Literature DB >> 17177043

[Cost-benefit analysis of intraoperative 3D imaging].

T Hüfner1, T Stübig, T Gösling, D Kendoff, J Geerling, C Krettek.   

Abstract

BACKGROUND: With intraoperative 3D imaging, inevitable corrections may be done already during the operation, and a second procedure can be avoided. The purpose of this study was to perform a cost-benefit analysis during the first year of intraoperative 3D application in order to provide a cost transparency for the surgeon within the current DRG system.
METHODS: On the basis of internal data and the literature, the annual operating costs of the ISO-C(3D) were calculated at 27,940 euros (purchase price, depreciation, maintenance, repair), the costs of an average revision as a secondary operation at 2,385 euros (costs avoided with the ISO-C(3D), Siremobil, Siemens, Erlangen, Germany), and the dynamic costs of an intraoperative 3D scan were averaged to 131.08 euros (draping, additional time, personnel).
RESULTS: In the year 2003 intraoperative 3D scanning was done in 126 patients, and intraoperative revision was performed in 24 (19%) due to the additional intraoperative 3D information provided by the Siremobil. In 11 (8.7%) patients the implant position was corrected and in 13 (10.3%) patients the reduction was improved. Taking only fixed costs into consideration, 29,311.52 euros could be saved, and when fixed and dynamic costs are taken into account 12,795.44 euros could be saved. Since the parameters for each hospital are different, the following formula for an individual computation is suitable. For the calculation of the cost the following mathematical relationship results: (annual fixed costs) + (costs per scan x number of cases) - (revision costs x revision rate [p]) x number of cases [N]) = 0. DISCUSSION: Although the costs of an ISO-C(3D) are considerably high, an economic benefit can also accrue with frequent application and high rates of avoided revision. However, if the rate of avoided revision adds up to only 5%, a substantial deficit may result.

Entities:  

Mesh:

Year:  2007        PMID: 17177043     DOI: 10.1007/s00113-006-1202-6

Source DB:  PubMed          Journal:  Unfallchirurg        ISSN: 0177-5537            Impact factor:   1.000


  14 in total

1.  A whole-body registration-free navigation system for image-guided surgery and interventional radiology.

Authors:  A L Jacob; P Messmer; A Kaim; N Suhm; P Regazzoni; B Baumann
Journal:  Invest Radiol       Date:  2000-05       Impact factor: 6.016

2.  Use of intraoperative isocentric C-arm 3D fluoroscopy for sextant percutaneous pedicle screw placement: case report and review of the literature.

Authors:  Frank L Acosta; Timothy L Thompson; Stacey Campbell; Philip R Weinstein; Christopher P Ames
Journal:  Spine J       Date:  2005 May-Jun       Impact factor: 4.166

3.  [Intraoperative three-dimensional imaging with an isocentric mobile C-arm at the wrist].

Authors:  R Meier; M Kfuri; J Geerling; T Hüfner; H Krimmer; C Krettek
Journal:  Handchir Mikrochir Plast Chir       Date:  2005-08       Impact factor: 1.018

4.  Intraoperative three-dimensional imaging with a motorized mobile C-arm (SIREMOBIL ISO-C-3D) in foot and ankle trauma care: a preliminary report.

Authors:  Martinus Richter; Jens Geerling; Stefan Zech; Thomas Goesling; Christian Krettek
Journal:  J Orthop Trauma       Date:  2005-04       Impact factor: 2.512

5.  Computer-assisted screw insertion into the first sacral vertebra using a three-dimensional image intensifier: results of a controlled experimental investigation.

Authors:  D Briem; W Linhart; W Lehmann; P G Begemann; G Adam; U Schumacher; D M Cullinane; J M Rueger; J Windolf
Journal:  Eur Spine J       Date:  2005-07-12       Impact factor: 3.134

6.  Navigated Iso-C3D-based percutaneous osteoid osteoma resection: a preliminary clinical report.

Authors:  Daniel Kendoff; Tobias Hüfner; Musa Citak; Jens Geerling; Eckhard Mössinger; Leonhard Bastian; Christian Krettek
Journal:  Comput Aided Surg       Date:  2005-05

7.  [Comparative study of the quality of C-arm based 3D imaging of the talus].

