Literature DB >> 15758672

A biomechanical comparison of a dorsal 3.5-mm T-plate and a volar fixed-angle plate in a model of dorsally unstable distal radius fractures.

Frank A Liporace1, Salil Gupta, Gerard K Jeong, Michael Stracher, Fredrick Kummer, Kenneth A Egol, Kenneth J Koval.   

Abstract

OBJECTIVES: To compare the biomechanical stability of internal fixation of extra-articular, dorsally unstable distal radius fractures fixed by 1 of 2 methods, either a standard dorsal nonlocked T-plate or a volar locked fixed-angle plate.
DESIGN: Biomechanical cadaveric study.
SETTING: Biomechanical testing laboratory. INTERVENTION: In 6 matched pairs of fresh-frozen cadaveric specimens, a simulated unstable extra-articular distal radius fracture was created. The fractures were stabilized with either a dorsal 3.5-mm stainless steel T-plate or a titanium locked volar fixed-angle plate. Specimens were axially loaded at 5 points (centrally, volarly, dorsally, radially, and ulnarly) and then cyclically loaded for 5000 cycles with an 80 N central load. Postcyclical loading, specimens were once again axially loaded at the 5 points. MAIN OUTCOME MEASURES: Initial fixation stiffness and stiffness after midaxial cyclical loading was compared at the 5 points.
RESULTS: With the volar locked fixed-angle plate, fixation was significantly stiffer than with the dorsal nonlocked T-plate for ulnar and volar loading in single-cycle testing. After cyclic loading, the locked volar fixed-angle plate maintained more of its initial stiffness than the dorsal nonlocked T-plate. The dorsal 3.5-mm stainless steel T-plate's stiffness when dorsally loaded significantly decreased after cyclical loading.
CONCLUSIONS: The volar locked fixed-angle plate maintained a greater percentage of its initial stiffness after cyclic loading compared to the dorsal nonlocked plate. Also, the volar locked plate was stiffer than the dorsal nonlocked plate for all loading configurations tested except when subjected to a dorsally applied eccentric load.

Entities:  

Mesh:

Year:  2005        PMID: 15758672     DOI: 10.1097/00005131-200503000-00006

Source DB:  PubMed          Journal:  J Orthop Trauma        ISSN: 0890-5339            Impact factor:   2.512


  11 in total

1.  Radiographic evaluation of dorsal screw penetration after volar fixed-angle plating of the distal radius: a cadaveric study.

Authors:  Steven D Maschke; Peter J Evans; David Schub; Richard Drake; Jeffrey N Lawton
Journal:  Hand (N Y)       Date:  2007-04-26

2.  Multidirectional volar fixed-angle plating using cancellous locking screws for distal radius fractures--evaluation of three screw configurations in an extra-articular fracture model.

Authors:  Patrick Weninger; Enrico Dall'Ara; Herwig Drobetz; Wolfgang Nemec; Markus Figl; Heinz Redl; Harald Hertz; Philippe Zysset
Journal:  Wien Klin Wochenschr       Date:  2011-01       Impact factor: 1.704

3.  Biomechanical comparison of different volar fracture fixation plates for distal radius fractures.

Authors:  Kareem Sobky; Todd Baldini; Kenneth Thomas; Joel Bach; Allison Williams; Jennifer Moriatis Wolf
Journal:  Hand (N Y)       Date:  2007-09-07

4.  Radiographic outcomes of volar locked plating for distal radius fractures.

Authors:  Megan E Mignemi; Ian R Byram; Carmen C Wolfe; Kang-Hsien Fan; Elizabeth A Koehler; John J Block; Martin I Jordanov; Jeffry T Watson; Douglas R Weikert; Donald H Lee
Journal:  J Hand Surg Am       Date:  2012-12-04       Impact factor: 2.230

5.  Promontory of radius: a new anatomical description on the distal radius.

Authors:  G Windisch; H Clement; K Tanzer; G Feigl; W Grechenig; F Anderhuber; W Pichler
Journal:  Surg Radiol Anat       Date:  2007-10-11       Impact factor: 1.246

6.  The role of locking technology in the upper extremity.

Authors:  Siddharth B Joglekar; Asif M Ilyas
Journal:  J Hand Microsurg       Date:  2010-01-08

7.  Biomechanical stability of four fixation constructs for distal radius fractures.

Authors:  John T Capo; Tosca Kinchelow; Kenneth Brooks; Virak Tan; Michaele Manigrasso; Kristin Francisco
Journal:  Hand (N Y)       Date:  2009-02-05

8.  Comparison of the effect on bone healing process of different implants used in minimally invasive plate osteosynthesis: limited contact dynamic compression plate versus locking compression plate.

Authors:  Zichao Xue; Haitao Xu; Haoliang Ding; Hui Qin; Zhiquan An
Journal:  Sci Rep       Date:  2016-11-25       Impact factor: 4.379

9.  Retrospective analysis of extra-articular distal humerus shaft fractures treated with the use of pre-contoured lateral column metaphyseal LCP by triceps-sparing posterolateral approach.

Authors:  Yatinder Kharbanda; Yashwant Singh Tanwar; Vishal Srivastava; Vikas Birla; Ashok Rajput; Ramsagar Pandit
Journal:  Strategies Trauma Limb Reconstr       Date:  2016-11-03

10.  Assessment of a novel biomechanical fracture model for distal radius fractures.

Authors:  Sebastian Felix Baumbach; Enrico Dall'Ara; Patrick Weninger; Anna Antoni; Hannes Traxler; Martin Dörr; Philippe K Zysset
Journal:  BMC Musculoskelet Disord       Date:  2012-12-18       Impact factor: 2.362

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.