Intertester agreement in refractive error measurements.

Jiayan Huang1, Maureen G Maguire, Elise Ciner, Marjean T Kulp, Graham E Quinn, Deborah Orel-Bixler, Lynn A Cyert, Bruce Moore, Gui-Shuang Ying.   

Abstract

PURPOSE: To determine the intertester agreement of refractive error measurements between lay and nurse screeners using the Retinomax Autorefractor and the SureSight Vision Screener.
METHODS: Trained lay and nurse screeners measured refractive error in 1452 preschoolers (3 to 5 years old) using the Retinomax and the SureSight in a random order for screeners and instruments. Intertester agreement between lay and nurse screeners was assessed for sphere, cylinder, and spherical equivalent (SE) using the mean difference and the 95% limits of agreement. The mean intertester difference (lay minus nurse) was compared between groups defined based on the child's age, cycloplegic refractive error, and the reading's confidence number using analysis of variance. The limits of agreement were compared between groups using the Brown-Forsythe test. Intereye correlation was accounted for in all analyses.
RESULTS: The mean intertester differences (95% limits of agreement) were -0.04 (-1.63, 1.54) diopter (D) sphere, 0.00 (-0.52, 0.51) D cylinder, and -0.04 (1.65, 1.56) D SE for the Retinomax and 0.05 (-1.48, 1.58) D sphere, 0.01 (-0.58, 0.60) D cylinder, and 0.06 (-1.45, 1.57) D SE for the SureSight. For either instrument, the mean intertester differences in sphere and SE did not differ by the child's age, cycloplegic refractive error, or the reading's confidence number. However, for both instruments, the limits of agreement were wider when eyes had significant refractive error or the reading's confidence number was below the manufacturer's recommended value.
CONCLUSIONS: Among Head Start preschool children, trained lay and nurse screeners agree well in measuring refractive error using the Retinomax or the SureSight. Both instruments had similar intertester agreement in refractive error measurements independent of the child's age. Significant refractive error and a reading with low confidence number were associated with worse intertester agreement.

Entities:  

Mesh:

Substances:

Year:  2013        PMID: 23974664      PMCID: PMC3894690          DOI: 10.1097/OPX.0000000000000025

Source DB:  PubMed          Journal:  Optom Vis Sci        ISSN: 1040-5488            Impact factor:   1.973


  17 in total

Review 1.  Measuring agreement in method comparison studies.

Authors:  J M Bland; D G Altman
Journal:  Stat Methods Med Res       Date:  1999-06       Impact factor: 3.021

2.  Measurement of refractive error in Native American preschoolers: validity and reproducibility of autorefraction.

Authors:  E M Harvey; J M Miller; V Dobson; R Tyszko; A L Davis
Journal:  Optom Vis Sci       Date:  2000-03       Impact factor: 1.973

3.  The amblyopia treatment study visual acuity testing protocol.

Authors:  J M Holmes; R W Beck; M X Repka; D A Leske; R T Kraker; R C Blair; P S Moke; E E Birch; R A Saunders; R W Hertle; G E Quinn; K A Simons; J M Miller
Journal:  Arch Ophthalmol       Date:  2001-09

4.  Testing young infants with the Welch Allyn suresight non-cycloplegic autorefractor.

Authors:  Russell J Adams; Susan M Dalton; Angela M Murphy; Heather L Hall; Mary L Courage
Journal:  Ophthalmic Physiol Opt       Date:  2002-11       Impact factor: 3.117

5.  Screening for refractive errors in children: accuracy of the hand held refractor Retinomax to screen for astigmatism.

Authors:  M Cordonnier; M Dramaix
Journal:  Br J Ophthalmol       Date:  1999-02       Impact factor: 4.638

6.  Repeatability of subjective and objective refraction.

Authors:  M Rosenfield; N N Chiu
Journal:  Optom Vis Sci       Date:  1995-08       Impact factor: 1.973

7.  ROC analysis of the accuracy of Noncycloplegic retinoscopy, Retinomax Autorefractor, and SureSight Vision Screener for preschool vision screening.

Authors:  Gui-shuang Ying; Maureen Maguire; Graham Quinn; Marjean Taylor Kulp; Lynn Cyert
Journal:  Invest Ophthalmol Vis Sci       Date:  2011-12-28       Impact factor: 4.799

8.  Screening for refractive errors in children: the plusoptiX S08 and the Retinomax K-plus2 performed by a lay screener compared to cycloplegic retinoscopy.

Authors:  Tamara Paff; Anne Marie Oudesluys-Murphy; Ron Wolterbeek; Marietta Swart-van den Berg; Johan M de Nie; Els Tijssen; Nicoline E Schalij-Delfos
Journal:  J AAPOS       Date:  2010-12       Impact factor: 1.220

9.  Reproducibility and accuracy of measurements with a hand held autorefractor in children.

Authors:  E M Harvey; J M Miller; L K Wagner; V Dobson
Journal:  Br J Ophthalmol       Date:  1997-11       Impact factor: 4.638

10.  Comparison of preschool vision screening tests as administered by licensed eye care professionals in the Vision In Preschoolers Study.

Authors:  Paulette Schmidt; Maureen Maguire; Velma Dobson; Graham Quinn; Elise Ciner; Lynn Cyert; Marjean Taylor Kulp; Bruce Moore; Deborah Orel-Bixler; Maryann Redford; Gui-shuang Ying
Journal:  Ophthalmology       Date:  2004-04       Impact factor: 12.079

View more
  3 in total

1.  Comparison of cycloplegic refraction between Grand Seiko autorefractor and Retinomax autorefractor in the Vision in Preschoolers-Hyperopia in Preschoolers (VIP-HIP) Study.

Authors:  Gui-Shuang Ying; Maureen G Maguire; Marjean Taylor Kulp; Elise Ciner; Bruce Moore; Maxwell Pistilli; Rowan Candy
Journal:  J AAPOS       Date:  2017-05-18       Impact factor: 1.220

Review 2.  Vision Screening, Vision Disorders, and Impacts of Hyperopia in Young Children: Outcomes of the Vision in Preschoolers (VIP) and Vision in Preschoolers - Hyperopia in Preschoolers (VIP-HIP) Studies.

Authors:  Marjean Taylor Kulp; Elise Ciner; Gui-Shuang Ying; T Rowan Candy; Bruce D Moore; Deborah Orel-Bixler
Journal:  Asia Pac J Ophthalmol (Phila)       Date:  2022-01-18

3.  Evidence-based preschool-age vision screening: health policy considerations.

Authors:  Deena Rachel Zimmerman; Hadas Ben-Eli; Bruce Moore; Monique Toledano; Chen Stein-Zamir; Ariela Gordon-Shaag
Journal:  Isr J Health Policy Res       Date:  2019-09-12
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.