Literature DB >> 14995935

Time-frequency balanced spectral entropy as a measure of anesthetic drug effect in central nervous system during sevoflurane, propofol, and thiopental anesthesia.

A Vakkuri1, A Yli-Hankala, P Talja, S Mustola, H Tolvanen-Laakso, T Sampson, H Viertiö-Oja.   

Abstract

BACKGROUND: Time-frequency balanced spectral entropy of electroencephalogram (EEG) and frontal electromyogram (FEMG) is a novel measure of hypnosis during anesthesia. Two Entropy parameters are described: Response entropy (RE) is calculated from EEG and FEMG; and State Entropy (SE) is calculated mainly from EEG. This study was performed to validate their performance during transition from consciousness to unconsciousness under different anesthetic agents.
METHODS: Response entropy, SE [S/5 Entropy Module, M-ENTROPY (later in text: Entropy), Datex-Ohmeda Division, Instrumentarium Corp., Helsinki, Finland] and BIS (BIS XP, A-2000, Aspect Medical Systems, Newton, MA) data were collected from 70 patients; 30 anesthetized with propofol 2 mg kg-1, 20 with sevoflurane inhalation, and 20 with thiopental 5 mg kg-1. Loss and regaining of consciousness (LOC, ROC) was tested every 10 s, and sensitivity, specificity, and prediction probability (Pk) were calculated. Behavior of the indices was studied.
RESULTS: Sensitivity, specificity, and Pk values for consciousness were high and similar for all indices. During regaining of consciousness after propofol bolus, RE, SE, and BIS values recovered by 81 +/- 22%, 75 +/- 26%, and 59 +/- 18% (mean +/- SD), respectively, from the minimum relative to their baseline. After thiopental bolus, RE, SE, and BIS values recovered by 86+/-21%, 88 +/- 13%, and 63 +/- 14%, respectively. The relative rise was higher in RE and SE compared with BIS (P < 0.01). During deep levels of hypnosis, RE and SE decreased monotonously as a function of burst suppression ratio, while BIS showed biphasic behavior. On average, RE indicated emergence from anesthesia 11 s earlier than SE, and 12.4 s earlier than BIS.
CONCLUSIONS: All indices, RE, SE, and BIS, distinguished excellently between conscious and unconscious states during propofol, sevoflurane, and thiopental anesthesia. During burst suppression, Entropy parameters RE and SE, but not BIS, behave monotonously. During regaining of consciousness after a thiopental or propofol bolus, RE and SE values recovered significantly closer to their baseline values than did BIS. Response entropy indicates emergence from anesthesia earlier than SE or BIS.

Entities:  

Mesh:

Substances:

Year:  2004        PMID: 14995935     DOI: 10.1111/j.0001-5172.2004.00323.x

Source DB:  PubMed          Journal:  Acta Anaesthesiol Scand        ISSN: 0001-5172            Impact factor:   2.105


  51 in total

1.  Regional entropy of functional imaging signals varies differently in sensory and cognitive systems during propofol-modulated loss and return of behavioral responsiveness.

Authors:  Xiaolin Liu; Kathryn K Lauer; B Douglas Ward; Christopher J Roberts; Suyan Liu; Suneeta Gollapudy; Robert Rohloff; William Gross; Zhan Xu; Shanshan Chen; Lubin Wang; Zheng Yang; Shi-Jiang Li; Jeffrey R Binder; Anthony G Hudetz
Journal:  Brain Imaging Behav       Date:  2019-04       Impact factor: 3.978

2.  The effect of addition of nitrous oxide to a sevoflurane anesthetic on BIS, PSI, and entropy.

Authors:  Roy G Soto; Robert A Smith; Amy L Zaccaria; Rafael V Miguel
Journal:  J Clin Monit Comput       Date:  2006-06-21       Impact factor: 2.502

3.  Comparing Entropy and the Bispectral index with the Ramsay score in sedated ICU patients.

Authors:  Carmen Hernández-Gancedo; David Pestaña; Hanna Pérez-Chrzanowska; Elena Martinez-Casanova; Antonio Criado
Journal:  J Clin Monit Comput       Date:  2007-08-16       Impact factor: 2.502

Review 4.  [Measurement of the depth of anaesthesia].

Authors:  G N Schmidt; J Müller; P Bischoff
Journal:  Anaesthesist       Date:  2008-01       Impact factor: 1.041

5.  BIS values during resuscitation: the role of the suppression ratio (case report).

Authors:  Agnes Declerck; Ngai Liu; Thierry Chazot; Marc Fischler
Journal:  J Clin Monit Comput       Date:  2009-08-21       Impact factor: 2.502

6.  Response entropy is more reactive than bispectral index during laparoscopic gastric banding.

Authors:  James Feld; William E Hoffman
Journal:  J Clin Monit Comput       Date:  2006-07-04       Impact factor: 2.502

7.  Comparison of spectral entropy and BIS VISTA™ monitor during general anesthesia for cardiac surgery.

Authors:  Tadeusz Musialowicz; Pasi Lahtinen; Otto Pitkänen; Jouni Kurola; Ilkka Parviainen
Journal:  J Clin Monit Comput       Date:  2011-04-22       Impact factor: 2.502

8.  What is New about Neuroanaesthesia?

Authors:  G S Umamaheswara Rao
Journal:  Indian J Anaesth       Date:  2009-08

9.  Performance of Surgical Stress Index during Sevoflurane-Fentanyl and Isoflurane-Fentanyl Anesthesia.

Authors:  S Mustola; T Parkkari; K Uutela; M Huiku; M Kymäläinen; J Toivonen
Journal:  Anesthesiol Res Pract       Date:  2010-04-06

10.  Facial muscle activity, Response Entropy, and State Entropy indices during noxious stimuli in propofol-nitrous oxide or propofol-nitrous oxide-remifentanil anaesthesia without neuromuscular block.

Authors:  A J Aho; A Yli-Hankala; L-P Lyytikäinen; V Jäntti
Journal:  Br J Anaesth       Date:  2008-12-25       Impact factor: 9.166