Abstract
Four cases of failure of distal femoral Less Invasive Stabilization System plates are reported. Two modes of failure were observed: in 2 cases, the Less Invasive Stabilization System plates broke, and in 2 cases, the plates gradually displaced from the femur as the shaft screws lost purchase in the diaphyseal femoral cortex. Delayed union/nonunion, early weight bearing, and improper plate placement were identified as the etiologies for plate failure.Entities:
Mesh:
Year: 2004 PMID: 15475855 DOI: 10.1097/00005131-200409000-00015