Comparison between two hand-held autorefractors: the Sure-Sight and the Retinomax.

M Cordonnier1, V De Maertelaer.   

Abstract

PURPOSE: To compare the results of manifest refraction obtained with two different hand-held autorefractors (Sure-Sight, Welch Allyn Co. and Retinomax 1, Nikon Inc.) and with the Topcon RMA 6000 on-table autorefractor in order to estimate any potential bias between these refractometers and to compare the diagnostic performances of these two hand-held autorefractors as screening devices.
METHODS: Ninety-eight children were refracted under manifest conditions with the three above-mentioned refractometers and under cycloplegic conditions with the Topcon on-table autorefractor, or by means of retinoscopy. The agreement between the manifest measurements obtained with the three different autorefractors was studied using the method of Bland and Altman. The validity of several thresholds of manifest refractive anomalies as measured with the Sure-Sight and with the Retinomax was estimated by receiver operating characteristic (ROC) curves using cycloplegic measures as reference. results There is a spherical positive bias of 1 D between the Sure-Sight and the Retinomax and better agreement between the Topcon and the Retinomax. The surface area indexes of the ROC curves and the diagnostic performances in term of sensitivity and specificity are better with the Retinomax in cases of hyperopia, astigmatism and anisometropia. For myopia, the Sure-Sight has better performance.
CONCLUSIONS: The results suggest that either device may be sufficient for assessing refractive errors in children in a screening setting. However, because of a bias between both refractometers in measuring the sphere, distinctive referral criteria must be chosen for the detection of hyperopia and myopia. The diagnostic performance is slightly in favor of the Retinomax.

Entities:  

Mesh:

Year:  2004        PMID: 15545144     DOI: 10.1080/09273970490886594

Source DB:  PubMed          Journal:  Strabismus        ISSN: 0927-3972


  11 in total

1.  Comparison of the Retinomax and Palm-AR Auto-Refractors: a pilot study.

Authors:  Elise Ciner; Ashanti Carter; Gui-Shuang Ying; Maureen Maguire; Marjean Taylor Kulp
Journal:  Optom Vis Sci       Date:  2011-07       Impact factor: 1.973

2.  [Possibilities and limitations of amblyopia screening with auto-refractometers].

Authors:  O Ehrt
Journal:  Ophthalmologe       Date:  2016-04       Impact factor: 1.059

3.  Accuracy of noncycloplegic retinoscopy, retinomax autorefractor, and SureSight vision screener for detecting significant refractive errors.

Authors:  Marjean Taylor Kulp; Gui-Shuang Ying; Jiayan Huang; Maureen Maguire; Graham Quinn; Elise B Ciner; Lynn A Cyert; Deborah A Orel-Bixler; Bruce D Moore
Journal:  Invest Ophthalmol Vis Sci       Date:  2014-03-06       Impact factor: 4.799

4.  Receding and disparity cues aid relaxation of accommodation.

Authors:  Anna M Horwood; Patricia M Riddell
Journal:  Optom Vis Sci       Date:  2009-11       Impact factor: 1.973

5.  Comparison of the Retinomax hand-held autorefractor versus table-top autorefractor and retinoscopy.

Authors:  Ibrahim Tuncer; Mehmet Ozgur Zengin; Eyyup Karahan
Journal:  Int J Ophthalmol       Date:  2014-06-18       Impact factor: 1.779

6.  Accuracy of the Welch Allyn SureSight for measurement of magnitude of astigmatism in 3- to 7-year-old children.

Authors:  Erin M Harvey; Velma Dobson; Joseph M Miller; Candice E Clifford-Donaldson; Tina K Green; Dawn H Messer; Katherine A Garvey
Journal:  J AAPOS       Date:  2009-10       Impact factor: 1.220

7.  Impact of confidence number on accuracy of the SureSight Vision Screener.

Authors: 
Journal:  Optom Vis Sci       Date:  2010-02       Impact factor: 1.973

8.  Utility of an open field Shack-Hartmann aberrometer for measurement of refractive error in infants and young children.

Authors:  Erin M Harvey; Joseph M Miller; Jim Schwiegerling
Journal:  J AAPOS       Date:  2013-10       Impact factor: 1.220

9.  Intertester agreement in refractive error measurements.

Authors:  Jiayan Huang; Maureen G Maguire; Elise Ciner; Marjean T Kulp; Graham E Quinn; Deborah Orel-Bixler; Lynn A Cyert; Bruce Moore; Gui-Shuang Ying
Journal:  Optom Vis Sci       Date:  2013-10       Impact factor: 1.973

10.  Prevalence and Associated Factors of Amblyopia Among School Age Children at Bahir Dar City, Northwest Ethiopia: A Community-Based Cross-Sectional Study.

Authors:  Mebratu Mulusew Tegegne; Abel Sinshaw Assem; Yosef Antehun Merie
Journal:  Clin Optom (Auckl)       Date:  2021-05-18
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.