Using a wrist-worn device based on peripheral arterial tonometry to diagnose obstructive sleep apnea: in-laboratory and ambulatory validation.

Stephen D Pittman1, Najib T Ayas, Mary M MacDonald, Atul Malhotra, Robert B Fogel, David P White.   

Abstract

STUDY
OBJECTIVES: To assess the accuracy of a wrist-worn device (Watch_PAT 100) to diagnose obstructive sleep apnea in the home.
DESIGN: Participants completed 2 overnight diagnostic studies with the test device: 1 night in the laboratory with concurrent polysomnography and 1 night in the home with only the Watch_PAT. The order of the laboratory and home study nights was random. The frequency of respiratory events on the PSG was quantified using indexes based on 2 definitions of hypopnea: the respiratory disturbance index (RDI) using American Academy of Sleep Medicine Task Force criteria for clinical research, also referred to as the Chicago criteria (RDI.C), and the Medicare guidelines (RDI.M). The Watch_PAT RDI (PAT RDI) and oxygen desaturation index (PAT ODI) were then evaluated against the polysomnography RDI.C and RDI.M, respectively, for both Watch_PAT diagnostic nights, yielding IN-LAB and HOME-LAB comparisons.
SETTING: Sleep laboratory affiliated with a tertiary-care academic medical center. PATIENTS: 30 patients referred with suspected OSA.
INTERVENTIONS: N/A. MEASUREMENTS AND
RESULTS: The polysomnography and PAT measures were compared using the mean [2 SD] of the differences and the intra-class correlation coefficient (ICC). The receiver-operator characteristic curve was used to assess optimum sensitivity and specificity and calculate likelihood ratios. For the IN-LAB comparison, there was high concordance between RDI.C and PAT RDI (ICC = 0.88, mean difference 2.5 [18.9] events per hour); RDI.M and PAT ODI (ICC = 0.95, mean difference 1.4 [12.9] events per hour; and sleep time (ICC = 0.70, mean difference 7.0 [93.1] minutes) between the test device and PSG. For the HOME-LAB comparison, there was good concordance between RDI.C and PAT RDI (ICC = 0.72, mean difference 1.4 [30.1] events per hour) and RDI.M and PAT ODI (ICC = 0.80, mean difference 1.6 [26.4] events per hour) for the test device and PSG. Home studies were performed with no technical failures.
CONCLUSIONS: In a population of patients suspected of having obstructive sleep apnea, the Watch_PAT can quantify an ODI that compares very well with Medicare criteria for defining respiratory events and an RDI that compares favorably with Chicago criteria for defining respiratory events. The device can be used with a low failure rate for single use in the lab and home for self-administered testing.

Entities:  

Mesh:

Substances:

Year:  2004        PMID: 15453551      PMCID: PMC3818284          DOI: 10.1093/sleep/27.5.923

Source DB:  PubMed          Journal:  Sleep        ISSN: 0161-8105            Impact factor:   5.849


  41 in total

1.  Proposed supplements and amendments to 'A Manual of Standardized Terminology, Techniques and Scoring System for Sleep Stages of Human Subjects', the Rechtschaffen & Kales (1968) standard.

Authors:  T Hori; Y Sugita; E Koga; S Shirakawa; K Inoue; S Uchida; H Kuwahara; M Kousaka; T Kobayashi; Y Tsuji; M Terashima; K Fukuda; N Fukuda
Journal:  Psychiatry Clin Neurosci       Date:  2001-06       Impact factor: 5.188

2.  Hypopnea in sleep-disordered breathing in adults.

Authors:  A L Meoli; K R Casey; R W Clark; J A Coleman; R W Fayle; R J Troell; C Iber
Journal:  Sleep       Date:  2001-06-15       Impact factor: 5.849

3.  The effect of upper airway obstruction and arousal on peripheral arterial tonometry in obstructive sleep apnea.

Authors:  Christopher P O'Donnell; Lawrence Allan; Paul Atkinson; Alan R Schwartz
Journal:  Am J Respir Crit Care Med       Date:  2002-10-01       Impact factor: 21.405

Review 4.  Home diagnosis of sleep apnea: a systematic review of the literature. An evidence review cosponsored by the American Academy of Sleep Medicine, the American College of Chest Physicians, and the American Thoracic Society.

Authors:  W Ward Flemons; Michael R Littner; James A Rowley; Peter Gay; W McDowell Anderson; David W Hudgel; R Douglas McEvoy; Daniel I Loube
Journal:  Chest       Date:  2003-10       Impact factor: 9.410

5.  Measuring agreement between diagnostic devices.

Authors:  W Ward Flemons; Michael R Littner
Journal:  Chest       Date:  2003-10       Impact factor: 9.410

6.  Short-term variability of respiration and sleep during unattended nonlaboratory polysomnography--the Sleep Heart Health Study. [corrected].

