Intraoperative three-dimensional imaging with a motorized mobile C-arm (SIREMOBIL ISO-C-3D) in foot and ankle trauma care: a preliminary report.

Martinus Richter1, Jens Geerling, Stefan Zech, Thomas Goesling, Christian Krettek.   

Abstract

OBJECTIVE: The aim of the study was to assess the feasibility and benefit of the intraoperative use of a mobile C-arm with 3-dimensional imaging (ISO-C-3D).
DESIGN: Prospective consecutive clinical study.
SETTING: University hospital, level I trauma center.
METHODS: The ISO-C-3D was used for intraoperative visualization in foot and ankle trauma care. Conventional C-arms were used to judge the reduction and implant position before the ISO-C-3D was used. Time spent, changes resulting from use of the ISO-C-3D, and surgeons' ratings (visual analogue scale, 0-10 points) were recorded. PATIENTS: Between January 1, 2003 and March 15, 2004, the ISO-C-3D was used in 62 cases (factures: pilon, n = 1; Weber-C ankles, n = 7; isolated dorsal Volkmann, n = 1; talus, n = 3; calcaneus, n = 20; navicular, n = 1; cuboid, n = 1; Lisfranc fracture-dislocation, n = 6; hindfoot arthrodesis with or without correction, n = 12).
RESULTS: On average, the operation was interrupted for 440 seconds (range 330-700); 120 seconds, on average, for the ISO-C-3D scan and 210 seconds, on average, for evaluation of the images by the surgeon. In 39% of the cases (24 of 62), the reduction and/or implant position was corrected during the same procedure after the ISO-C-3D scan. The ratings of the 8 surgeons who used the ISO-C-3D were 9.2(5.2-10) for feasibility, 9.5 (6.1-10) for accuracy, and 8.2 (4.5-10) for clinical benefit.
CONCLUSION: Intraoperative 3-dimensional visualization with the ISO-C-3D can provide useful information in foot and ankle trauma care that cannot be obtained from plain films or conventional C-arms. During the same procedure, after conventional C-arm scans judged the positioning to be correct and an ISO-C-3D scan was done, the reduction and/or implant position was corrected in 39% of the cases in this study, although not unnecessarily prolonging the operation. The ISO-C-3D appears to be most helpful in procedures with a closed reduction and internal fixation, and/or when axial reformations provide information that is not possible to obtain with a conventional C-arm and/or direct visualization during open reduction and internal fixation.

Entities:  

Mesh:

Year:  2005        PMID: 15795575     DOI: 10.1097/01.bot.0000151822.10254.db

Source DB:  PubMed          Journal:  J Orthop Trauma        ISSN: 0890-5339            Impact factor:   2.512


  31 in total

1.  Radiation exposure during 3D fluoroscopy of the knee: an experimental study.

Authors:  T Tharmviboonsri; K Riansuwan; A Nitising; B Mahaisavariya
Journal:  Eur J Trauma Emerg Surg       Date:  2011-11-19       Impact factor: 3.693

2.  Calibration of mobile-gantry computed tomography for surgical navigation.

Authors:  Anna Belkova; David R Pichora; Randy E Ellis
Journal:  Int J Comput Assist Radiol Surg       Date:  2015-10-08       Impact factor: 2.924

3.  [Cost-benefit analysis of intraoperative 3D imaging].

Authors:  T Hüfner; T Stübig; T Gösling; D Kendoff; J Geerling; C Krettek
Journal:  Unfallchirurg       Date:  2007-01       Impact factor: 1.000

4.  [Integrated navigation. Preclinical evaluation and initial clinical experience].

Authors:  U Stöckle; A Schäffler; B König; N P Haas
Journal:  Unfallchirurg       Date:  2006-11       Impact factor: 1.000

5.  Computer aided surgery in foot and ankle: applications and perspectives.

Authors:  Martinus Richter
Journal:  Int Orthop       Date:  2013-05-25       Impact factor: 3.075

6.  [Comparison of C-arm CT and standard imaging in osteosyntheses of fractured distal radius specimens].

Authors:  S Wirth; E Euler; D Kotsianos-Hermle; M Treitl; U Linsenmaier; K-J Pfeifer; M Reiser; W Mutschler
Journal:  Unfallchirurg       Date:  2007-01       Impact factor: 1.000

7.  [Intraoperative pedography].

Authors:  Martinus Richter; Stefan Zech
Journal:  Oper Orthop Traumatol       Date:  2010-03       Impact factor: 1.154

8.  Volume slicing of cone-beam computed tomography images for navigation of percutaneous scaphoid fixation.

Authors:  Erin J Smith; Hisham A Al-Sanawi; Braden Gammon; Paul J St John; David R Pichora; Randy E Ellis
Journal:  Int J Comput Assist Radiol Surg       Date:  2011-06-25       Impact factor: 2.924

Review 9.  [Injuries of the midfoot].

Authors:  T Mittlmeier; M Beck
Journal:  Chirurg       Date:  2011-02       Impact factor: 0.955

10.  [Intraoperative 3D C-arm imaging. State of the art].

Authors:  J von Recum; K Wendl; B Vock; P A Grützner; J Franke
Journal:  Unfallchirurg       Date:  2012-03       Impact factor: 1.000

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.