Is a single anterolateral screw-plate fixation sufficient for the treatment of spinal fractures in the thoracolumbar junction? A biomechanical in vitro investigation.

Ulrich Schreiber1, Tibor Bence, Thomas Grupp, Erwin Steinhauser, Thomas Mückley, Wolfram Mittelmeier, Rudolf Beisse.   

Abstract

Controversy exists about the indications, advantages and disadvantages of various surgical techniques used for anterior interbody fusion of spinal fractures in the thoracolumbar junction. The purpose of this study was to evaluate the stabilizing effect of an anterolateral and thoracoscopically implantable screw-plate system. Six human bisegmental spinal units (T12-L2) were used for the biomechanical in vitro testing procedure. Each specimen was tested in three different scenarios: (1) intact spinal segments vs (2) monosegmental (T12/L1) anterolateral fixation (macsTL, Aesculap, Germany) with an interbody bone strut graft from the iliac crest after both partial corpectomy (L1) and discectomy (T12/L1) vs (3) bisegmental anterolateral instrumentation after extended partial corpectomy (L1), and bisegmental discectomy (T12/L1 and L1/L2). Specimens were loaded with an alternating, nondestructive maximum bending moment of +/-7.5 Nm in six directions: flexion/extension, right and left lateral bending, and right and left axial rotation. Motion analysis was performed by a contact-less three-dimensional optical measuring system. Segmental stiffness of the three different scenarios was evaluated by the relative alteration of the intervertebral angles in the three main anatomical planes. With each stabilization technique, the specimens were more rigid, compared with the intact spine, for flexion/extension (sagittal plane) as well as in left and right lateral bending (frontal plane). In these planes the bisegmental instrumentation compared to the monosegmental case had an even larger stiffening effect on the specimens. In contrast to these findings, axial rotation showed a modest increase of motion after bisegmental instrumentation. To conclude, the immobilization of monosegmental fractures in the thoracolumbar junction can be secured by means of bone grafting and the implant used in this study for all three anatomical planes. After bisegmental anterolateral stabilization a sufficient reduction of the movements was registered for flexion/extension and lateral bending. However, the observed slight increase of the range of motion in the transversal plane may lead to loosening of the implant before union. Therefore, the use of an additional dorsal fixation device should be considered.

Entities:  

Mesh:

Year:  2004        PMID: 15243790      PMCID: PMC3476694          DOI: 10.1007/s00586-004-0770-9

Source DB:  PubMed          Journal:  Eur Spine J        ISSN: 0940-6719            Impact factor:   3.134


  32 in total

1.  [Biomechanical stability with a new artificial vertebral body implant. 3-dimensional movement analysis of instrumented human vertebral segments].

Authors:  C Knop; U Lange; L Bastian; M Blauth
Journal:  Unfallchirurg       Date:  2001-10       Impact factor: 1.000

2.  Combined anteroposterior spinal fixation provides superior stabilisation to a single anterior or posterior procedure.

Authors:  H J Wilke; V Kemmerich; L E Claes; M Arand
Journal:  J Bone Joint Surg Br       Date:  2001-05

3.  [Stability of ventral, dorsal and combined spondylodesis in vertebral body prosthesis implantation].

Authors:  M Vahldiek; F Gossè; M M Panjabi
Journal:  Orthopade       Date:  2002-05       Impact factor: 1.087

4.  [Mechanical testing of implant properties of thoracoscopic implantation of ventral spinal stabilizing systems. Comparative study with the ISO/DIS 12189-2 corpectomy model and an improved synthetic model].

Authors:  T M Grupp; R Beisse; M Potulski; T Marnay; J Beger; W Blömer
Journal:  Orthopade       Date:  2002-04       Impact factor: 1.087

5.  [Development and clinical application of a thoracoscopy implantable plate frame for treatment of thoracolumbar fractures and instabilities].

Authors:  R Beisse; M Potulski; J Beger; V Bühren
Journal:  Orthopade       Date:  2002-04       Impact factor: 1.087

6.  Stability potential of spinal instrumentations in tumor vertebral body replacement surgery.

Authors:  M J Vahldiek; M M Panjabi
Journal:  Spine (Phila Pa 1976)       Date:  1998-03-01       Impact factor: 3.468

7.  Bone grafts. The basic science rationale for clinical applications.

Authors:  G E Friedlaender
Journal:  J Bone Joint Surg Am       Date:  1987-06       Impact factor: 5.284

8.  The biology of bone graft repair.

Authors:  H Burchardt
Journal:  Clin Orthop Relat Res       Date:  1983-04       Impact factor: 4.176

9.  Minimally invasive ventral spondylodesis for thoracolumbar fracture treatment: surgical technique and first clinical outcome.

Authors:  Markus Schultheiss; Lothar Kinzl; Lutz Claes; Hans-Joachim Wilke; Erich Hartwig
Journal:  Eur Spine J       Date:  2003-07-31       Impact factor: 3.134

10.  [MAC-TL twin screw. A new thoracoscopic implantable stabilization system for treatment of vertebral fractures--implant design, implantation technique and in vitro testing].

Authors:  M Schultheiss; H J Wilke; L Claes; L Kinzl; E Hartwig
Journal:  Orthopade       Date:  2002-04       Impact factor: 1.087

View more
  5 in total

1.  Contribution of Round vs. Rectangular Expandable Cage Endcaps to Spinal Stability in a Cadaveric Corpectomy Model.

Authors:  Gregory M Mundis; Robert K Eastlack; Payam Moazzaz; Alexander W L Turner; G Bryan Cornwall
Journal:  Int J Spine Surg       Date:  2015-10-22

2.  Two column lesions in the thoracolumbar junction: anterior, posterior or combined approach? A comparative biomechanical in vitro investigation.

Authors:  Tibor Bence; Ulrich Schreiber; Thomas Grupp; Erwin Steinhauser; Wolfram Mittelmeier
Journal:  Eur Spine J       Date:  2006-08-30       Impact factor: 3.134

3.  En bloc spondylectomy reconstructions in a biomechanical in-vitro study.

Authors:  A C Disch; K D Schaser; I Melcher; A Luzzati; F Feraboli; W Schmoelz
Journal:  Eur Spine J       Date:  2008-01-15       Impact factor: 3.134

4.  The Effect of Polymethyl Methacrylate Augmentation on the Primary Stability of Cannulated Bone Screws in an Anterolateral Plate in Osteoporotic Vertebrae: A Human Cadaver Study.

Authors:  Matthias Rüger; Richard M Sellei; Marcus Stoffel; Christian von Rüden
Journal:  Global Spine J       Date:  2015-06-15

5.  Stepwise resection of the posterior ligamentous complex for stability of a thoracolumbar compression fracture: An in vitro biomechanical investigation.

Authors:  Yao Li; Zhonghai Shen; Mingyu Huang; Xiangyang Wang
Journal:  Medicine (Baltimore)       Date:  2017-09       Impact factor: 1.889