Radiation exposure during 3D fluoroscopy of the knee: an experimental study.

T Tharmviboonsri1, K Riansuwan2, A Nitising3, B Mahaisavariya1.   

Abstract

PURPOSE: To measure the radiation exposure at the knee and surrounding area during the use of 3D fluoroscopy.
METHODS: An experimental study was conducted by using a human cadaveric knee as a focus point for the 3D fluoroscope. An isocentric C-arm fluoroscope machine was applied on the lateral side of the knee. The radiation dosage at the focus point and surrounding area was measured. The mean radiation exposure in each location was compared between low- and high-resolution scanning.
RESULTS: The mean radiation sustained at the focus point was 44.0 ± 5.6 μSv and 20.0 ± 1.0 μSv in high- and low-resolution scanning, respectively. Radiation exposure on the opposite side of the C-arm machine was found to be higher than that on the other locations with the same distance from the focus point. In low-resolution scanning, radiation could not be detected beyond 75 cm from the focus point at the proximal, distal and same side of the machine. Radiation could be measured at a distance of up to 1.25 m on the opposite side of the machine. In high-resolution scanning, radiation could be measured at a distance of up to 1 m at the proximal, distal and same side of the C-arm, but up to 1.5 m on the opposite side.
CONCLUSION: Radiation exposure during 3D fluoroscopy of the knee decreases with increasing distance from the focus point. A higher number of scans in the high-resolution mode causes greater radiation exposure. In isocentric 3D fluoroscopy of the knee, a safe zone is located at least 1.5 m away from the focus point.

Entities:  

Keywords:  3D fluoroscopy; C-arm; Exposure; Knee; Radiation

Year:  2011        PMID: 26815963     DOI: 10.1007/s00068-011-0165-7

Source DB:  PubMed          Journal:  Eur J Trauma Emerg Surg        ISSN: 1863-9933            Impact factor:   3.693


  13 in total

1.  Occupational exposure from common fluoroscopic projections used in orthopaedic surgery.

Authors:  Nicholas Theocharopoulos; Kostas Perisinakis; John Damilakis; George Papadokostakis; Alexander Hadjipavlou; Nicholas Gourtsoyiannis
Journal:  J Bone Joint Surg Am       Date:  2003-09       Impact factor: 5.284

2.  Intraoperative three-dimensional imaging with a motorized mobile C-arm (SIREMOBIL ISO-C-3D) in foot and ankle trauma care: a preliminary report.

Authors:  Martinus Richter; Jens Geerling; Stefan Zech; Thomas Goesling; Christian Krettek
Journal:  J Orthop Trauma       Date:  2005-04       Impact factor: 2.512

3.  Percutaneous placement of posterior cervical screws using three-dimensional fluoroscopy.

Authors:  Langston T Holly; Kevin T Foley
Journal:  Spine (Phila Pa 1976)       Date:  2006-03-01       Impact factor: 3.468

4.  [The clinical use of the ISO-C(3D) imaging system in calcaneus fracture surgery].

Authors:  A Rübberdt; R Feil; D Stengel; N Spranger; S Mutze; M Wich; A Ekkernkamp
Journal:  Unfallchirurg       Date:  2006-02       Impact factor: 1.000

5.  Radiation exposure to the orthopaedic surgical team during fluoroscopy: "how far away is far enough?".

Authors:  C T Mehlman; T G DiPasquale
Journal:  J Orthop Trauma       Date:  1997-08       Impact factor: 2.512

6.  Radiation exposure and associated risks to operating-room personnel during use of fluoroscopic guidance for selected orthopaedic surgical procedures.

Authors:  M E Miller; M L Davis; C R MacClean; J G Davis; B L Smith; J R Humphries
Journal:  J Bone Joint Surg Am       Date:  1983-01       Impact factor: 5.284

7.  Irradiation of the surgeon during pinning of femoral fractures.

Authors:  A A Giachino; M Cheng
Journal:  J Bone Joint Surg Br       Date:  1980-05

8.  Accuracy analysis of Iso-C3D versus fluoroscopy-based navigated retrograde drilling of osteochondral lesions: a pilot study.

Authors:  M Citak; D Kendoff; M Kfuri; A Pearle; C Krettek; T Hüfner
Journal:  J Bone Joint Surg Br       Date:  2007-03

9.  Intraoperative 3-dimensional reconstructed multiplanar fluoroscopic imaging for immediate evaluation of spinal decompression.

Authors:  Vikas V Patel; Anthony Dwyer; Susan Estes; Evalina Burger
Journal:  J Spinal Disord Tech       Date:  2008-05

10.  Odontoid screw placement using Isocentric 3-dimensional C-arm fluoroscopy.

Authors:  Lori E Summers; Joshua G Kouri; Mu Yang; R Patrick Jacob
Journal:  J Spinal Disord Tech       Date:  2008-02
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.