Literature DB >> 15529910

Radiation peak skin dose to risk stratify electrophysiological procedures for deterministic skin damage.

J R Paisey1, A M Yue, A White, A Moss, J M Morgan, P R Roberts.   

Abstract

UNLABELLED: Ionising radiation is has the potential to cause harm both by increasing the probability future malignancy (stochastic mechanisms) and by direct physical injury (deterministic mechanisms). Several measures have been developed to quantify radiation exposure during a procedure and cardiologists usually refer to fluoroscopic screening time (FST). FST, however, has limitations for predicting deterministic injury which is directly dependant on peak skin dose (PSD). We compared FST to PSD for a range of interventional cardiac electrophysiology procedures.
METHODS: All patients undergoing electrophysiology procedures during a 2-month period in our institution were studied. Demographic details, nature of procedure, FST and PSD were measured. The FST to PSD ratio was calculated and compared between patient and procedural factors.
RESULTS: 67 procedures on patients (23 female) with body mass index (BMI) of 28 (SD 5) Kg/m2 were studied. Screening times ranged from 0.2 to 96.6 min (median 11.2). PSD ranged from <0.1 to 1108 mGy (median 141). There was a positive correlation between PSD to FST ratio and BMI (r = 0.59, p < 0.001). The PSD to FST ratio was higher in cardiac resynchronization therapy (CRT) devices than single or dual chamber ICDs (p = 0.002).
CONCLUSION: FST is not a reliable predictor of deterministic skin injury and in high-risk procedures such as CRT devices and those on individuals of high BMI PSD should be measured.

Entities:  

Mesh:

Year:  2004        PMID: 15529910     DOI: 10.1023/b:caim.0000041943.73199.d3

Source DB:  PubMed          Journal:  Int J Cardiovasc Imaging        ISSN: 1569-5794            Impact factor:   2.357


  14 in total

1.  Skin dose and dose-area product values for interventional cardiology procedures.

Authors:  E Vano; L Gonzalez; J I Ten; J M Fernandez; E Guibelalde; C Macaya
Journal:  Br J Radiol       Date:  2001-01       Impact factor: 3.039

Review 2.  Skin injuries from fluoroscopically guided procedures: part 1, characteristics of radiation injury.

Authors:  T R Koenig; D Wolff; F A Mettler; L K Wagner
Journal:  AJR Am J Roentgenol       Date:  2001-07       Impact factor: 3.959

Review 3.  The skin: its structure and response to ionizing radiation.

Authors:  J W Hopewell
Journal:  Int J Radiat Biol       Date:  1990-04       Impact factor: 2.694

4.  A method for permanent transvenous left ventricular pacing.

Authors:  J J Blanc; D G Benditt; M Gilard; Y Etienne; J Mansourati; K G Lurie
Journal:  Pacing Clin Electrophysiol       Date:  1998-11       Impact factor: 1.976

5.  Radiation exposure during standard and complex interventional procedures.

Authors:  T Vehmas
Journal:  Br J Radiol       Date:  1997-03       Impact factor: 3.039

Review 6.  Dose, effect severity, and imparted energy in assessing biological effects.

Authors:  V P Bond
Journal:  Stem Cells       Date:  1995-05       Impact factor: 6.277

Review 7.  Potential biological effects following high X-ray dose interventional procedures.

Authors:  L K Wagner; P J Eifel; R A Geise
Journal:  J Vasc Interv Radiol       Date:  1994 Jan-Feb       Impact factor: 3.464

Review 8.  Patient dosimetry measurement methods.

Authors:  K Faulkner; D A Broadhead; R M Harrison
Journal:  Appl Radiat Isot       Date:  1999-01       Impact factor: 1.513

9.  Pulmonary vein disconnection using the LocaLisa three-dimensional nonfluoroscopic catheter imaging system.

Authors:  Laurent Macle; Pierre Jaïs; Christophe Scavée; Rukshen Weerasooriya; Mélèze Hocini; Dipen C Shah; Florence Raybaud; Kee-Joon Choi; Jacques Clémenty; Michel Haïssaguerre
Journal:  J Cardiovasc Electrophysiol       Date:  2003-07

10.  Entrance skin dose estimates derived from dose-area product measurements in interventional radiological procedures.

Authors:  B J McParland
Journal:  Br J Radiol       Date:  1998-12       Impact factor: 3.039

View more
  1 in total

1.  Patient radiation doses in interventional cardiology procedures.

Authors:  Ioannis Pantos; Georgios Patatoukas; Demosthenes G Katritsis; Efstathios Efstathopoulos
Journal:  Curr Cardiol Rev       Date:  2009-01
  1 in total