The effect of addition of nitrous oxide to a sevoflurane anesthetic on BIS, PSI, and entropy.

Roy G Soto1, Robert A Smith, Amy L Zaccaria, Rafael V Miguel.   

Abstract

OBJECTIVE: N(2)O is a commonly used anesthetic that has amnestic and analgesic properties. Recently, devices that estimate depth of consciousness have been introduced in an attempt to better titrate anesthesia, however the effect of N(2)O on these monitors is unclear.
METHODS: General anesthesia was induced and titrated to maintain normal blood pressure and pulse in healthy adults. Data were collected in three 10 minute intervals (Sevo, Sevo + N(2)O, Sevo). In Phase A, sevoflurane concentration was held constant during the N(2)O trial in 60 subjects monitored with either BIS, PSI, or Entropy. In Phase B, sevoflurane concentration was reduced as N(2)O was added, maintaining a constant overall "MAC" in 20 subjects monitored concurrently with BIS and Entropy. Sample size for both phases was designed to detect a 10 unit change in measure of processed EEG with alpha = .05 and statistical power = .80.
RESULTS: In Phase A, supplementing sevoflurane with > 65% N(2)O increased MAC from 1.3 +/- 0.05 to 2.2 +/- 0.10, but did not significantly alter BIS nor PSI (p-value for differential MAC is < 0.05). Entropy, however, dropped significantly, with a change in state entropy (SE) from 31.1 +/- 7.3 to 18.9 +/- 3.7 and a corresponding rise when N(2)O was discontinued. In Phase B, supplementing sevoflurane with > 65% N(2)O with a concomitant reduction in sevoflurane resulted in an increase in both BIS (from 34 +/- 5 to 53.9 +/- 11.5) and SE (from 32 +/- 8.2 to 55.4 +/- 21.3).
CONCLUSION: Supplementing sevoflurane with > 65% N(2)O did not result in a significant change in either BIS or PSI when sevoflurane concentration was kept constant. Entropy, however, significantly decreased as anesthetic depth increased. When sevoflurane concentration was reduced during N(2)O administration, both BIS and Entropy rose despite maintenance of anesthetic depth, indicating a variable concentration effect between volatiles and N(2)O.

Entities:  

Mesh:

Substances:

Year:  2006        PMID: 16794892     DOI: 10.1007/s10877-006-9009-0

Source DB:  PubMed          Journal:  J Clin Monit Comput        ISSN: 1387-1307            Impact factor:   2.502


  19 in total

Review 1.  Development and clinical application of electroencephalographic bispectrum monitoring.

Authors:  J W Johansen; P S Sebel
Journal:  Anesthesiology       Date:  2000-11       Impact factor: 7.892

2.  Time-frequency balanced spectral entropy as a measure of anesthetic drug effect in central nervous system during sevoflurane, propofol, and thiopental anesthesia.

Authors:  A Vakkuri; A Yli-Hankala; P Talja; S Mustola; H Tolvanen-Laakso; T Sampson; H Viertiö-Oja
Journal:  Acta Anaesthesiol Scand       Date:  2004-02       Impact factor: 2.105

3.  Minimum alveolar concentrations (MAC) of isoflurande with and without nitrous oxide in patients of various ages.

Authors:  W C Stevens; W M Dolan; R T Gibbons; A White; E I Eger; R D Miller; R H DeJong; R M Elashoff
Journal:  Anesthesiology       Date:  1975-02       Impact factor: 7.892

4.  Pharmacodynamic interaction of nitrous oxide with sevoflurane, desflurane, isoflurane and enflurane in surgical patients: measurements by effects on EEG median power frequency.

Authors:  H Röpcke; S Wirz; T Bouillon; J Bruhn; A Hoeft
Journal:  Eur J Anaesthesiol       Date:  2001-07       Impact factor: 4.330

5.  Effect on nitrous oxide on the anaesthetic requirement of enflurane.

Authors:  G Torri; G Damia; M L Fabiani
Journal:  Br J Anaesth       Date:  1974-07       Impact factor: 9.166

6.  A multicenter study of bispectral electroencephalogram analysis for monitoring anesthetic effect.

Authors:  P S Sebel; E Lang; I J Rampil; P F White; R Cork; M Jopling; N T Smith; P S Glass; P Manberg
Journal:  Anesth Analg       Date:  1997-04       Impact factor: 5.108

7.  Bispectral EEG index during nitrous oxide administration.

Authors:  I J Rampil; J S Kim; R Lenhardt; C Negishi; D I Sessler
Journal:  Anesthesiology       Date:  1998-09       Impact factor: 7.892

8.  The Patient State Index as an indicator of the level of hypnosis under general anaesthesia.

Authors:  L S Prichep; L D Gugino; E R John; R J Chabot; B Howard; H Merkin; M L Tom; S Wolter; L Rausch; W J Kox
Journal:  Br J Anaesth       Date:  2004-01-22       Impact factor: 9.166

9.  Interaction of isoflurane and nitrous oxide combinations similar for median electroencephalographic frequency and clinical anesthesia.

Authors:  H Röpcke; H Schwilden
Journal:  Anesthesiology       Date:  1996-04       Impact factor: 7.892

10.  No correlation between quantitative electroencephalographic measurements and movement response to noxious stimuli during isoflurane anesthesia in rats.

Authors:  I J Rampil; M J Laster
Journal:  Anesthesiology       Date:  1992-11       Impact factor: 7.892

View more
  4 in total

Review 1.  [Measurement of the depth of anaesthesia].

Authors:  G N Schmidt; J Müller; P Bischoff
Journal:  Anaesthesist       Date:  2008-01       Impact factor: 1.041

2.  Offline comparison of processed electroencephalogram monitors for anaesthetic-induced electroencephalogram changes in older adults.

Authors:  Sarah L Eagleman; Caitlin M Drover; Xi Li; M Bruce MacIver; David R Drover
Journal:  Br J Anaesth       Date:  2021-02-25       Impact factor: 9.166

3.  Remifentanil prevents tourniquet-induced arterial pressure increase in elderly orthopedic patients under sevoflurane/N2O general anesthesia.

Authors:  Jun-Young Jung; Jin-Hee Han; Jae-Woo Yi; Jong-Man Kang
Journal:  Int J Med Sci       Date:  2012-06-08       Impact factor: 3.738

4.  PRISMA: accuracy of response entropy and bispectral index to predict the transition of consciousness during sevoflurane anesthesia: A systematic review and meta-analysis.

Authors:  Tao Liang; Fan Wu; Baoguo Wang; Feng Mu
Journal:  Medicine (Baltimore)       Date:  2021-04-30       Impact factor: 1.817

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.