Literature DB >> 21512777

Comparison of spectral entropy and BIS VISTA™ monitor during general anesthesia for cardiac surgery.

Tadeusz Musialowicz1, Pasi Lahtinen, Otto Pitkänen, Jouni Kurola, Ilkka Parviainen.   

Abstract

OBJECTIVES: We compared the primary metrics of the Spectral entropy M-ENTROPY™ module and BIS VISTA™ monitor-i.e., bispectral index (BIS), state entropy (SE), and response entropy (RE) in terms of agreement and correlation during general anesthesia for cardiac surgery. We also evaluated responsiveness of electroencephalogram (EEG)-based and hemodynamic parameters to surgical noxious stimulation, skin incision, and sternotomy, hypothesizing that RE would be a better responsiveness predictor.
METHODS: BIS and entropy sensors were applied before anesthesia induction in 32 patients having elective cardiac surgery. Total intravenous anesthesia was standardized and guided by the BIS index with neuromuscular blockade tested with train-of-four monitoring. Parameters included SE, RE, BIS, forehead electromyography (EMG), and hemodynamic variables. Time points for analyzing BIS, entropy, and hemodynamic values were 1 min before and after: anesthesia induction, intubation, skin incision, sternotomy, cannulation of the aorta, cardiopulmonary bypass (CPB), cross-clamping the aorta, de-clamping the aorta, and end of CPB; also after starting the re-warming phase and at 10, 20, 30, and 40 min following.
RESULTS: The mean difference between BIS and SE (Bland-Altman) was 2.14 (+16/- 11; 95% CI 1.59-2.67), and between BIS and RE it was 0.02 (+14/- 14; 95% CI 0.01-0.06). BIS and SE (r(2) = 0.66; P = 0.001) and BIS and RE (r(2) = 0.7; P = 0.001) were closely correlated (Pearson's). EEG parameters, EMG values, and systolic blood pressure significantly increased after skin incision, and sternotomy. The effect of surgical stimulation (Cohen's d) was highest for RE after skin incision (-0.71; P = 0.0001) and sternotomy (-0.94; P = 0.0001).
CONCLUSION: Agreement was poor between the BIS index measured by BIS VISTA™ and SE values at critical anesthesia time points in patients undergoing cardiac surgery. RE was a good predictor of arousal after surgical stimulation regardless of the surgical level of muscle relaxation. Index differences most likely resulted from different algorithms for calculating consciousness level.

Entities:  

Mesh:

Year:  2011        PMID: 21512777     DOI: 10.1007/s10877-011-9280-6

Source DB:  PubMed          Journal:  J Clin Monit Comput        ISSN: 1387-1307            Impact factor:   2.502


  28 in total

1.  Evaluation of the electroencephalographic bispectral index during fentanyl-midazolam anaesthesia for cardiac surgery. Does it predict haemodynamic responses during endotracheal intubation and sternotomy?

Authors:  J J Driessen; J B Harbers; J van Egmond; L H Booij
Journal:  Eur J Anaesthesiol       Date:  1999-09       Impact factor: 4.330

2.  Description of the Entropy algorithm as applied in the Datex-Ohmeda S/5 Entropy Module.

Authors:  H Viertiö-Oja; V Maja; M Särkelä; P Talja; N Tenkanen; H Tolvanen-Laakso; M Paloheimo; A Vakkuri; A Yli-Hankala; P Meriläinen
Journal:  Acta Anaesthesiol Scand       Date:  2004-02       Impact factor: 2.105

3.  The effect of skin incision on the electroencephalogram during general anesthesia maintained with propofol or desflurane.

Authors:  James W Sleigh; Kate Leslie; Logan Voss
Journal:  J Clin Monit Comput       Date:  2010-08-01       Impact factor: 2.502

4.  Effect of an intubation dose of rocuronium on Spectral Entropy and Bispectral Index responses to laryngoscopy during propofol anaesthesia.

Authors:  P Hans; J Giwer; J F Brichant; P-Y Dewandre; V Bonhomme
Journal:  Br J Anaesth       Date:  2006-09-21       Impact factor: 9.166

5.  Relationship between Bispectral Index, electroencephalographic state entropy and effect-site EC50 for propofol at different clinical endpoints.

