Comparison of the psychophysical properties of various intraoral film and digital systems by means of the perceptibility curve test.

K Yoshiura1, U Welander, W D McDavid, G Li, X-Q Shi, E Nakayama, M Shimizu, K Okamura, S Kanda.   

Abstract

OBJECTIVES: To compare psychophysical properties of two intraoral films and three digital systems using the perceptibility curve (PC) test.
MATERIALS AND METHODS: A test object was used to determine the exposures and exposure differences between the total thickness of the test object and details consisting of holes of increasing depth. The PCs for the two intraoral films, UltraSpeed and EktaSpeed Plus, were constructed employing exposure and exposure differences from dose response functions. Integrals of the PCs were calculated to obtain the psychophysical properties of the two films. Psychophysical properties of the two films were compared with those of the three digital systems published previously (CDR, Dixel and Digora).
RESULTS: The PC for the EktaSpeed Plus showed a slightly higher peak than that for the UltraSpeed. Available exposure ranges were comparable. The PC for the EktaSpeed Plus was shifted to the left of the exposure axis indicating its higher sensitivity as compared with UltraSpeed. All three digital systems had narrower but higher peaks compared with the films. The integrals for the digital systems were considerably larger than those for the two film types.
CONCLUSIONS: All the three digital systems have superior psychophysical properties compared with the two tested films.

Entities:  

Mesh:

Year:  2004        PMID: 15314001     DOI: 10.1259/dmfr/29102849

Source DB:  PubMed          Journal:  Dentomaxillofac Radiol        ISSN: 0250-832X            Impact factor:   2.419


  3 in total

1.  Comparison of psychophysical properties of two intraoral digital sensors on low-contrast perceptibility.

Authors:  X-Q Shi; D Benchimol; K Näsström
Journal:  Dentomaxillofac Radiol       Date:  2013-10-29       Impact factor: 2.419

2.  Evaluation of low-contrast perceptibility in dental restorative materials under the influence of ambient light conditions.

Authors:  A D Cruz; I C Lobo; A L B Lemos; M F Aguiar
Journal:  Dentomaxillofac Radiol       Date:  2015-01-28       Impact factor: 2.419

Review 3.  The missing link in image quality assessment in digital dental radiography.

Authors:  Kazutoshi Okamura; Kazunori Yoshiura
Journal:  Oral Radiol       Date:  2019-07-13       Impact factor: 1.852

  3 in total