Reducing the time period of steady state does not affect the accuracy of energy expenditure measurements by indirect calorimetry.

Marina M Reeves1, Peter S W Davies, Judith Bauer, Diana Battistutta.   

Abstract

Achievement of steady state during indirect calorimetry measurements of resting energy expenditure (REE) is necessary to reduce error and ensure accuracy in the measurement. Steady state is often defined as 5 consecutive min (5-min SS) during which oxygen consumption and carbon dioxide production vary by +/-10%. These criteria, however, are stringent and often difficult to satisfy. This study aimed to assess whether reducing the time period for steady state (4-min SS or 3-min SS) produced measurements of REE that were significantly different from 5-min SS. REE was measured with the use of open-circuit indirect calorimetry in 39 subjects, of whom only 21 (54%) met the 5-min SS criteria. In these 21 subjects, median biases in REE between 5-min SS and 4-min SS and between 5-min SS and 3-min SS were 0.1 and 0.01%, respectively. For individuals, 4-min SS measured REE within a clinically acceptable range of +/-2% of 5-min SS, whereas 3-min SS measured REE within a range of -2-3% of 5-min SS. Harris-Benedict prediction equations estimated REE for individuals within +/-20-30% of 5-min SS. Reducing the time period of steady state to 4 min produced measurements of REE for individuals that were within clinically acceptable, predetermined limits. The limits of agreement for 3-min SS fell outside the predefined limits of +/-2%; however, both 4-min SS and 3-min SS criteria greatly increased the proportion of subjects who satisfied steady state within smaller limits than would be achieved if relying on prediction equations.

Entities:  

Mesh:

Substances:

Year:  2004        PMID: 15020579     DOI: 10.1152/japplphysiol.01212.2003

Source DB:  PubMed          Journal:  J Appl Physiol (1985)        ISSN: 0161-7567


  20 in total

1.  Resting energy expenditure in male athletes with a spinal cord injury.

Authors:  Fiona E Pelly; Elizabeth M Broad; Natalie Stuart; Mark A Holmes
Journal:  J Spinal Cord Med       Date:  2017-05-04       Impact factor: 1.985

Review 2.  Clinical use of volumetric capnography in mechanically ventilated patients.

Authors:  Peter Kremeier; Stephan H Böhm; Gerardo Tusman
Journal:  J Clin Monit Comput       Date:  2019-05-31       Impact factor: 2.502

3.  Marital discord, past depression, and metabolic responses to high-fat meals: Interpersonal pathways to obesity.

Authors:  Janice K Kiecolt-Glaser; Lisa Jaremka; Rebecca Andridge; Juan Peng; Diane Habash; Christopher P Fagundes; Ronald Glaser; William B Malarkey; Martha A Belury
Journal:  Psychoneuroendocrinology       Date:  2014-12-03       Impact factor: 4.905

4.  Comparison of the Cosmed K4b(2) portable metabolic system in measuring steady-state walking energy expenditure.

Authors:  Jennifer A Schrack; Eleanor M Simonsick; Luigi Ferrucci
Journal:  PLoS One       Date:  2010-02-18       Impact factor: 3.240

5.  Abbreviated Steady State Intervals for Measuring Resting Energy Expenditure in Patients on Maintenance Hemodialysis.

Authors:  Laura A Olejnik; Emily N Peters; J Scott Parrott; Andrea F Marcus; Rebecca A Brody; Rosa K Hand; Justin J Fiutem; Laura D Byham-Gray
Journal:  JPEN J Parenter Enteral Nutr       Date:  2016-07-27       Impact factor: 4.016

6.  Daily stressors, past depression, and metabolic responses to high-fat meals: a novel path to obesity.

Authors:  Janice K Kiecolt-Glaser; Diane L Habash; Christopher P Fagundes; Rebecca Andridge; Juan Peng; William B Malarkey; Martha A Belury
Journal:  Biol Psychiatry       Date:  2014-07-14       Impact factor: 13.382

7.  [Foundations of Volumetric capnography : Principles of monitoring of metabolism and hemodynamics].

Authors:  S H Böhm; P Kremeier; G Tusman; D A Reuter; S Pulletz
Journal:  Anaesthesist       Date:  2020-04       Impact factor: 1.041

8.  Approximate Time to Steady-state Resting Energy Expenditure Using Indirect Calorimetry in Young, Healthy Adults.

Authors:  Collin J Popp; Jocelyn J Tisch; Kenan E Sakarcan; William C Bridges; Elliot D Jesch
Journal:  Front Nutr       Date:  2016-11-03

9.  Comparison of Energy Expenditure and Oxygen Consumption of Spontaneous Breathing Trial Conducted With and Without Automatic Tube Compensation.

Authors:  Alessandra Fabiane Lago; Elaine Cristina Goncalves; Elaine Caetano Silva; Mayra Goncalves Menegueti; Edson Antonio Nicolini; Maria Auxiliadora-Martins; Edson Zangiacomi Martinez; Ada Clarice Gastaldi; Anibal Basile-Filho
Journal:  J Clin Med Res       Date:  2015-07-24

10.  VO2 and VCO2 variabilities through indirect calorimetry instrumentation.

Authors:  Miguel Cadena-Méndez; Boris Escalante-Ramírez; Joaquín Azpiroz-Leehan; Oscar Infante-Vázquez
Journal:  Springerplus       Date:  2013-12-23
View more