Literature DB >> 15042289

Advantages and disadvantages of different nasal CPAP systems in newborns.

V Buettiker1, M I Hug, O Baenziger, C Meyer, B Frey.   

Abstract

OBJECTIVE: To compare three different systems of continuous positive airway pressure (CPAP): the naso-pharyngeal tube and two-prong systems in newborns, focusing on duration of CPAP, side effects and cost.
DESIGN: Randomized clinical study. PATIENTS: Between July 2000 and September 2001 newborns were randomized to three different CPAP systems. Forty infants in two weight groups (>2500 g and 1250-2500 g; 20 patients in each group) were included.
RESULTS: In the group >2500 g the median duration of CPAP was 1.1 days (range 0.25-14.3 days). The median time on a naso-pharyngeal CPAP was 1 day (range 0.25-14.3 days), on Hudson prongs 1.6 days (range 0.5-3.3 days) and on the Infant Flow system 0.7 days (range 0.3-13.6 days; p>0.05 for comparison between groups, Fisher's exact test). With naso-pharyngeal CPAP, 2 patients developed moderate nasal injuries. On Hudson, 2 patients developed moderate and three mild nasal injuries. One patient on the Infant Flow showed mild and one moderate nasal injuries. In the weight group 1250-2500 g the median duration of CPAP was 1.1 days (range 0.1-7.0 days). The median time on the naso-pharyngeal tube was 0.9 days (range 0.1-7 days), on Hudson prongs 1.1 days (range 0.7-6.6 days) and on the Infant Flow system 1.3 days (range 0.25-5.9 days; p>0.05 for comparison between groups, Fisher's exact test). With a naso-pharyngeal tube, one infant developed mild and one moderate nasal injuries. On Hudson prongs, two had moderate nasal injuries. On Infant Flow, one newborn showed a severe nasal injury and two mild injuries. None of the patients developed a pneumothorax.
CONCLUSION: The naso-pharyngeal tube is an easy, safe and economical CPAP system usable with every common ventilator. For very low birth weight newborns, a prong system may have advantages.

Entities:  

Mesh:

Year:  2004        PMID: 15042289     DOI: 10.1007/s00134-004-2267-8

Source DB:  PubMed          Journal:  Intensive Care Med        ISSN: 0342-4642            Impact factor:   17.440


  10 in total

Review 1.  Continuous distending pressure.

Authors:  C Morley
Journal:  Arch Dis Child Fetal Neonatal Ed       Date:  1999-09       Impact factor: 5.747

2.  In vitro comparison of nasal continuous positive airway pressure devices for neonates.

Authors:  A G De Paoli; C J Morley; P G Davis; R Lau; E Hingeley
Journal:  Arch Dis Child Fetal Neonatal Ed       Date:  2002-07       Impact factor: 5.747

3.  [Necrosis of the nasal columella secondary to nasal continuous positive airway pressure].

Authors:  V Millet; V Lacroze; J M Bartoli; D Unal
Journal:  Arch Pediatr       Date:  1997-05       Impact factor: 1.180

4.  Nasal deformities resulting from flow driver continuous positive airway pressure.

Authors:  N J Robertson; L S McCarthy; P A Hamilton; A L Moss
Journal:  Arch Dis Child Fetal Neonatal Ed       Date:  1996-11       Impact factor: 5.747

5.  Lung recruitment and breathing pattern during variable versus continuous flow nasal continuous positive airway pressure in premature infants: an evaluation of three devices.

Authors:  S E Courtney; K H Pyon; J G Saslow; G K Arnold; P B Pandit; R H Habib
Journal:  Pediatrics       Date:  2001-02       Impact factor: 7.124

6.  Thrombophlebitis after vasectomy.

Authors:  H J Roberts
Journal:  N Engl J Med       Date:  1971-06-10       Impact factor: 91.245

7.  A randomised control study comparing the Infant Flow Driver with nasal continuous positive airway pressure in preterm infants.

Authors:  M Mazzella; C Bellini; M G Calevo; F Campone; D Massocco; P Mezzano; E Zullino; F Scopesi; C Arioni; W Bonacci; G Serra
Journal:  Arch Dis Child Fetal Neonatal Ed       Date:  2001-09       Impact factor: 5.747

8.  A randomised controlled trial of two methods of delivering nasal continuous positive airway pressure after extubation to infants weighing less than 1000 g: binasal (Hudson) versus single nasal prongs.

