal:  AJNR Am J Neuroradiol        ISSN: 0195-6108            Impact factor:   3.825


  17 in total

1.  A comparison between endovascular and surgical management of basilar artery apex aneurysms.

Authors:  D P Gruber; G A Zimmerman; T A Tomsick; H R van Loveren; M J Link; J M Tew
Journal:  J Neurosurg       Date:  1999-05       Impact factor: 5.115

Review 2.  Treatment of intracranial aneurysms by embolization with coils: a systematic review.

Authors:  E H Brilstra; G J Rinkel; Y van der Graaf; W J van Rooij; A Algra
Journal:  Stroke       Date:  1999-02       Impact factor: 7.914

3.  Treatment of unruptured cerebral aneurysms in California.

Authors:  S C Johnston; S Zhao; R A Dudley; M F Berman; D R Gress
Journal:  Stroke       Date:  2001-03       Impact factor: 7.914

4.  Pulsation-pressure relationship in experimental aneurysms: observation of aneurysmal hysteresis.

Authors:  P Novak; R Glikstein; G Mohr
Journal:  Neurol Res       Date:  1996-08       Impact factor: 2.448

5.  Gross and microscopic histopathological findings in aneurysms of the human brain treated with Guglielmi detachable coils.

Authors:  G Bavinzski; V Talazoglu; M Killer; B Richling; A Gruber; C E Gross; H Plenk
Journal:  J Neurosurg       Date:  1999-08       Impact factor: 5.115

6.  New expandable hydrogel-platinum coil hybrid device for aneurysm embolization.

Authors:  David F Kallmes; Naomi H Fujiwara
Journal:  AJNR Am J Neuroradiol       Date:  2002-10       Impact factor: 3.825

7.  Endovascular treatment of unruptured aneurysms.

Authors:  D Roy; G Milot; J Raymond
Journal:  Stroke       Date:  2001-09       Impact factor: 7.914

8.  Ex vivo study of the physical effect of coils on pressure and flow dynamics in experimental aneurysms.

Authors:  H G Boecher-Schwarz; K Ringel; L Kopacz; A Heimann; O Kempski
Journal:  AJNR Am J Neuroradiol       Date:  2000-09       Impact factor: 3.825

9.  Effect of Guglielmi detachable coil placement on intraaneurysmal pressure: experimental study in canines.

Authors:  A Sorteberg; W Sorteberg; A S Turk; A Rappe; P H Nakstad; C M Strother
Journal:  AJNR Am J Neuroradiol       Date:  2001-10       Impact factor: 3.825

10.  Aneurysm packing with HydroCoil Embolic System versus platinum coils: initial clinical experience.

Authors:  Harry J Cloft; David F Kallmes
Journal:  AJNR Am J Neuroradiol       Date:  2004-01       Impact factor: 3.825

View more
  8 in total

1.  HydroCoil as an adjuvant to bare platinum coil treatment of 100 cerebral aneurysms.

Authors:  Noel F Fanning; Zsolt Berentei; Paul R Brennan; John Thornton
Journal:  Neuroradiology       Date:  2006-11-22       Impact factor: 2.804

2.  Changes of intra-aneurysmal pressure during coiling.

Authors:  Ajay K Wakhloo; Baruch B Lieber
Journal:  AJNR Am J Neuroradiol       Date:  2006-03       Impact factor: 3.825

3.  In vitro study of near-wall flow in a cerebral aneurysm model with and without coils.

Authors:  L Goubergrits; B Thamsen; A Berthe; J Poethke; U Kertzscher; K Affeld; C Petz; H-C Hege; H Hoch; A Spuler
Journal:  AJNR Am J Neuroradiol       Date:  2010-05-20       Impact factor: 3.825

4.  Transvenous treatment of cranial dural arteriovenous fistulas with hydrogel coated coils.

Authors:  P Klurfan; T Gunnarsson; I Shelef; K G Terbrugge; R A Willinsky
Journal:  Interv Neuroradiol       Date:  2007-01-19       Impact factor: 1.610

5.  Challenging the surgical rodent hindlimb ischemia model with the miniinterventional technique.

Authors:  Zhen W Zhuang; Jing Shi; John M Rhodes; Michael J Tsapakos; Michael Simons
Journal:  J Vasc Interv Radiol       Date:  2011-04-03       Impact factor: 3.464

6.  The Hydrodynamic Effect of Bare Platinum Coils versus Hydrogel-Coated Platinum Coils.

Authors:  A Sorteberg; S J Bakke
Journal:  Interv Neuroradiol       Date:  2009-01-05       Impact factor: 1.610

Review 7.  Silicone models as basic training and research aid in endovascular neurointervention--a single-center experience and review of the literature.

Authors:  Srinivasan Paramasivam; Gerasimos Baltsavias; Evlampia Psatha; Georgios Matis; Anton Valavanis
Journal:  Neurosurg Rev       Date:  2014-01-25       Impact factor: 3.042

8.  Usefulness of Hydrogel-Coated Coils in Embolization of Pulmonary Arteriovenous Malformations.

Authors:  Masashi Shimohira; Tatsuya Kawai; Takuya Hashizume; Masahiro Muto; Masanori Kitase; Yuta Shibamoto
Journal:  Cardiovasc Intervent Radiol       Date:  2018-01-17       Impact factor: 2.740

  8 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.