Authors:  E Euler; S Wirth; U Linsenmaier; W Mutschler; K J Pfeifer; A Hebecker
Journal:  Unfallchirurg       Date:  2001-09       Impact factor: 1.000

8.  [3-D imaging with a mobile surgical image enhancement equipment (ISO-C-3D). Initial examples of fracture diagnosis of peripheral joints in comparison with spiral CT and conventional radiography].

Authors:  D Kotsianos; C Rock; E Euler; S Wirth; U Linsenmaier; R Brandl; W Mutschler; K J Pfeifer
Journal:  Unfallchirurg       Date:  2001-09       Impact factor: 1.000

9.  Intraoperative 3D imaging of the facial skeleton using the SIREMOBIL Iso-C3D.

Authors:  M Heiland; R Schmelzle; A Hebecker; D Schulze
Journal:  Dentomaxillofac Radiol       Date:  2004-03       Impact factor: 2.419

10.  [Cost effectiveness of intraoperative three-dimensional imaging with a mobile surgical C-arm].

Authors:  M Bischoff; A Hebecker; E Hartwig; F Gebhard
Journal:  Unfallchirurg       Date:  2004-08       Impact factor: 1.000

View more
  18 in total

1.  [Handling modern imaging procedures in a high-tech operating room].

Authors:  T Hüfner; M Citak; J Imrecke; C Krettek; T Stübig
Journal:  Unfallchirurg       Date:  2012-03       Impact factor: 1.000

2.  3D-based navigation in posterior stabilisations of the cervical and thoracic spine: problems and benefits. Results of 451 screws.

Authors:  J-S Jarvers; S Katscher; A Franck; S Glasmacher; C Schmidt; T Blattert; C Josten
Journal:  Eur J Trauma Emerg Surg       Date:  2011-04-01       Impact factor: 3.693

3.  Computer-Assisted Orthopedic and Trauma Surgery.

Authors:  Timo Stübig; Henning Windhagen; Christian Krettek; Max Ettinger
Journal:  Dtsch Arztebl Int       Date:  2020-11-20       Impact factor: 5.594

4.  [Hybrid operating room : Application in trauma surgery and orthopaedics].

Authors:  Frederic Braun; Simon Peh; Matthias Weuster; Tim Klueter; Andreas Seekamp; Sebastian Lippross
Journal:  Unfallchirurg       Date:  2020-02       Impact factor: 1.000

5.  [Intraoperative 3D C-arm imaging. State of the art].

Authors:  J von Recum; K Wendl; B Vock; P A Grützner; J Franke
Journal:  Unfallchirurg       Date:  2012-03       Impact factor: 1.000

Review 6.  [The hybrid operating room. Home of high-end intraoperative imaging].

Authors:  F Gebhard; C Riepl; P Richter; A Liebold; H Gorki; R Wirtz; R König; F Wilde; A Schramm; M Kraus
Journal:  Unfallchirurg       Date:  2012-02       Impact factor: 1.000

7.  Preclinical usability study of multiple augmented reality concepts for K-wire placement.

Authors:  Marius Fischer; Bernhard Fuerst; Sing Chun Lee; Javad Fotouhi; Severine Habert; Simon Weidert; Ekkehard Euler; Greg Osgood; Nassir Navab
Journal:  Int J Comput Assist Radiol Surg       Date:  2016-03-19       Impact factor: 2.924

Review 8.  [Intraoperative 3D imaging in spinal surgery].

Authors:  O Gonschorek; S Hauck; V Bühren
Journal:  Unfallchirurg       Date:  2016-10       Impact factor: 1.000

9.  [Comparison of early total care (ETC) and damage control orthopedics (DCO) in the treatment of multiple trauma with femoral shaft fractures: benefit and costs].

Authors:  T Stübig; P Mommsen; C Krettek; C Probst; M Frink; C Zeckey; H Andruszkow; F Hildebrand
Journal:  Unfallchirurg       Date:  2010-11       Impact factor: 1.000

10.  [Multiplanar reconstruction with mobile 3D image intensifier. Surgical treatment of proximal humerus fractures].

Authors:  P Hepp; J Theopold; J-S Jarvers; B Marquaß; N von Dercks; C Josten
Journal:  Unfallchirurg       Date:  2014-05       Impact factor: 1.000

View more