Effect of Guglielmi detachable coil placement on intraaneurysmal pressure: experimental study in canines.

A Sorteberg1, W Sorteberg, A S Turk, A Rappe, P H Nakstad, C M Strother.   

Abstract

BACKGROUND AND
PURPOSE: Treatment of ruptured aneurysms with Guglielmi detachable coils (GDCs) has been shown to prevent repeat bleeding. To assess whether GDC coiling alters aneurysmal pressure, we measured intraaneurysmal pressure in two canine types of carotid artery aneurysms before and after GDC placement.
METHODS: A 0.014-inch guidewire with a pressure transducer was inserted into parent arteries and domes of surgically created aneurysms. Intravascular static pressures were recorded before and during saline power injections (10, 20, and 30 mL over 1 and 2 s), before and after GDC placement. Common femoral arterial pressure was monitored.
RESULTS: Saline power injections reproducibly and abruptly increased pressure in parent arteries and aneurysms. Mean intraaneurysmal pressure varied (18 +/- 4 [10 mL] to 75 +/- 15 mm Hg [30 mL]), independent of injection duration. Intraaneurysmal baseline pressures were higher after GDC placement (111 +/- 10 versus 93 +/- 15 mm Hg; P =.05). Aneurysmal pressure increases with saline injections were slightly higher after GDC placement, which dampened intraaneurysmal pressure amplitude at baseline (26.5 +/- 5.6 versus 19.6 +/- 7.4 mm Hg; P =.003) and during hypertension (25.3 +/- 5.4 versus 19.8 +/- 7.5 mm Hg, P =.002). The pressure increase slope with saline injection was delayed with GDC placement (0.24 +/- 0.1 versus 0.38 +/- 0.19 s; P <.001).
CONCLUSION: Graded saline power injections into parent arteries can rapidly increase intraaneurysmal pressure. GDC treatment did not attenuate mean intraaneurysmal pressures, but both dampened the pressure amplitude and delayed pressure increases during locally induced hypertension.

Entities:  

Mesh:

Year:  2001        PMID: 11673173      PMCID: PMC7974449     

Source DB:  PubMed          Journal:  AJNR Am J Neuroradiol        ISSN: 0195-6108            Impact factor:   3.825


  13 in total

1.  Experimental production of carotid aneurysms.

Authors:  W J GERMAN; S P BLACK
Journal:  N Engl J Med       Date:  1954-01-21       Impact factor: 91.245

2.  Pathophysiology of development and rupture of cerebral aneurysms.

Authors:  H J Steiger
Journal:  Acta Neurochir Suppl (Wien)       Date:  1990

3.  Treatment of carotid artery aneurysms with platinum coils: an experimental study in dogs.

Authors:  V B Graves; C R Partington; D A Rüfenacht; A H Rappe; C M Strother
Journal:  AJNR Am J Neuroradiol       Date:  1990 Mar-Apr       Impact factor: 3.825

Review 4.  Treatment of intracranial aneurysms by embolization with coils: a systematic review.

Authors:  E H Brilstra; G J Rinkel; Y van der Graaf; W J van Rooij; A Algra
Journal:  Stroke       Date:  1999-02       Impact factor: 7.914

5.  Pulsation-pressure relationship in experimental aneurysms: observation of aneurysmal hysteresis.

Authors:  P Novak; R Glikstein; G Mohr
Journal:  Neurol Res       Date:  1996-08       Impact factor: 2.448

6.  Controlled pressure-volume factors in the enlargement of intracranial aneurysms.

Authors:  G M Austin; W Schievink; R Williams
Journal:  Neurosurgery       Date:  1989-05       Impact factor: 4.654

7.  Ex vivo study of the physical effect of coils on pressure and flow dynamics in experimental aneurysms.

Authors:  H G Boecher-Schwarz; K Ringel; L Kopacz; A Heimann; O Kempski
Journal:  AJNR Am J Neuroradiol       Date:  2000-09       Impact factor: 3.825

8.  Early treatment of ruptured aneurysms with Guglielmi detachable coils: effect on subsequent bleeding.

Authors:  V B Graves; C M Strother; T A Duff; J Perl
Journal:  Neurosurgery       Date:  1995-10       Impact factor: 4.654

9.  Computer modeling of intracranial saccular and lateral aneurysms for the study of their hemodynamics.

Authors:  A C Burleson; C M Strother; V T Turitto
Journal:  Neurosurgery       Date:  1995-10       Impact factor: 4.654

Review 10.  Etiology of intracranial berry aneurysms.

