[Mechanical testing of implant properties of thoracoscopic implantation of ventral spinal stabilizing systems. Comparative study with the ISO/DIS 12189-2 corpectomy model and an improved synthetic model].

T M Grupp1, R Beisse, M Potulski, T Marnay, J Beger, W Blömer.   

Abstract

A new modular anterior fixation system MACS TL (modular anterior construct system for the thoracic and lumbar spine) has been developed for use in thoracoscopic spondylodesis. This system demonstrates high angular stability and meets the surgical requirements for an endoscopic approach. The objective of the current study was fatigue testing of the MACS TL implant system using a corpectomy model according to ISO/DIS 12189-2 and a synthetic model recently developed by Kotani et al. [6]. The MACS TL system demonstrated good mechanical properties with a high stiffness compared to the published data reviewed. The importance of dynamic testing in a corpectomy model has been demonstrated by comparing the MACS TL plate system with an early prototype, which has not yet been clinically evaluated. The corpectomy model according to Kotani et al. offers an interesting alternative to the ISO/DIS 12189-2 test method for asymmetrically designed and antero-laterally positioned spinal implants due to the unconstrained ball joint.

Entities:  

Mesh:

Year:  2002        PMID: 12056284     DOI: 10.1007/s00132-001-0284-7

Source DB:  PubMed          Journal:  Orthopade        ISSN: 0085-4530            Impact factor:   1.087


  5 in total

1.  Comparison of the biomechanical effects of posterior and anterior spine-stabilizing implants.

Authors:  Antonius Rohlmann; T Zander; G Bergmann
Journal:  Eur Spine J       Date:  2005-02-17       Impact factor: 3.134

2.  Primary stability of anterior lumbar stabilization: interdependence of implant type and endplate retention or removal.

Authors:  Christian H Flamme; Nadine von der Heide; Caroline Heymann; Christof Hurschler
Journal:  Eur Spine J       Date:  2005-08-10       Impact factor: 3.134

3.  Blood loss and operative duration using monopolar electrosurgery versus ultrasound scissors for surgical preparation during thoracoscopic ventral spondylodesis: results of a randomized, blinded, controlled trial.

Authors:  Christina Otto; Gereon Schiffer; Thorsten Tjardes; Henning Kunter; Peer Eysel; Thomas Paffrath
Journal:  Eur Spine J       Date:  2014-04-24       Impact factor: 3.134

4.  Is a single anterolateral screw-plate fixation sufficient for the treatment of spinal fractures in the thoracolumbar junction? A biomechanical in vitro investigation.

Authors:  Ulrich Schreiber; Tibor Bence; Thomas Grupp; Erwin Steinhauser; Thomas Mückley; Wolfram Mittelmeier; Rudolf Beisse
Journal:  Eur Spine J       Date:  2004-07-09       Impact factor: 3.134

5.  [Incomplete cranial burst fracture in the thoracolumbar junction. Results 6 years after thoracoscopic monosegmental spondylodesis].

Authors:  U J A Spiegl; S Hauck; P Merkel; V Bühren; O Gonschorek
Journal:  Unfallchirurg       Date:  2014-08       Impact factor: 1.000

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.