Optimum tension for partial ossicular replacement prosthesis reconstruction in the human middle ear.

David P Morris1, Manohar Bance, Rene G van Wijhe, Michael Kiefte, Rachael Smith.   

Abstract

OBJECTIVE: Hearing results from ossiculoplasty are unpredictable. There are many potentially modifiable parameters. One parameter that has not been adequately investigated in the past is the effect of tension on the mechanical functioning of the prosthesis. Our goal was to investigate this parameter further, with the hypothesis that the mechanical functioning of partial ossicular replacement prostheses (PORP) from the stapes head to the eardrum will be affected by the tension that they are placed under.
METHODS: Fresh temporal bones were used to reconstruct a missing incus defect with a PORP-type prosthesis. Three different lengths of PORP were used, and the stapes vibrations were measured with a laser Doppler vibrometer using a calibrated standard sound in the ear canal. Eight temporal bones were used.
RESULTS: Tension had a very significant effect on stapes vibration. In general, loose prostheses resulted in the best overall vibration transmission. The effects were most marked at the lower frequencies. There was a slight advantage to tight prostheses in the higher frequencies, but much less than the decrement in lower frequencies with tight prostheses.
CONCLUSION: In ossicular reconstruction, best stapes vibration results in our model are achieved by shorter prostheses, which result in lower tension.

Entities:  

Mesh:

Year:  2004        PMID: 14755209     DOI: 10.1097/00005537-200402000-00024

Source DB:  PubMed          Journal:  Laryngoscope        ISSN: 0023-852X            Impact factor:   3.325


  11 in total

1.  Reconstructive methods in hearing disorders - surgical methods.

Authors:  Thomas Zahnert
Journal:  GMS Curr Top Otorhinolaryngol Head Neck Surg       Date:  2005-09-28

2.  Parameters for novel incus replacement prostheses.

Authors:  Holger Kaftan; Andrea Böhme; Heiner Martin
Journal:  Eur Arch Otorhinolaryngol       Date:  2015-11-04       Impact factor: 2.503

3.  Clinical investigation of flat panel CT following middle ear reconstruction: a study of 107 patients.

Authors:  K Zaoui; J Kromeier; M Neudert; T Beleites; T Zahnert; R Laszig; C Offergeld
Journal:  Eur Radiol       Date:  2013-11-26       Impact factor: 5.315

Review 4.  [Reconstruction of the middle ear with passive implants].

Authors:  T Zahnert
Journal:  HNO       Date:  2011-10       Impact factor: 1.284

5.  Comparisons of the mechanics of partial and total ossicular replacement prostheses with cartilage in a cadaveric temporal bone preparation.

Authors:  Cagatay Han Ulku; Jeffrey Tao Cheng; Jeremie Guignard; John J Rosowski
Journal:  Acta Otolaryngol       Date:  2014-05-22       Impact factor: 1.494

6.  Experimental investigation of rotational tomography in reconstructed middle ears with clinical implications.

Authors:  Christian Offergeld; Jan Kromeier; Saumil N Merchant; Nicoloz Lasurashvili; Marcus Neudert; Matthias Bornitz; Roland Laszig; Thomas Zahnert
Journal:  Hear Res       Date:  2009-12-05       Impact factor: 3.208

7.  First results of a novel adjustable-length ossicular reconstruction prosthesis in temporal bones.

Authors:  Peter K Gottlieb; Xiping Li; Ashkan Monfared; Nikolas Blevins; Sunil Puria
Journal:  Laryngoscope       Date:  2016-03-12       Impact factor: 3.325

8.  [Variable length titanium prostheses for type III tympanoplasty. Intraoperative length adjustment and fixation of the cartilage overlay].

Authors:  H-P Zenner; R Zimmermann; U Steinhardt; M M Maassen
Journal:  HNO       Date:  2006-04       Impact factor: 1.284

Review 9.  Tympanoplasty - news and new perspectives.

Authors:  Marcus Neudert; Thomas Zahnert
Journal:  GMS Curr Top Otorhinolaryngol Head Neck Surg       Date:  2017-12-18

10.  Handheld laser-fiber vibrometry probe for assessing auditory ossicles displacement.

Authors:  Marcin Masalski; Adam Wąż; Przemysław Błauciak; Tomasz Zatonski; Krzysztof Morawski
Journal:  J Biomed Opt       Date:  2021-07       Impact factor: 3.170

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.