Authors:  W E Stehbens
Journal:  J Neurosurg       Date:  1989-06       Impact factor: 5.115

View more
  12 in total

1.  Recording CSF pressure with a transducer-tipped wire in an animal model of Chiari I.

Authors:  A Turk; B J Iskandar; V Haughton; D Consigny
Journal:  AJNR Am J Neuroradiol       Date:  2006-02       Impact factor: 3.825

2.  Changes in the intraaneurysmal pressure due to HydroCoil embolization.

Authors:  Gádor Cantón; David I Levy; Juan C Lasheras
Journal:  AJNR Am J Neuroradiol       Date:  2005-04       Impact factor: 3.825

3.  Effect of Guglielmi detachable coils on intraaneurysmal flow: experimental study in canines.

Authors:  Angelika Sorteberg; Wilhelm Sorteberg; Alan Rappe; Charles M Strother
Journal:  AJNR Am J Neuroradiol       Date:  2002-02       Impact factor: 3.825

4.  The Underlying Mechanisms of Endovascular Exclusion of Intracranial Aneurysms by Coils. How Important is Electrothrombosis?

Authors:  H Henkes; S Brew; E Miloslavski; S Fischer; I Tavrovski; D Kühne
Journal:  Interv Neuroradiol       Date:  2004-10-22       Impact factor: 1.610

5.  Development and validation of models for the investigation of blood clotting in idealized stenoses and cerebral aneurysms.

Authors:  Andrew Narracott; Stephen Smith; Patricia Lawford; Hao Liu; Ryutaro Himeno; Iain Wilkinson; Paul Griffiths; Rodney Hose
Journal:  J Artif Organs       Date:  2005       Impact factor: 1.731

6.  Fibered electrolytically detachable platinum coils used for the endovascular treatment of intracranial aneurysms. Initial experiences and mid-term results in 474 aneurysms.

Authors:  T Liebig; H Henkes; S Fischer; W Weber; E Miloslavski; W Mariushi; S Brew; D Kühne
Journal:  Interv Neuroradiol       Date:  2004-10-22       Impact factor: 1.610

7.  In vitro and in vivo Studies of the Extent of Electrothrombotic Deposition of Blood Elements on the Surface of Electrolytically Detachable Coils.

Authors:  H Henkes; S Brew; S Felber; E Miloslavski; G Mogilevski; I Tavrovski; D Kühne
Journal:  Interv Neuroradiol       Date:  2005-01-05       Impact factor: 1.610

8.  Optical coherence tomography: a new method to assess aneurysm healing.

Authors:  William E Thorell; Michael M Chow; Richard A Prayson; Mark A Shure; Sung W Jeon; David Huang; Emil Zeynalov; Henry H Woo; Peter A Rasmussen; Andrew M Rollins; Thomas J Masaryk
Journal:  J Neurosurg       Date:  2005-02       Impact factor: 5.115

9.  The influence of aneurysm configuration on intra-aneurysmal pressure and flow.

Authors:  A Sorteberg; D Farhoudi
Journal:  Interv Neuroradiol       Date:  2006-12-13       Impact factor: 1.610

10.  Hemodynamic versus hydrodynamic effects of Guglielmi detachable coils on intra-aneurysmal pressure and flow at varying pulse rate and systemic pressure.

Authors:  Angelika Sorteberg; Wilhelm Sorteberg; Beverly D L Aagaard; Alan Rappe; Charles M Strother
Journal:  AJNR Am J Neuroradiol       Date:  2004 Jun-Jul       Impact factor: 3.825

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.