The influence of head and neck geometry on stability of total hip replacement: a mechanical test study.

Rainer Bader1, Roger Scholz, Erwin Steinhauser, Susanne Zimmermann, Raymonde Busch, Wolfram Mittelmeier.   

Abstract

BACKGROUND: Dislocation after replacement may be caused by poor implant design or positioning, or by the surgical approach taken. We evaluated the influence of head and neck design on range of motion and stability (with respect to risk of dislocation) in total hip endoprostheses.
MATERIAL AND METHODS: Using a test device, we determined the stability afforded by different head sizes and neck geometries for various implant positions.
RESULTS: Increasing head diameter led to an enhancement of range of motion as well as resistance against subluxation, and thus to improved stability in any movement combination and implant orientation. Smaller femoral heads were associated with increased risk of dislocation, especially in a poor implant position such as retroversion, and steep positioning of the cup. Skirted metal or mushroom-shaped ceramic heads had a reduced range of motion until impingement of approx. 20 degrees, as compared to spherical standard heads. Furthermore, after identical joint loading, skirted heads dislocated more readily than standard heads with corresponding diameters.
INTERPRETATION: To obtain sufficient joint mobility and stability, neck geometry and implant position should be considered when choosing the femoral head size. Copyright 2004 Taylor & Francis

Entities:  

Mesh:

Year:  2004        PMID: 15370584     DOI: 10.1080/00016470410001178-1

Source DB:  PubMed          Journal:  Acta Orthop Scand        ISSN: 0001-6470


  11 in total

1.  Accuracy and reliability of different methods to evaluate the acetabular cup version from plain radiographs.

Authors:  Maximilian Haenle; Wolfram Mittelmeier; Rodrigo Barbano; Klaus Wörtler; Roger Scholz; Rainer Bader
Journal:  Surg Radiol Anat       Date:  2010-06-25       Impact factor: 1.246

Review 2.  Intra-prosthetic dislocation of dual-mobility cups after total hip arthroplasty: potential causes from a clinical and biomechanical perspective.

Authors:  Christian Fabry; Jean Langlois; Moussa Hamadouche; Rainer Bader
Journal:  Int Orthop       Date:  2015-10-01       Impact factor: 3.075

3.  [Resurfacing arthroplasty of the hip].

Authors:  M Rudert; L Gerdesmeyer; H Rechl; P Juhnke; R Gradinger
Journal:  Orthopade       Date:  2007-04       Impact factor: 1.087

4.  Bony impingement limits design-related increases in hip range of motion.

Authors:  Adam Bunn; Clifford W Colwell; Darryl D D'Lima
Journal:  Clin Orthop Relat Res       Date:  2012-02       Impact factor: 4.176

5.  Effects of design parameters of total hip components on the impingement angle and determination of the preferred liner skirt shape with an adequate oscillation angle.

Authors:  Hsiao-Che Lin; Wei-Min Chi; Ying-Jui Ho; Jian-Horng Chen
Journal:  Med Biol Eng Comput       Date:  2012-11-29       Impact factor: 2.602

6.  Instability after total hip arthroplasty.

Authors:  Brian C Werner; Thomas E Brown
Journal:  World J Orthop       Date:  2012-08-18

7.  Effects of ball head diameter and stem neck shape in range of motion after total hip arthroplasty: A simulation study.

Authors:  Tsutomu Higashi; Nobuhiro Kaku; Shouhei Noda; Tomonori Tabata; Hiroaki Tagomori; Hiroshi Tsumura
Journal:  J Orthop       Date:  2019-09-11

Review 8.  [The importance of wear couples for younger endoprosthesis patients].

Authors:  J Kircher; P Bergschmidt; R Bader; D Kluess; E Besser-Mahuzir; A Leder; W Mittelmeier
Journal:  Orthopade       Date:  2007-04       Impact factor: 1.087

Review 9.  [Biomechanical aspects of the implant fixation and kinematics of hip resurfacing systems].

Authors:  R Bader; D Klüss; L Gerdesmeyer; E Steinhauser
Journal:  Orthopade       Date:  2008-07       Impact factor: 1.087

10.  Similar range of motion and function after resurfacing large-head or standard total hip arthroplasty.

Authors:  Jeannette Østergaard Penny; Ole Ovesen; Jens-Erik Varmarken; Søren Overgaard
Journal:  Acta Orthop       Date:  2013-03-26       Impact factor: 3.717

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.