Authors:  Stuart F Quan; Michael E Griswold; Conrad Iber; F Javier Nieto; David M Rapoport; Susan Redline; Mark Sanders; Terry Young
Journal:  Sleep       Date:  2002-12       Impact factor: 5.849

7.  Home unattended vs hospital telemonitored polysomnography in suspected obstructive sleep apnea syndrome: a randomized crossover trial.

Authors:  Frédéric Gagnadoux; Nathalie Pelletier-Fleury; Carole Philippe; Dominique Rakotonanahary; Bernard Fleury
Journal:  Chest       Date:  2002-03       Impact factor: 9.410

8.  Evaluation of a portable device based on peripheral arterial tone for unattended home sleep studies.

Authors:  Amir Bar; Giora Pillar; Itsik Dvir; Jacob Sheffy; Robert P Schnall; Peretz Lavie
Journal:  Chest       Date:  2003-03       Impact factor: 9.410

9.  Assessment of a wrist-worn device in the detection of obstructive sleep apnea.

Authors:  Najib T Ayas; Stephen Pittman; Mary MacDonald; David P White
Journal:  Sleep Med       Date:  2003-09       Impact factor: 3.492

10.  Practice parameters for the use of portable monitoring devices in the investigation of suspected obstructive sleep apnea in adults.

Authors:  Andrew L Chesson; Richard B Berry; Allan Pack
Journal:  Sleep       Date:  2003-11-01       Impact factor: 5.849

View more
  53 in total

1.  Preeclampsia and Sleep-Disordered Breathing: A Case-Control Study.

Authors:  Francesca L Facco; Justin Lappen; Courtney Lim; Phyllis C Zee; William A Grobman
Journal:  Pregnancy Hypertens       Date:  2013-04       Impact factor: 2.899

Review 2.  Obstructive sleep apnea devices for out-of-center (OOC) testing: technology evaluation.

Authors:  Nancy A Collop; Sharon L Tracy; Vishesh Kapur; Reena Mehra; David Kuhlmann; Sam A Fleishman; Joseph M Ojile
Journal:  J Clin Sleep Med       Date:  2011-10-15       Impact factor: 4.062

3.  Monitoring peripheral arterial tone (PAT) to diagnose sleep apnea in the home.

Authors:  David P White
Journal:  J Clin Sleep Med       Date:  2008-02-15       Impact factor: 4.062

4.  Development of a pregnancy-specific screening tool for sleep apnea.

Authors:  Francesca L Facco; David W Ouyang; Phyllis C Zee; William A Grobman
Journal:  J Clin Sleep Med       Date:  2012-08-15       Impact factor: 4.062

5.  Survey regarding limited diagnostic systems for sleep apnea.

Authors:  Philip R Westbrook
Journal:  J Clin Sleep Med       Date:  2007-04-15       Impact factor: 4.062

6.  Follow-up assessment of CPAP efficacy in patients with obstructive sleep apnea using an ambulatory device based on peripheral arterial tonometry.

Authors:  Stephen D Pittman; Giora Pillar; Richard B Berry; Atul Malhotra; Mary M MacDonald; David P White
Journal:  Sleep Breath       Date:  2006-09       Impact factor: 2.816

7.  Reliability of the Watch-PAT 200 in detecting sleep apnea in highway bus drivers.

Authors:  Melike Yuceege; Hikmet Firat; Ahmet Demir; Sadik Ardic
Journal:  J Clin Sleep Med       Date:  2013-04-15       Impact factor: 4.062

8.  Portable monitoring and autotitration versus polysomnography for the diagnosis and treatment of sleep apnea.

Authors:  Richard B Berry; Gilbert Hill; Linda Thompson; Valorea McLaurin
Journal:  Sleep       Date:  2008-10       Impact factor: 5.849

9.  Validating the Watch-PAT for Diagnosing Obstructive Sleep Apnea in Adolescents.

Authors:  Ji Ho Choi; Bora Lee; Jae Yong Lee; Hyun Jun Kim
Journal:  J Clin Sleep Med       Date:  2018-10-15       Impact factor: 4.062

10.  Evaluation of a portable recording device (ApneaLink) for case selection of obstructive sleep apnea.

Authors:  Hui Chen; Alan A Lowe; Yuxing Bai; Peter Hamilton; John A Fleetham; Fernanda R Almeida
Journal:  Sleep Breath       Date:  2008-12-04       Impact factor: 2.816

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.