The relationship between standard automated perimetry and GDx VCC measurements.

Nicolaas J Reus1, Hans G Lemij.   

Abstract

PURPOSE: To investigate the relationship between retinal light sensitivity measured with standard automated perimetry (SAP) and retardation of the peripapillary retinal nerve fiber layer (RNFL) measured with the GDx VCC (Laser Diagnostic Technologies, Inc., San Diego, CA).
METHODS: Forty-seven healthy subjects and 101 patients with glaucoma were examined with SAP and with the commercially available scanning laser polarimeter GDx VCC, with automated individualized compensation of anterior segment birefringence. Individual visual field test points and peripapillary RNFL retardation measurements were grouped into six corresponding sectors. The correlation between perimetry and GDx VCC measurements was determined, and the relationship between RNFL retardation and perimetry, expressed both in the standard decibel scale and in an unlogged scale, was described with linear regression analysis.
RESULTS: A statistically significant correlation was found in most sectors between perimetry and GDx VCC measurements in patients with glaucoma, but not in healthy subjects. A linear relationship was found between the unlogged sensitivities and GDx VCC measurements for the superotemporal and inferotemporal sectors. In the decibel scale, this relationship was curvilinear.
CONCLUSIONS: GDx VCC measurements of the peripapillary RNFL relate well with functional loss in glaucoma. Based on the observed relationships between function and structure, patients with mild to moderate visual field loss in glaucoma may be better monitored with the GDx VCC and patients who have severe loss, with SAP.

Entities:  

Mesh:

Year:  2004        PMID: 14985299     DOI: 10.1167/iovs.03-0646

Source DB:  PubMed          Journal:  Invest Ophthalmol Vis Sci        ISSN: 0146-0404            Impact factor:   4.799


  30 in total

1.  GDx VCC measurements and blue-on-yellow perimetry in ocular hypertension.

Authors:  Yisheng Zhong; Liping Chen; Yu Cheng; Fangzhi Yu
Journal:  Lasers Med Sci       Date:  2012-02-03       Impact factor: 3.161

2.  Longitudinal relationship between retinal nerve fiber layer thickness parameters assessed by scanning laser polarimetry (GDxVCC) and visual field in glaucoma.

Authors:  Kenichi Makabe; Kazuo Takei; Tetsuro Oshika
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2012-04       Impact factor: 3.117

3.  Development and evaluation of a linear staircase strategy for the measurement of perimetric sensitivity.

Authors:  Rizwan Malik; William H Swanson; David F Garway-Heath
Journal:  Vision Res       Date:  2006-06-09       Impact factor: 1.886

4.  Ganglion cell loss and age-related visual loss: a cortical pooling analysis.

Authors:  Pauline M Pearson; Laura A Schmidt; Emily Ly-Schroeder; William H Swanson
Journal:  Optom Vis Sci       Date:  2006-07       Impact factor: 1.973

5.  Scanning laser polarimetry with variable and enhanced corneal compensation in normal and glaucomatous eyes.

Authors:  Mitra Sehi; Delia C Guaqueta; William J Feuer; David S Greenfield
Journal:  Am J Ophthalmol       Date:  2006-10-25       Impact factor: 5.258

6.  GDx-VCC performance in discriminating normal from glaucomatous eyes with early visual field loss.

Authors:  Stefano Da Pozzo; Mirko Fuser; Odilla Vattovani; Giuseppe Di Stefano; Giuseppe Ravalico
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2005-11-16       Impact factor: 3.117

7.  Retinal nerve fiber layer thickness changes following optic neuritis caused by multiple sclerosis.

Authors:  Toru Urano; Toyoaki Matsuura; Eiichi Yukawa; Mikki Arai; Yoshiaki Hara; Ryoji Yamakawa
Journal:  Jpn J Ophthalmol       Date:  2011-02-18       Impact factor: 2.447

8.  Scanning laser polarimetry with variable corneal compensation and detection of glaucomatous optic neuropathy.

Authors:  Stefano Da Pozzo; Pierluigi Iacono; Roberta Marchesan; Anna Fantin; Giuseppe Ravalico
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2005-03-09       Impact factor: 3.117

9.  Comparison of Spectralis-OCT, GDxVCC and GDxECC in assessing retinal nerve fiber layer (RNFL) in glaucomatous patients.

Authors:  Maurice Schallenberg; Dirk Dekowski; Stephan Kremmer; J Michael Selbach; Klaus-Peter Steuhl
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2012-12-19       Impact factor: 3.117

10.  Retinal nerve fibre layer assessment in myopic glaucomatous eyes: comparison of GDx variable corneal compensation with GDx enhanced corneal compensation.

Authors:  S Morishita; T Tanabe; S Yu; M Hangai; T Ojima; H Aikawa; N Yoshimura
Journal:  Br J Ophthalmol       Date:  2008-07-29       Impact factor: 4.638

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.