[Variable length titanium prostheses for type III tympanoplasty. Intraoperative length adjustment and fixation of the cartilage overlay].

H-P Zenner1, R Zimmermann, U Steinhardt, M M Maassen.   

Abstract

INTRODUCTION: For type III tympanoplasty by partial ossicular replacement prosthesis (PORP) or total ossicular replacement prosthesis (TORP), the length of the prosthesis must match the individual intraoperative anatomical and physiological characteristics.
MATERIALS AND METHODS: Databanks were used to determine the necessary sizer length of the sizer disc. The measurement template for the size of the cartilage to overlay the prosthesis headplate was derived from the headplates of the Tübinger titanium prostheses (TTP) and the Dresdener titanium prostheses. Finally all functions were integrated into a synthetic plate.
RESULTS: The result was a simple and reasonably priced disposable multifunctional instrument (Tübinger sizer disc TSD) which allowed an exact measurement for every prosthesis in TORP and PORP. For the TTP-Variac, the TSD enabled the simple intraoperative production of prostheses with the length desired by the surgeon. For PORP the TSD enabled an adaptation of the diameter of the prosthesis foot for TTP, TTP-Vario and TTP-Variac and provided a template for the size determination of the cartilage overlay of the titanium prosthesis head. The sizers and the resulting prostheses were used for initial tympanoplastic operations. Audiometric investigations carried out 6 weeks postoperatively gave results corresponding to those previously obtained in a study with TTP and TTP-Vario using the old instrumentation.
CONCLUSIONS: The new instrumentation leads to an improvement of the intraoperative practicability and a simplification. The audiological results remain the same.

Entities:  

Mesh:

Substances:

Year:  2006        PMID: 16557413     DOI: 10.1007/s00106-006-1391-x

Source DB:  PubMed          Journal:  HNO        ISSN: 0017-6192            Impact factor:   1.284


  19 in total

1.  Optimum tension for partial ossicular replacement prosthesis reconstruction in the human middle ear.

Authors:  David P Morris; Manohar Bance; Rene G van Wijhe; Michael Kiefte; Rachael Smith
Journal:  Laryngoscope       Date:  2004-02       Impact factor: 3.325

Review 2.  [Function tests for the Eustachian tube. Current knowledge].

Authors:  E Martino; R Di Thaden; G A Krombach; M Westhofen
Journal:  HNO       Date:  2004-11       Impact factor: 1.284

3.  Ossiculoplasty using incus interposition: hearing results and analysis of the middle ear risk index.

Authors:  Robert C O'Reilly; Steven P Cass; Barry E Hirsch; Donald B Kamerer; Richard A Bernat; Sherri P Poznanovic
Journal:  Otol Neurotol       Date:  2005-09       Impact factor: 2.311

4.  Malleus-to-footplate versus malleus-to-stapes-head ossicular reconstruction prostheses: temporal bone pressure gain measurements and clinical audiological data.

Authors:  Euan Murugasu; Sunil Puria; Joseph B Roberson
Journal:  Otol Neurotol       Date:  2005-07       Impact factor: 2.311

5.  Early results with titanium ossicular implants.

Authors:  Steven Y Ho; Robert A Battista; Richard J Wiet
Journal:  Otol Neurotol       Date:  2003-03       Impact factor: 2.311

6.  [Morphological and functional results of Palisade Cartilage Tympanoplasty].

Authors:  A Neumann; A Hennig; H J Schultz-Coulon
Journal:  HNO       Date:  2002-10       Impact factor: 1.284

7.  Tympano-ossiculoplasty utilizing the Spiggle and Theis titanium total ossicular replacement prosthesis.

Authors:  Brian A Neff; Franklin M Rizer; Arnold G Schuring; William H Lippy
Journal:  Laryngoscope       Date:  2003-09       Impact factor: 3.325

8.  Titanium clip prosthesis.

Authors:  Karl-Bernd Hüttenbrink; Th Zahnert; E G Wüstenberg; G Hofmann
Journal:  Otol Neurotol       Date:  2004-07       Impact factor: 2.311

9.  [Experimental investigations of CO2 laser application in middle ear ossicles].

Authors:  S Dazert; D Russ; R Mlynski; D Brors; A Greiner; C Aletsee; J Helms
Journal:  HNO       Date:  2003-04-09       Impact factor: 1.284

10.  A new L-shaped titanium prosthesis for total reconstruction of the ossicular chain.

Authors:  Ugo Fisch; John May; Thomas Linder; Ilka Charlotte Naumann
Journal:  Otol Neurotol       Date:  2004-11       Impact factor: 2.311

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.