Literature DB >> 1497153

Length-based endotracheal tube and emergency equipment in pediatrics.

R C Luten1, R L Wears, J Broselow, A Zaritsky, T M Barnett, T Lee, A Bailey, R Vally, R Brown, B Rosenthal.   

Abstract

HYPOTHESIS: Pediatric endotracheal (ET) tubes can be accurately selected based on body length using a specialized emergency tape. POPULATION: Derivation set: Two hundred five children undergoing elective surgery. Validation set: Two hundred thirteen children undergoing elective surgery. Each child served as his or her own control.
METHODS: Derivation phase: Two hundred twenty-one children undergoing ET intubation for elective surgery had their body length and leak pressures measured. The 205 children who had leak pressures between 10 and 40 cm H2O constituted the derivation set. The body length for a given ET tube size was derived from the interquartile range of patient lengths in this derivation set. Sizes for other resuscitative equipment items were chosen by a panel of experts using a modified Delphi technique. This information was placed by length on a color-coded tape. Validation phase: The tape was validated by using it to select ET tube size in another group of 203 children undergoing elective surgery. Criteria for acceptable fit in this group included leak pressure as above and the anesthesiologists' decision to accept the tube size or to retintubate. In the validation phase, length-based ET tube selections were compared with age-based rules: (age + 16)/4, and (age + 18)/4.
RESULTS: The tape selected the appropriate ET tube size by leak pressure criterion in 77% of the cases and was within +/- 0.5 mm of the "correct" size 99% of the time. This was significantly better (P less than .005) than two widely used age-based rules, which gave the correct initial size in only 47% and 9% of these cases, and were within +/- 0.5 mm for 86% and 59%. The anesthesiologists chose to continue with the tape-sized tube rather than to reintubate in 89% of cases.
CONCLUSION: A system for length-based selection of emergency equipment represents a significant adjunct to emergency physicians and paramedics who must deal with critically ill children. Length-based ET tube selection is clearly superior to age-based rules, which are difficult to remember and require accurate estimation of a patient's age.

Entities:  

Mesh:

Year:  1992        PMID: 1497153     DOI: 10.1016/s0196-0644(05)82924-5

Source DB:  PubMed          Journal:  Ann Emerg Med        ISSN: 0196-0644            Impact factor:   5.721


  20 in total

1.  Prospective comparison of live evaluation and video review in the evaluation of operator performance in a pediatric emergency airway simulation.

Authors:  Joseph B House; Suzanne Dooley-Hash; Terry Kowalenko; Athina Sikavitsas; Desiree M Seeyave; John G Younger; Stanley J Hamstra; Michele M Nypaver
Journal:  J Grad Med Educ       Date:  2012-09

2.  Relationships of pediatric anthropometrics for CT protocol selection.

Authors:  Grace S Phillips; Arta-Luana Stanescu; Adam M Alessio
Journal:  AJR Am J Roentgenol       Date:  2014-07       Impact factor: 3.959

3.  The Accuracy of the Broselow™ Pediatric Emergency Tape for Weight Estimation in an Omani Paediatric Population.

Authors:  Azher A Al-Busaidi; Lakshmanan Jeyaseelan; Hilal M Al-Barwani
Journal:  Sultan Qaboos Univ Med J       Date:  2017-06-20

Review 4.  [Tools for drug dosing in life-threatening pediatric emergencies].

Authors:  C G Erker; M Santamaria; M Möllmann
Journal:  Anaesthesist       Date:  2012-11       Impact factor: 1.041

5.  A smartphone application to determine body length for body weight estimation in children: a prospective clinical trial.

Authors:  Oliver Wetzel; Alexander R Schmidt; Michelle Seiler; Davide Scaramuzza; Burkhardt Seifert; Donat R Spahn; Philipp Stein
Journal:  J Clin Monit Comput       Date:  2017-06-28       Impact factor: 2.502

6.  Weight estimation in resuscitation: is the current formula still valid?

Authors:  Mark Luscombe; Ben Owens
Journal:  Arch Dis Child       Date:  2007-01-09       Impact factor: 3.791

7.  Prediction of optimal endotracheal tube cuff volume from tracheal diameter and from patient height and age: a prospective cohort trial.

Authors:  Masayuki Shibasaki; Yasufumi Nakajima; Nobuaki Shime; Teiji Sawa; Daniel I Sessler
Journal:  J Anesth       Date:  2012-03-22       Impact factor: 2.078

Review 8.  Principles of pediatric emergency care.

Authors:  Jochen Meyburg; Michael Bernhard; Georg F Hoffmann; Johann Motsch
Journal:  Dtsch Arztebl Int       Date:  2009-11-06       Impact factor: 5.594

9.  Comparison of Two High-Dose Magnesium Infusion Regimens in the Treatment of Status Asthmaticus.

Authors:  Danish Vaiyani; Jose E Irazuzta
Journal:  J Pediatr Pharmacol Ther       Date:  2016 May-Jun

10.  Guidelines for paediatric life support. Paediatric Life Support Working Party of the European Resuscitation Council.

Authors: 
Journal:  BMJ       Date:  1994-05-21
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.