Image brightening in samples of high dielectric constant.

James Tropp1.   

Abstract

An analytic solution is given for the electromagnetic problem of a lossy dielectric cylinder of infinite length, irradiated by a circularly polarized radiofrequency (RF) magnetic field; the NMR-active components of the field inside the cylinder are projected out by transforming the RF Hamiltonian to the rotating frame and retaining only those terms independent of time; it is noted that the resulting cartesian field components are required to be real. The squared magnitude of the NMR-active fields are then used to calculate the gradient-recalled images of the cylinder, for small tip angles of the magnetization; and the result is shown to predict almost quantitatively the intensity patterns of experimental proton images at 3.0 and 4.0T, in a cylindrical phantom of radius 9.25cm, filled with 0.05M aqueous NaCl. In particular, the artifactual brightening at the center of the recorded image is convincingly reproduced in a simulation, whose underlying model excludes wave propagation along the direction of the cylinder axis. Formation of the artifact is explained in terms of the focussing of the RF magnetic field at the center of the cylinder, as illustrated by contour plots showing the time evolution of the rotating flux. An extended electromagnetic model--having the dielectric cylinder enclosed in a long, shielded volume resonator (e.g., of bird cage type)--is then sketched. The mathematical details appear in Appendix A; and the simulated images are shown to be virtually indistinguishable from those of the simpler original model. The theory of the Q, or quality factor, of the dielectric cylinder--considered itself as a resonant object--is developed for the enclosed cylinder model, where flux containment by the shield permits an unambiguous treatment of both the stored energy and the radiative losses. This is extended to treat the Q of a lossy dielectric sphere without shielding. Further plots of flux contours are given for the sphere, excited at 208 MHz with a uniform circularly polarized field, as well as by a surface coil, and for the enclosed cylinder in the range 140-160 MHz. It is then argued that the center brightening artifacts in magnetic resonance images are due to the underdamped dielectric resonance of the sample, i.e., at Q >0.5, while the overdamped condition, Q < 0.5, leads to exclusion of flux from the center, i.e., to the classic skin effect. The term "dielectric resonance" is shown to require careful interpretation for mixed-mode excitation, such as occurs with a surface coil. An extended reciprocity formula for NMR reception, valid for an arbitrary electromagnetic Green's function, is also given in Appendix B.

Entities:  

Mesh:

Year:  2004        PMID: 14987593     DOI: 10.1016/j.jmr.2003.11.003

Source DB:  PubMed          Journal:  J Magn Reson        ISSN: 1090-7807            Impact factor:   2.229


  16 in total

1.  Dual optimization method of radiofrequency and quasistatic field simulations for reduction of eddy currents generated on 7T radiofrequency coil shielding.

Authors:  Yujuan Zhao; Tiejun Zhao; Shailesh B Raval; Narayanan Krishnamurthy; Hai Zheng; Chad T Harris; William B Handler; Blaine A Chronik; Tamer S Ibrahim
Journal:  Magn Reson Med       Date:  2014-11-03       Impact factor: 4.668

2.  Multi-center reproducibility of structural, diffusion tensor, and resting state functional magnetic resonance imaging measures.

Authors:  S Deprez; Michiel B de Ruiter; S Bogaert; R Peeters; J Belderbos; D De Ruysscher; S Schagen; S Sunaert; P Pullens; E Achten
Journal:  Neuroradiology       Date:  2018-04-14       Impact factor: 2.804

3.  A 128-channel receive-only cardiac coil for highly accelerated cardiac MRI at 3 Tesla.

Authors:  Melanie Schmitt; Andreas Potthast; David E Sosnovik; Jonathan R Polimeni; Graham C Wiggins; Christina Triantafyllou; Lawrence L Wald
Journal:  Magn Reson Med       Date:  2008-06       Impact factor: 4.668

4.  Radiofrequency transmit calibration: A multi-center evaluation of vendor-provided radiofrequency transmit mapping methods.

Authors:  Yannick Bliesener; Xinran Zhong; Yi Guo; Michael Boss; Ryan Bosca; Hendrik Laue; Caroline Chung; Kyunghyun Sung; Krishna S Nayak
Journal:  Med Phys       Date:  2019-04-15       Impact factor: 4.071

5.  A highly decoupled transmit-receive array design with triangular elements at 7T.

Authors:  Gang Chen; Bei Zhang; Martijn A Cloos; Daniel K Sodickson; Graham C Wiggins
Journal:  Magn Reson Med       Date:  2018-03-23       Impact factor: 4.668

6.  Dependence of B1+ and B1- Field Patterns of Surface Coils on the Electrical Properties of the Sample and the MR Operating Frequency.

Authors:  Manushka V Vaidya; Christopher M Collins; Daniel K Sodickson; Ryan Brown; Graham C Wiggins; Riccardo Lattanzi
Journal:  Concepts Magn Reson Part B Magn Reson Eng       Date:  2016-02-04       Impact factor: 1.176

7.  Ideal current patterns yielding optimal signal-to-noise ratio and specific absorption rate in magnetic resonance imaging: computational methods and physical insights.

Authors:  Riccardo Lattanzi; Daniel K Sodickson
Journal:  Magn Reson Med       Date:  2011-11-29       Impact factor: 4.668

8.  Effect of MR contrast agents on quantitative accuracy of PET in combined whole-body PET/MR imaging.

Authors:  Cristina Lois; Ilja Bezrukov; Holger Schmidt; Nina Schwenzer; Matthias K Werner; Jürgen Kupferschläger; Thomas Beyer
Journal:  Eur J Nucl Med Mol Imaging       Date:  2012-08-14       Impact factor: 9.236

9.  Performance evaluation of a 32-element head array with respect to the ultimate intrinsic SNR.

Authors:  Riccardo Lattanzi; Aaron K Grant; Jonathan R Polimeni; Michael A Ohliger; Graham C Wiggins; Lawrence L Wald; Daniel K Sodickson
Journal:  NMR Biomed       Date:  2010-02       Impact factor: 4.044

10.  Directly detected (55)Mn MRI: application to phantoms for human hyperpolarized (13)C MRI development.

Authors:  Cornelius von Morze; Lucas Carvajal; Galen D Reed; Christine Leon Swisher; James Tropp; Daniel B Vigneron
Journal:  Magn Reson Imaging       Date:  2014-08-29       Impact factor: 2.546

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.