Comparison of the Tono-Pen and Goldmann tonometer for measuring intraocular pressure in patients with glaucoma.

Gregory S Horowitz1, Julie Byles, John Lee, Catherine D'Este.   

Abstract

AIM: To estimate agreement between measurement of intraocular pressure (IOP) by the Tono-Pen and by the Goldmann tonometer. The hypothesis tested was that the mean difference in measurements would be <-2 mmHg.
METHODS: A prospective, single-centre, comparison study was carried out. The Tono-Pen was compared to the Goldmann tonometer in 138 patients systematically selected from two glaucoma clinics and among selected patients known to have high pressures (n = 22). For both groups the investigators were masked to the value of the Goldmann readings, the instruments were used in random order, and duplicated readings were taken with both tonometers.
RESULTS: In the systematically selected group, for right eyes, the mean difference in measurement between the Tono-Pen and Goldmann tonometer was -0.41 mmHg (SD: 2.59). The 95% limits of agreement were -5.5 to 4.7 mmHg, and the intraclass correlation coefficient (ICC) was 0.90. Results for left eyes were very similar. For patients with high pressures (> or =25 mmHg), the mean difference between instruments (Tono-Pen reading minus Goldmann reading): was -4.2 mmHg (P = 0.0004), SD: 4.6, 95% limits of agreement: -13.2 to 4.8 mmHg and ICC = 0.76. Combining the analysis for both groups, the Tono-Pen significantly underestimated the IOP when the pressure was >20 mmHg.
CONCLUSIONS: The Tono-Pen cannot replace the Goldmann tonometer in the sense that it will give the same readings of IOP. The accuracy of the Tono-Pen is increased, if at least two measurements are taken per eye and then averaged.

Entities:  

Mesh:

Year:  2004        PMID: 15575828     DOI: 10.1111/j.1442-9071.2004.00907.x

Source DB:  PubMed          Journal:  Clin Exp Ophthalmol        ISSN: 1442-6404            Impact factor:   4.207


  11 in total

1.  Agreement between diurnal variations of intraocular pressure by Tono-Pen and Goldmann applanation tonometer in patients on topical anti-glaucoma medication.

Authors:  Shikha Gupta; Gautam Sinha; Reetika Sharma; Bhagabat Nayak; Bharat Patil; Bibhuti Kashyap; Abdul Shameer; Tanuj Dada
Journal:  Int Ophthalmol       Date:  2015-03-28       Impact factor: 2.031

2.  [Goldmann applanation tonometry and dynamic contour tonometry. Comparison of intraocular pressure measurements in the course of product certification according to EN ISO 8612:2001].

Authors:  E M Hoffmann; N Pfeiffer; L Barleon; F H Grus
Journal:  Ophthalmologe       Date:  2006-04       Impact factor: 1.059

3.  Comparison of intraocular tonometry using three different non-invasive tonometers in children.

Authors:  Julia Lasseck; Thomas Jehle; Nicolas Feltgen; Wolf Alexander Lagrèze
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2008-06-14       Impact factor: 3.117

4.  Choroidal thickness changes after dynamic exercise as measured by spectral-domain optical coherence tomography.

Authors:  Nihat Sayin; Necip Kara; Gokhan Pekel; Hasan Altinkaynak
Journal:  Indian J Ophthalmol       Date:  2015-05       Impact factor: 1.848

5.  Evaluation of the new digital goldmann applanation tonometer for measuring intraocular pressure.

Authors:  Yuta Sakaue; Jun Ueda; Masaaki Seki; Takayuki Tanaka; Tetsuya Togano; Takaiko Yoshino; Takeo Fukuchi
Journal:  J Ophthalmol       Date:  2014-07-10       Impact factor: 1.909

6.  Detecting IOP Fluctuations in Glaucoma Patients.

Authors:  Brenda Nuyen; Kaweh Mansouri
Journal:  Open Ophthalmol J       Date:  2016-02-29

7.  Comparison of Intraocular Pressure Measurements in Healthy Pediatric Patients using Three Types of Tonometers.

Authors:  Muhsin Eraslan; Eren Çerman; Sena Sümmen
Journal:  Turk J Ophthalmol       Date:  2017-01-17

8.  Strategies to estimate the characteristics of 24-hour IOP curves of treated glaucoma patients during office hours.

Authors:  Leonardo Colombo; Paolo Fogagnolo; Giovanni Montesano; Stefano De Cillà; Nicola Orzalesi; Luca Rossetti
Journal:  BMC Ophthalmol       Date:  2016-01-27       Impact factor: 2.209

Review 9.  Ageing and ocular surface immunity.

Authors:  Alireza Mashaghi; Jiaxu Hong; Sunil K Chauhan; Reza Dana
Journal:  Br J Ophthalmol       Date:  2016-07-04       Impact factor: 4.638

10.  Comparisons of the Tono-Pen® and Goldmann Applanation Tonometer in the Measurement of Intraocular Pressure of Primary Open Angle Glaucoma Patients in a Hospital Population in Southwest Nigeria.

Authors:  Chizoba Onochie; Obiekwe Okoye; Adunola Ogunro; Tade Aribaba; Kunle Hassan; Adeola Onakoya
Journal:  Med Princ Pract       Date:  2016-08-09       Impact factor: 1.927

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.