In vitro comparison of six commercially available thermodilution cardiac output systems.

D J Powner, J V Snyder.   

Abstract

Six commercially available thermodilution cardiac output catheter-and-computer systems were evaluated over a variety of simulated clinical conditions using a fixed-output pulsatile-flow pumping unit. Simultaneous testing was performed at pumping rates of 60, 80, and 120 beats per minute and over an output range of 1.57 liters/min to 14.11 liters/min under normothermic, hyperthermic, and hypothermic conditions. Precision within each thermodilution system, as measured by the standard deviation (S.D.) of serial determinations, was acceptable with an overall S.D. range between 0.015 and 0.588 liters/min. When averaged over all experimental situations, the mean percent deviation of five or ten serial measurements was within 10 percent of the known pump output for all systems and maintained a fairly constant relationship above or below the pump output. In this in vitro test, iced injectate did not consistently improve precision or accuracy above that obtained with room-temperature injectate. Features other than performance should also be considered when selecting a particular unit to fulfill a clinical or research need.

Entities:  

Mesh:

Year:  1978        PMID: 150486

Source DB:  PubMed          Journal:  Med Instrum        ISSN: 0090-6689


  6 in total

1.  Reproducibility of measurements of cardiac output in newborn infants by Doppler ultrasound.

Authors:  I Hudson; A Houston; T Aitchison; B Holland; T Turner
Journal:  Arch Dis Child       Date:  1990-01       Impact factor: 3.791

2.  Pulsatile flow simulator for comparison of cardiac output measurements by electromagnetic flow meter and thermodilution.

Authors:  P J Jebson; W S Karkow
Journal:  J Clin Monit       Date:  1986-01

3.  Method of assessing the reproducibility of blood flow measurement: factors influencing the performance of thermodilution cardiac output computers.

Authors:  J D Mackenzie; N E Haites; J M Rawles
Journal:  Br Heart J       Date:  1986-01

4.  Cardiac output measurement by the thermodilution method: an in vitro test of accuracy of three commercially available automatic cardiac output computers.

Authors:  A Rubini; D Del Monte; V Catena; I Attar; M Cesaro; D Soranzo; G Rattazzi; G L Alati
Journal:  Intensive Care Med       Date:  1995-02       Impact factor: 17.440

5.  Non-invasive measurement of cardiac output: whole-body impedance cardiography in simultaneous comparison with thermodilution and direct oxygen Fick methods.

Authors:  T Kööbi; S Kaukinen; T Ahola; V M Turjanmaa
Journal:  Intensive Care Med       Date:  1997-11       Impact factor: 17.440

6.  Oxygen consumption in infants and children during heart catheterization.

Authors:  B P Lundell; M L Casas; C G Wallgren
Journal:  Pediatr Cardiol       Date:  1996 Jul-Aug       Impact factor: 1.655

  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.