Literature DB >> 15218941

Performance of epimysial stimulating electrodes in the lower extremities of individuals with spinal cord injury.

James P Uhlir1, Ronald J Triolo, John A Davis, Carol Bieri.   

Abstract

This study describes the performance of surgically-implanted epimysial stimulating electrodes in the muscles of the lower extremities for use in functional neuromuscular stimulation (FNS) systems for standing after spinal cord injury. A total of 86 epimysial electrodes were implanted in 13 volunteers with low tetraplegia or paraplegia receiving the Case Western Reserve University/Veteran Affairs (CWRU/VA)-implanted standing/transfer neuroprosthesis. The neuroprosthesis consisted of bilateral epimysial electrodes in the knee and hip extensors (vastus lateralis, gluteus maximus, and adductor magnus or semimembranosus) and intramuscular electrodes at the T12/L1 or L1/L2 spinal roots for trunk extension. Recruitment properties, stimulated knee and hip extension moments, standing performance, and mechanical integrity over time were measured for a period up to four years post-implantation. Stimulated thresholds were stable and recruitment was sufficient to generate joint moments adequate for standing, with up to 97% body weight supported by the legs. Four mechanical failures were observed, all in the posterior muscles of the thigh, leaving 95% of all electrodes operational at all followup intervals. Probability of 24-month survival is estimated to be 93% plateauing to a steady state of 90% at four years. These results indicate that epimysial designs are appropriate for long-term clinical use in the large muscles of the lower extremities with implanted motor system neuroprostheses.

Entities:  

Mesh:

Year:  2004        PMID: 15218941     DOI: 10.1109/TNSRE.2004.827224

Source DB:  PubMed          Journal:  IEEE Trans Neural Syst Rehabil Eng        ISSN: 1534-4320            Impact factor:   3.802


  18 in total

1.  Comprehensive joint feedback control for standing by functional neuromuscular stimulation-a simulation study.

Authors:  Raviraj Nataraj; Musa L Audu; Robert F Kirsch; Ronald J Triolo
Journal:  IEEE Trans Neural Syst Rehabil Eng       Date:  2010-10-04       Impact factor: 3.802

2.  Center of mass acceleration feedback control for standing by functional neuromuscular stimulation: a simulation study.

Authors:  Raviraj Nataraj; Musa L Audu; Robert F Kirsch; Ronald J Triolo
Journal:  J Rehabil Res Dev       Date:  2012

3.  A model of selective activation of the femoral nerve with a flat interface nerve electrode for a lower extremity neuroprosthesis.

Authors:  Matthew A Schiefer; Ronald J Triolo; Dustin J Tyler
Journal:  IEEE Trans Neural Syst Rehabil Eng       Date:  2008-04       Impact factor: 3.802

4.  The design of and chronic tissue response to a composite nerve electrode with patterned stiffness.

Authors:  M J Freeberg; M A Stone; R J Triolo; D J Tyler
Journal:  J Neural Eng       Date:  2017-03-13       Impact factor: 5.379

5.  Effects of intramuscular trunk stimulation on manual wheelchair propulsion mechanics in 6 subjects with spinal cord injury.

Authors:  Ronald J Triolo; Stephanie Nogan Bailey; Lisa M Lombardo; Michael E Miller; Kevin Foglyano; Musa L Audu
Journal:  Arch Phys Med Rehabil       Date:  2013-04-26       Impact factor: 3.966

6.  Selective stimulation of the human femoral nerve with a flat interface nerve electrode.

Authors:  M A Schiefer; K H Polasek; R J Triolo; G C J Pinault; D J Tyler
Journal:  J Neural Eng       Date:  2010-03-08       Impact factor: 5.379

7.  Standing after spinal cord injury with four-contact nerve-cuff electrodes for quadriceps stimulation.

Authors:  Lee E Fisher; Michael E Miller; Stephanie N Bailey; John A Davis; James S Anderson; Lori Rhode; Dustin J Tyler; Ronald J Triolo
Journal:  IEEE Trans Neural Syst Rehabil Eng       Date:  2008-10       Impact factor: 3.802

8.  Intraoperative evaluation of the spiral nerve cuff electrode on the femoral nerve trunk.

Authors:  K H Polasek; M A Schiefer; G C J Pinault; R J Triolo; D J Tyler
Journal:  J Neural Eng       Date:  2009-11-09       Impact factor: 5.379

9.  Intraoperative demonstration of selective stimulation of the common human femoral nerve with a FINE.

Authors:  Matthew A Schiefer; Katharine H Polasek; Ronald J Triolo; Gilles C Pinault; Dustin J Tyler
Journal:  Conf Proc IEEE Eng Med Biol Soc       Date:  2009

10.  Chronic stability and selectivity of four-contact spiral nerve-cuff electrodes in stimulating the human femoral nerve.

Authors:  L E Fisher; D J Tyler; J S Anderson; R J Triolo
Journal:  J Neural Eng       Date:  2009-07-15       Impact factor: 5.379

View more