Literature DB >> 15191293

Scanning-slit photon counting x-ray imaging system using a microchannel plate detector.

Polad M Shikhaliev1, Tong Xu, Huy Le, Sabee Molloi.   

Abstract

An experimental prototype of a novel photon counting x-ray imaging system was evaluated. This system is based on an "edge-on" microchannel plate (MCP) detector and utilizes scanning slit imaging configuration. The detector is capable of photon counting, direct conversion, high spatial resolution, controllable physical charge amplification, quantum limited and scatter free operation. The detector provides a 60 mm wide field of view (FOV) and its count rate is 200 kHz for the entire FOV. The count rate of the current system is limited by the position encoding electronics, which has a single input for all events from the entire detector, and incorporates a single channel ADC with 1 micros conversion time. It is shown that the count rate can potentially be improved to clinically acceptable levels using multichannel application specific integrated circuit (ASIC) electronics and multi-slit image acquisition geometry. For a typical acquisition time used in this study, the image noise was measured to be less than the typically acceptable noise level for medical x-ray imaging. It is anticipated that the noise level will be also low after the implementation of the ASIC electronics. The quantum efficiency of the detector was measured to be 40%-56% for an energy range of 50-90 kVp for MCPs used in this study and can be improved to > 80% using MCPs with the optimized parameters. Images of resolution and anthropomorphic phantoms were acquired at an x-ray tube voltage of 50 kVp. The value of contrast transfer function for the detector was measured to be 0.5 at a spatial frequency of 5 lp/mm. The intrinsic spatial resolution of the system is 28 microm FWHM and was limited by the accuracy of the time-to-digital conversion of the position encoding electronics. Given the advantages of the edge-on MCP detector such as direct conversion and physical charge amplification, it can potentially be applied to mammography and chest radiography.

Mesh:

Substances:

Year:  2004        PMID: 15191293     DOI: 10.1118/1.1695651

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  6 in total

1.  A Prototype Micro-Angiographic Fluoroscope and Its Application in Animal Studies.

Authors:  Ye Wu; Stephen Rudin; Daniel R Bednarek
Journal:  Proc SPIE Int Soc Opt Eng       Date:  2005

2.  Cascaded systems analysis of photon counting detectors.

Authors:  J Xu; W Zbijewski; G Gang; J W Stayman; K Taguchi; M Lundqvist; E Fredenberg; J A Carrino; J H Siewerdsen
Journal:  Med Phys       Date:  2014-10       Impact factor: 4.071

3.  Radiation dose reduction using a CdZnTe-based computed tomography system: comparison to flat-panel detectors.

Authors:  Q Le Huy; Justin L Ducote; Sabee Molloi
Journal:  Med Phys       Date:  2010-03       Impact factor: 4.071

4.  Least squares parameter estimation methods for material decomposition with energy discriminating detectors.

Authors:  Q Le Huy; Sabee Molloi
Journal:  Med Phys       Date:  2011-01       Impact factor: 4.071

Review 5.  Breast cancer imaging: a perspective for the next decade.

Authors:  Andrew Karellas; Srinivasan Vedantham
Journal:  Med Phys       Date:  2008-11       Impact factor: 4.071

6.  Volumetric CT with sparse detector arrays (and application to Si-strip photon counters).

Authors:  A Sisniega; W Zbijewski; J W Stayman; J Xu; K Taguchi; E Fredenberg; Mats Lundqvist; J H Siewerdsen
Journal:  Phys Med Biol       Date:  2015-11-27       Impact factor: 3.609

  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.