Discrepancy between Doppler and catheter measurements of pressure gradients across small-size prosthetic valve.

Takuya Yamashita1, Yukinori Moriyama, Naoyuki Sata, Naokazu Hamada, Takashi Horinouchi, Shigeru Amitani, Kenkichi Miyahara, Kentaro Setoyama, Kazuhiro Misumi, Hiroshi Sakamoto.   

Abstract

OBJECTIVES: The exact role of pressure gradient across the prosthetic valve estimated from Doppler flow velocity remains controversial. This in-vivo study was designed to assess the actual discrepancy between Doppler and catheter measurements of the pressure gradients for small bileaflet prosthetic valves in the aortic position.
METHODS: Bileaflet prosthetic valves (19 mm-ATS) were implanted into the aortic position in pigs, and pressure gradients across the valves were examined by volume loading under right heart bypass. The pressure gradient obtained by catheter was defined as the conventional peak-to-peak gradient between the left ventricle and aorta. The peak Doppler gradients were calculated from the maximal instantaneous Doppler velocity with the ultrasound probe positioned on the diaphragm at the level of the cardiac apex.
RESULTS: There were strong correlations between pressure gradients and cardiac output. The Doppler gradient was constantly higher than the catheter values, and the resultant discrepancy between Doppler and catheter measurements was directly dependent on cardiac output (y=9.9x+0.6, r2=0.55). For cardiac output > or = 5.0 L/min, the difference between Doppler and catheter gradients reached 40 mmHg, and maximum differences of up to 80 mmHg were observed.
CONCLUSIONS: In view of the presence of striking overestimation of catheter gradient by Doppler measurement, Doppler ultrasound should be used cautiously to assess small-size bileaflet prosthetic valve function with consideration of the patient's hemodynamic state.

Entities:  

Mesh:

Year:  2005        PMID: 15782565     DOI: 10.1007/s11748-005-0002-1

Source DB:  PubMed          Journal:  Jpn J Thorac Cardiovasc Surg        ISSN: 1344-4964


  14 in total

1.  Comparative study of the hydrodynamic function of six size 19 mm bileaflet heart valves.

Authors:  J Fisher
Journal:  Eur J Cardiothorac Surg       Date:  1995       Impact factor: 4.191

2.  Importance of pressure recovery for the assessment of aortic stenosis by Doppler ultrasound. Role of aortic size, aortic valve area, and direction of the stenotic jet in vitro.

Authors:  J Niederberger; H Schima; G Maurer; H Baumgartner
Journal:  Circulation       Date:  1996-10-15       Impact factor: 29.690

3.  Prediction of the severity of aortic stenosis by Doppler aortic valve area determination: prospective Doppler-catheterization correlation in 100 patients.

Authors:  J K Oh; C P Taliercio; D R Holmes; G S Reeder; K R Bailey; J B Seward; A J Tajik
Journal:  J Am Coll Cardiol       Date:  1988-06       Impact factor: 24.094

4.  Pressure recovery in bileaflet heart valve prostheses. Localized high velocities and gradients in central and side orifices with implications for Doppler-catheter gradient relation in aortic and mitral position.

Authors:  P M Vandervoort; N L Greenberg; K A Powell; D M Cosgrove; J D Thomas
Journal:  Circulation       Date:  1995-12-15       Impact factor: 29.690

5.  Clinical experience with the first 100 ATS heart valve implants.

Authors:  G Van Nooten; F Caes; K François; L Missault; Y Van Belleghem
Journal:  Cardiovasc Surg       Date:  1996-06

6.  The initial experience with the ATS Medical mechanical cardiac valve prosthesis.

Authors:  Robert W Emery; Guido J Van Nooten; Peter J Tesar
Journal:  Ann Thorac Surg       Date:  2003-02       Impact factor: 4.330

7.  Aortic stenosis in adults. Non-invasive estimation of pressure differences by continuous wave Doppler echocardiography.

Authors:  L Hegrenaes; L Hatle
Journal:  Br Heart J       Date:  1985-10

8.  Discrepancies between catheter and Doppler estimates of valve effective orifice area can be predicted from the pressure recovery phenomenon: practical implications with regard to quantification of aortic stenosis severity.

Authors:  Damien Garcia; Jean G Dumesnil; Louis-Gilles Durand; Lyes Kadem; Philippe Pibarot
Journal:  J Am Coll Cardiol       Date:  2003-02-05       Impact factor: 24.094

9.  Prospective study of hemodynamic performances of standard ATS and AP-ATS valves.

Authors:  Louis M Labrousse; Emmanuel Choukroun; Donato Serena; Marc-Alain Billes; Francesco Madonna; Claude Deville
Journal:  J Heart Valve Dis       Date:  2003-05

10.  Long-term clinical outcome with small size Standard St Jude Medical valves implanted in the aortic position.

Authors:  K V Arom; I F Goldenberg; R W Emery
Journal:  J Heart Valve Dis       Date:  1994-09
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.