Authors:  M Iannuzzi; E Iannuzzi; F Rossi; L Berrino; M Chiefari
Journal:  Br J Anaesth       Date:  2005-01-21       Impact factor: 9.166

6.  Recovery from paralysis with succinylcholine increased Response entropy and EMG but not State entropy.

Authors:  Verna L Baughman; William E Hoffman; Heidi M Koenig; Peggy L Wheeler; Ranga C Ananda; Mathew Wang
Journal:  J Clin Monit Comput       Date:  2005-06       Impact factor: 2.502

7.  Response entropy increases during painful stimulation.

Authors:  Peggy Wheeler; William E Hoffman; Verna L Baughman; Heidi Koenig
Journal:  J Neurosurg Anesthesiol       Date:  2005-04       Impact factor: 3.956

8.  Entropy indices vs the bispectral index for estimating nociception during sevoflurane anaesthesia.

Authors:  I Takamatsu; M Ozaki; T Kazama
Journal:  Br J Anaesth       Date:  2006-03-10       Impact factor: 9.166

9.  Comparison of spectral entropy and bispectral index electroencephalography in coronary artery bypass graft surgery.

Authors:  Werner Baulig; Burkhardt Seifert; Edith R Schmid; Urs Schwarz
Journal:  J Cardiothorac Vasc Anesth       Date:  2009-11-27       Impact factor: 2.628

10.  Patients' conscious recollections from cardiac anesthesia.

Authors:  Seppo O-V Ranta; Pirkko Herranen; Markku Hynynen
Journal:  J Cardiothorac Vasc Anesth       Date:  2002-08       Impact factor: 2.628

View more
  8 in total

1.  Offline comparison of processed electroencephalogram monitors for anaesthetic-induced electroencephalogram changes in older adults.

Authors:  Sarah L Eagleman; Caitlin M Drover; Xi Li; M Bruce MacIver; David R Drover
Journal:  Br J Anaesth       Date:  2021-02-25       Impact factor: 9.166

2.  Decrease in spectral entropy by low tidal volume ventilation-associated severe hypercapnia: a case report.

Authors:  Junggun Ann; Sung Mee Jung; Sang-Jin Park
Journal:  Korean J Anesthesiol       Date:  2016-10-25

3.  Comparison of electroencephalogram between propofol- and thiopental-induced anesthesia for awareness risk in pregnant women.

Authors:  Hee-Sun Park; Yeon-Su Kim; Sung-Hoon Kim; A-Rom Jeon; Seong-Eun Kim; Woo-Jong Choi
Journal:  Sci Rep       Date:  2020-04-10       Impact factor: 4.379

4.  Monitoring of anesthetic depth and EEG band power using phase lag entropy during propofol anesthesia.

Authors:  Hye Won Shin; Hyun Jung Kim; Yoo Kyung Jang; Hae Sun You; Hyub Huh; Yoon Ji Choi; Seung Uk Choi; Ji Su Hong
Journal:  BMC Anesthesiol       Date:  2020-02-26       Impact factor: 2.217

5.  Impact of General Anesthesia Guided by State Entropy (SE) and Response Entropy (RE) on Perioperative Stability in Elective Laparoscopic Cholecystectomy Patients-A Prospective Observational Randomized Monocentric Study.

Authors:  Anca Raluca Dinu; Alexandru Florin Rogobete; Sonia Elena Popovici; Ovidiu Horea Bedreag; Marius Papurica; Corina Maria Dumbuleu; Raluca Ramona Velovan; Daiana Toma; Corina Maria Georgescu; Lavinia Ioana Trache; Claudiu Barsac; Loredana Luca; Bettina Buzzi; Andra Maghiar; Mihai Alexandru Sandesc; Samir Rimawi; Madalin Marian Vaduva; Lavinia Melania Bratu; Paul Manuel Luminosu; Dorel Sandesc
Journal:  Entropy (Basel)       Date:  2020-03-19       Impact factor: 2.524

6.  Comparative evaluation of the bispectral index (BIS) and BISpro during propofol anaesthesia.

Authors:  Zhitao Li; Jiachun Cai; Jiexiong Li; Xianghui Xu; Limin Zheng
Journal:  J Int Med Res       Date:  2021-04       Impact factor: 1.671

7.  Variations in Values of State, Response Entropy and Haemodynamic Parameters Associated with Development of Different Epileptiform Patterns during Volatile Induction of General Anaesthesia with Two Different Anaesthetic Regimens Using Sevoflurane in Comparison with Intravenous Induct: A Comparative Study.

Authors:  Michał Stasiowski; Anna Duława; Izabela Szumera; Radosław Marciniak; Ewa Niewiadomska; Wojciech Kaspera; Lech Krawczyk; Piotr Ładziński; Beniamin Oskar Grabarek; Przemysław Jałowiecki
Journal:  Brain Sci       Date:  2020-06-12

8.  Adequacy of Anesthesia Guidance for Colonoscopy Procedures.

Authors:  Michał Jan Stasiowski; Małgorzata Starzewska; Ewa Niewiadomska; Seweryn Król; Kaja Marczak; Jakub Żak; Aleksandra Pluta; Jerzy Eszyk; Beniamin Oskar Grabarek; Izabela Szumera; Michał Nycz; Anna Missir; Lech Krawczyk; Przemysław Jałowiecki
Journal:  Pharmaceuticals (Basel)       Date:  2021-05-14
  8 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.