Literature DB >> 12628865

Evaluation of a portable device based on peripheral arterial tone for unattended home sleep studies.

Amir Bar1, Giora Pillar, Itsik Dvir, Jacob Sheffy, Robert P Schnall, Peretz Lavie.   

Abstract

BACKGROUND: Diagnosis of obstructive sleep apnea syndrome (OSAS) by ambulatory systems is a growing practice in view of the large number of patients awaiting correct diagnosis. The Watch PAT100 (WP100) [Itamar Medical; Caesarea, Israel] is a portable device based on the peripheral arterial tone (PAT) signal, and is designed for unattended home sleep studies.
OBJECTIVES: To evaluate the efficacy, reliability, and reproducibility of the WP100 device for the diagnosis of OSAS as compared to in-laboratory, standard polysomnographic-based manual scoring. DESIGN AND METHODS: One hundred two subjects (78 men; 69 patients with OSAS and 33 normal volunteers; mean +/- SD age, 41.4 +/- 15.2 years; body mass index, 26.8 +/- 5.5) underwent in-laboratory full polysomnography simultaneously with WP100 recording. Fourteen subjects also underwent two additional unattended home sleep studies with the WP100 alone. The polysomnography recordings were blindly scored for apnea/hypopnea according to the American Academy of Sleep Medicine criteria (1999), and the polysomnography respiratory disturbance index (RDI) [PSG-RDI] was calculated. The WP100 data were analyzed automatically for the PAT RDI (PRDI) by a proprietary algorithm that was previously developed on an independent group of subjects.
RESULTS: Across a wide range of RDI levels, the PRDI was highly correlated with the PSG-RDI (r = 0.88, p < 0.0001), with an area under the receiver operating characteristic curve of 0.82 and 0.87 for thresholds of 10 events per hour and 20 events per hour, respectively. The PRDI scores were also highly reproducible, showing high correlation between home and in-laboratory sleep studies (r = 0.89, p < 0.001).
CONCLUSION: The WP100 may offer an accurate, robust, and reliable ambulatory method for the detection of OSAS, with minimal patient discomfort.

Entities:  

Mesh:

Year:  2003        PMID: 12628865     DOI: 10.1378/chest.123.3.695

Source DB:  PubMed          Journal:  Chest        ISSN: 0012-3692            Impact factor:   9.410


  61 in total

1.  Changes in cerebral hemoglobin indices in obstructive sleep apnea syndrome with nasal continuous positive airway pressure treatment.

Authors:  Akira Matsuo; Yuichi Inoue; Kazuyoshi Namba; Hiroshige Chiba
Journal:  Sleep Breath       Date:  2010-06-30       Impact factor: 2.816

2.  Preeclampsia and Sleep-Disordered Breathing: A Case-Control Study.

Authors:  Francesca L Facco; Justin Lappen; Courtney Lim; Phyllis C Zee; William A Grobman
Journal:  Pregnancy Hypertens       Date:  2013-04       Impact factor: 2.899

Review 3.  Obstructive sleep apnea devices for out-of-center (OOC) testing: technology evaluation.

Authors:  Nancy A Collop; Sharon L Tracy; Vishesh Kapur; Reena Mehra; David Kuhlmann; Sam A Fleishman; Joseph M Ojile
Journal:  J Clin Sleep Med       Date:  2011-10-15       Impact factor: 4.062

4.  Pre-eclampsia is associated with sleep-disordered breathing and endothelial dysfunction.

Authors:  D Yinon; L Lowenstein; S Suraya; R Beloosesky; O Zmora; A Malhotra; G Pillar
Journal:  Eur Respir J       Date:  2006-02       Impact factor: 16.671

5.  Monitoring peripheral arterial tone (PAT) to diagnose sleep apnea in the home.

Authors:  David P White
Journal:  J Clin Sleep Med       Date:  2008-02-15       Impact factor: 4.062

6.  Follow-up assessment of CPAP efficacy in patients with obstructive sleep apnea using an ambulatory device based on peripheral arterial tonometry.

Authors:  Stephen D Pittman; Giora Pillar; Richard B Berry; Atul Malhotra; Mary M MacDonald; David P White
Journal:  Sleep Breath       Date:  2006-09       Impact factor: 2.816

7.  Reliability of the Watch-PAT 200 in detecting sleep apnea in highway bus drivers.

Authors:  Melike Yuceege; Hikmet Firat; Ahmet Demir; Sadik Ardic
Journal:  J Clin Sleep Med       Date:  2013-04-15       Impact factor: 4.062

8.  Portable monitoring and autotitration versus polysomnography for the diagnosis and treatment of sleep apnea.

Authors:  Richard B Berry; Gilbert Hill; Linda Thompson; Valorea McLaurin
Journal:  Sleep       Date:  2008-10       Impact factor: 5.849

9.  Validating the Watch-PAT for Diagnosing Obstructive Sleep Apnea in Adolescents.

Authors:  Ji Ho Choi; Bora Lee; Jae Yong Lee; Hyun Jun Kim
Journal:  J Clin Sleep Med       Date:  2018-10-15       Impact factor: 4.062

10.  Effective Apnea-Hypopnea Index ("Effective AHI"): A New Measure of Effectiveness for Positive Airway Pressure Therapy.

Authors:  Scott B Boyd; Raghu Upender; Arthur S Walters; R Lucas Goodpaster; Jeffrey J Stanley; Li Wang; Rameela Chandrasekhar
Journal:  Sleep       Date:  2016-11-01       Impact factor: 5.849

View more