The diagnosis of breast implant rupture: MRI findings compared with findings at explantation.

Lisbet R Hölmich1, Ilse Vejborg, Carsten Conrad, Susanne Sletting, Joseph K McLaughlin.   

Abstract

STUDY
OBJECTIVE: The aim of this study was to evaluate the accuracy of Magnetic Resonance Imaging (MRI) as performed according to a strict study protocol in diagnosing rupture of silicone breast implants.
MATERIAL AND METHODS: The study population consisted of 64 women with 118 implants, who had participated in either one or two study MRI examinations, aiming at determining the prevalence and incidence of silent implant rupture, respectively, and who subsequently underwent explantation. Implant rupture status was determined by four independent readers and a consensus diagnosis of either rupture (intracapsular or extracapsular), possible rupture or intact implant was then obtained. Strict predetermined rupture criteria were applied as described in this report and findings at surgery were abstracted in a standardised manner and results compared.
RESULTS: At MRI, 66 implants were diagnosed as ruptured, nine as possibly ruptured and 43 as intact. Among the ruptured implants, 27 were categorized as extracapsular. At surgery, on average 297 days after the MRI, 65 of the 66 rupture diagnoses were confirmed, as were 20 of the cases with extracapsular silicone. Eight of the nine possibly ruptured implants were in fact ruptured at surgery. Thirty-four of the 43 intact implants were described as intact at surgery. When categorising possible ruptures as ruptures, there were one false positive and nine false negative rupture diagnoses at MRI yielding an accuracy of 92%, a sensitivity of 89%, and a specificity of 97%. Correspondingly, the predictive value of a positive MRI examination was 99% and the predictive value of a negative MRI examination was 79%.
CONCLUSIONS: We conclude that MRI is highly accurate for identification of silicone breast implant rupture, with a high sensitivity and specificity when evaluation of images are based on presence of well-defined rupture criteria.

Entities:  

Mesh:

Substances:

Year:  2005        PMID: 15664285     DOI: 10.1016/j.ejrad.2004.03.012

Source DB:  PubMed          Journal:  Eur J Radiol        ISSN: 0720-048X            Impact factor:   3.528


  13 in total

1.  Study of breast implant rupture: MRI versus surgical findings.

Authors:  A Vestito; F F Mangieri; A Ancona; C Minervini; V Perchinunno; S Rinaldi
Journal:  Radiol Med       Date:  2012-03-19       Impact factor: 3.469

2.  Role of US in evaluating breast implant integrity.

Authors:  Michele Telegrafo; Marco Moschetta
Journal:  J Ultrasound       Date:  2015-06-23

3.  MRI screening for silicone breast implant rupture: accuracy, inter- and intraobserver variability using explantation results as reference standard.

Authors:  M C Maijers; F B Niessen; J F H Veldhuizen; M J P F Ritt; R A Manoliu
Journal:  Eur Radiol       Date:  2014-03-06       Impact factor: 5.315

4.  Invited Discussion on: Breast Implants Follow-Up-Results of a Cross-Sectional Study on Patients Submitted to MRI Breast Examinations.

Authors:  Tim Papadopoulos
Journal:  Aesthetic Plast Surg       Date:  2020-11-19       Impact factor: 2.326

Review 5.  The effect of study design biases on the diagnostic accuracy of magnetic resonance imaging for detecting silicone breast implant ruptures: a meta-analysis.

Authors:  Jae W Song; Hyungjin Myra Kim; Lillian T Bellfi; Kevin C Chung
Journal:  Plast Reconstr Surg       Date:  2011-03       Impact factor: 4.730

6.  Punctual Breast Implant Rupture following Lipofilling: Only a Myth?

Authors:  Andres Rivera; Carlota González-Pozega; Gorka Ibarra; Borja Fernandez-Ibarburu; Ángela García-Ruano; Alfonso Vallejo-Valero
Journal:  Breast Care (Basel)       Date:  2020-10-28       Impact factor: 2.268

7.  Dual-energy CT for the evaluation of silicone breast implants.

Authors:  Thorsten R C Johnson; Isabelle Himsl; Karin Hellerhoff; Doris Mayr; Dorothea Rjosk-Dendorfer; Nina Ditsch; Bernhard Krauss; Klaus Friese; Maximilian F Reiser; Miriam S Lenhard
Journal:  Eur Radiol       Date:  2012-10-13       Impact factor: 5.315

8.  Imaging of breast implants-a pictorial review.

Authors:  Sergi Juanpere; Elsa Perez; Oscar Huc; Naiara Motos; Josep Pont; Salvador Pedraza
Journal:  Insights Imaging       Date:  2011-08-07

9.  Removing silicone artifacts in diffusion-weighted breast MRI by means of shift-resolved spatiotemporally encoding.

Authors:  Eddy Solomon; Noam Nissan; Rita Schmidt; Edna Furman-Haran; Uriel Ben-Aharon; Lucio Frydman
Journal:  Magn Reson Med       Date:  2015-06-22       Impact factor: 4.668

10.  Breast MRI: guidelines from the European Society of Breast Imaging.

Authors:  R M Mann; C K Kuhl; K Kinkel; C Boetes
Journal:  Eur Radiol       Date:  2008-04-04       Impact factor: 5.315

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.