Literature DB >> 14666034

Magnetic vascular coupling for distal anastomosis in coronary artery bypass grafting: a multicenter trial.

Uwe Klima1, Volkmar Falk, Michael Maringka, Steffen Bargenda, Steven Badack, Anton Moritz, Friedrich Mohr, Axel Haverich, Gerhard Wimmer-Greinecker.   

Abstract

OBJECTIVE: The hand-sewn anastomosis is the "gold standard" for performing coronary artery bypass grafts. However, performing a hand-sewn anastomosis is more demanding and time-consuming when used in less invasive approaches such as small access, totally endoscopic or beating heart surgery. In conjunction with attempts to reduce the surgical trauma of coronary artery bypass grafts by using these less invasive approaches, alternative methods for constructing distal anastomoses should be explored. These data report on predischarge angiographic findings and 30-day clinical follow up of patients who have received a new distal anastomotic device.
METHODS: In a multicenter trial, 32 patients (mean age: 65 +/- 9 years; 85% men) requiring multivessel coronary artery bypass surgery had 1 of the anastomoses performed using a novel anastomotic technology. The Magnetic Vascular Positioner System was used in 1 of the bypass grafts and the other bypasses were completed by conventional hand-sewn technique. The Magnetic Vascular Positioner System consists of 4 magnetic, gold-plated implants and 2 delivery devices that facilitate the creation of a functional end-to-side anastomosis. A predischarge angiogram was performed to evaluate graft patency.
RESULTS: There were no device-related major adverse events. The application of the Magnetic Vascular Positioner device was successful in 32 of 41 cases (78%). Nine patients were intended for treatment but did not receive the Magnetic Vascular Positioner System. In 5 of the cases the coronary artery was too small; 1 case had a posterior wall plaque in the target artery; and 3 patients had a nonhemostatic anastomosis after coupling of the port and were subsequently converted to hand-sewn anastomoses. The median total Magnetic Vascular Positioner anastomotic time was 137 seconds with a range from 65 to 370 seconds. Overall patency rate of the Magnetic Vascular Positioner anastomosis was 93.5% versus 91.7% (P = not significant) in hand-sewn grafts. One patient (3.1%) died due to low cardiac output but had patent grafts at autopsy. One myocardial infarction (3.1%) occurred the day after a percutaneous transluminal coronary angioplasty of a hand-sewn graft. One prolonged mechanical ventilation (3.1%) was required because of pneumonia and adult respiratory distress syndrome.
CONCLUSIONS: Magnetic vascular coupling in coronary surgery is safe and effective and has acceptable early patency rates. This new technique may facilitate beating heart and minimally invasive coronary artery bypass grafts.

Entities:  

Mesh:

Year:  2003        PMID: 14666034     DOI: 10.1016/s0022-5223(03)01314-x

Source DB:  PubMed          Journal:  J Thorac Cardiovasc Surg        ISSN: 0022-5223            Impact factor:   5.209


  9 in total

Review 1.  Will drug-eluting stents replace coronary artery bypass surgery?

Authors:  Ross M Reul
Journal:  Tex Heart Inst J       Date:  2005

2.  Development of a suturing device for anastomosis for small caliber arteries.

Authors:  Yoshifumi Itoda; Nirmal Panthee; Tsuruhito Tanaka; Takehiro Ando; Ichiro Sakuma; Minoru Ono
Journal:  J Artif Organs       Date:  2013-12-24       Impact factor: 1.731

3.  Choledochojejunostomy with an innovative magnetic compressive anastomosis: How to determine optimal pressure?

Authors:  Fei Xue; Hong-Chang Guo; Jian-Peng Li; Jian-Wen Lu; Hao-Hua Wang; Feng Ma; Ya-Xiong Liu; Yi Lv
Journal:  World J Gastroenterol       Date:  2016-02-21       Impact factor: 5.742

4.  Extrahepatic portacaval shunt via a magnetic compression technique: A cadaveric feasibility study.

Authors:  Xiao-Peng Yan; Wen-Yan Liu; Jia Ma; Jian-Peng Li; Yi Lv
Journal:  World J Gastroenterol       Date:  2015-07-14       Impact factor: 5.742

5.  Advanced Roux-en-Y hepaticojejunostomy with magnetic compressive anastomats in obstructive jaundice dog models.

Authors:  Chao Fan; Hongke Zhang; Xiaopeng Yan; Jia Ma; Chunbao Wang; Yi Lv
Journal:  Surg Endosc       Date:  2017-08-04       Impact factor: 4.584

6.  Novel magnetic compression technique for establishment of a canine model of tracheoesophageal fistula.

Authors:  Yi Gao; Rong-Qian Wu; Yi Lv; Xiao-Peng Yan
Journal:  World J Gastroenterol       Date:  2019-08-14       Impact factor: 5.742

Review 7.  Sutureless versus Hand-Sewn Coronary Anastomoses: A Systematic Review and Meta-Analysis.

Authors:  Marieke Hoogewerf; Jeroen Schuurkamp; Johannes C Kelder; Stephan Jacobs; Pieter A Doevendans
Journal:  J Clin Med       Date:  2022-01-29       Impact factor: 4.241

Review 8.  Minimally Invasive Coronary Revascularisation Surgery: A Focused Review of the Available Literature.

Authors:  Karel M Van Praet; Markus Kofler; Timo Z Nazari Shafti; Alaa Abd El Al; Antonia van Kampen; Andrea Amabile; Gianluca Torregrossa; Jörg Kempfert; Volkmar Falk; Husam H Balkhy; Stephan Jacobs
Journal:  Interv Cardiol       Date:  2021-05-19

9.  Gastrojejunal anastomosis in rats using the magnetic compression technique.

Authors:  Yingfeng An; Yanchao Zhang; Hao Liu; Sijie Ma; Shan Fu; Yi Lv; Xiaopeng Yan
Journal:  Sci Rep       Date:  2018-08-02       Impact factor: 4.379

  9 in total