Authors:  P Davis; M Davies; B Faber
Journal:  Arch Dis Child Fetal Neonatal Ed       Date:  2001-09       Impact factor: 5.747

9.  A randomized, controlled trial comparing two different continuous positive airway pressure systems for the successful extubation of extremely low birth weight infants.

Authors:  Beatrice M Stefanescu; W Paul Murphy; Brenda J Hansell; Mamta Fuloria; Timothy M Morgan; Judy L Aschner
Journal:  Pediatrics       Date:  2003-11       Impact factor: 7.124

10.  Effect of lung volume on expiratory time in the newborn infant.

Authors:  R J Martin; A Okken; P G Katona; M H Klaus
Journal:  J Appl Physiol Respir Environ Exerc Physiol       Date:  1978-07
  10 in total
  14 in total

Review 1.  Year in review in intensive care medicine, 2004. III. Outcome, ICU organisation, scoring, quality of life, ethics, psychological problems and communication in the ICU, immunity and hemodynamics during sepsis, pediatric and neonatal critical care, experimental studies.

Authors:  Peter Andrews; Elie Azoulay; Massimo Antonelli; Laurent Brochard; Christian Brun-Buisson; Geoffrey Dobb; Jean-Yves Fagon; Herwig Gerlach; Johan Groeneveld; Jordi Mancebo; Philipp Metnitz; Stefano Nava; Jerome Pugin; Michael Pinsky; Peter Radermacher; Christian Richard; Robert Tasker; Benoit Vallet
Journal:  Intensive Care Med       Date:  2005-02-18       Impact factor: 17.440

2.  Adaptive mechanical backup ventilation for preterm infants on respiratory assist modes - a pilot study.

Authors:  Susanne Herber-Jonat; Esther Rieger-Fackeldey; Helmut Hummler; Andreas Schulze
Journal:  Intensive Care Med       Date:  2006-01-24       Impact factor: 17.440

3.  Nasal masks or binasal prongs for delivering continuous positive airway pressure in preterm neonates-a randomised trial.

Authors:  Aparna Chandrasekaran; Anu Thukral; M Jeeva Sankar; Ramesh Agarwal; Vinod K Paul; Ashok K Deorari
Journal:  Eur J Pediatr       Date:  2017-01-13       Impact factor: 3.183

4.  Columella erosion secondary to nasal prongs in the neonate.

Authors:  Ewan Campbell; Marc D Pacifico
Journal:  BMJ Case Rep       Date:  2016-06-17

5.  Nasal injury and comfort with jet versus bubble continuous positive airway pressure delivery systems in preterm infants with respiratory distress.

Authors:  Jafar Khan; Venkataseshan Sundaram; Srinivas Murki; Anuj Bhatti; Shiv Sajan Saini; Praveen Kumar
Journal:  Eur J Pediatr       Date:  2017-09-15       Impact factor: 3.183

6.  Effects of nasal continuous positive airway pressure and cannula use in the neonatal intensive care unit setting.

Authors:  Kris R Jatana; Agnes Oplatek; Melanie Stein; Gary Phillips; D Richard Kang; Charles A Elmaraghy
Journal:  Arch Otolaryngol Head Neck Surg       Date:  2010-03

7.  Is volume and leak monitoring feasible during nasopharyngeal continuous positive airway pressure in neonates?

Authors:  Hendrik S Fischer; Charles C Roehr; Hans Proquitté; Hannes Hammer; Roland R Wauer; Gerd Schmalisch
Journal:  Intensive Care Med       Date:  2009-11       Impact factor: 17.440

8.  Hydrocolloid dressing in preventing nasal trauma secondary to nasal continuous positive airway pressure in preterm infants.

Authors:  Li-Hua Xie
Journal:  World J Emerg Med       Date:  2014

Review 9.  What is new in ventilation strategies for the neonate?

Authors:  Anne Greenough; Atul Sharma
Journal:  Eur J Pediatr       Date:  2007-06-02       Impact factor: 3.183

Review 10.  Devices and pressure sources for administration of nasal continuous positive airway pressure (NCPAP) in preterm neonates.

Authors:  A G De Paoli; P G Davis; B Faber; C J Morley
Journal:  Cochrane Database Syst Rev       Date:  2008-01-23
View more