S Badiou1, A M Dupuy, B Descomps, J P Cristolead.
Abstract
To evaluate the relationship between the enzymatic Vitros assay for creatinine determination and other methods, we determined creatinine concentration in 400 heparin samples. Plasma creatinine level was successively determined on the Vitros 750 analyzer (Johnson & Johnson Co., Rochester, NY) and on the Olympus AU2700 analyzer (Olympus France, Rungis, France), using three reagents in the same assay: Olympus-Jaffé and two enzymatic commercial kits-Crea Plus (Roche Diagnostics, Meylan, France) and Enzymatic Creatinine (Randox, Mauguio, France). Comparison of Jaffé and enzymatic measurements of plasma creatinine levels revealed a high correlation when considering all values ranged from 20-1000 micromol/l (r > 0.99). However, for values <60 micromol/l, enzymatic reagents provided best results. Enzymatic methodology is a better clinical choice for the accurate measurement of creatinine, especially for neonates, pediatrics, and hematology units. Because analytical performance and the costs of Randox creatinine were satisfactory for our laboratory, this method, adapted on the Olympus analyzer, was chosen for routine determination of creatinine levels. According to the Valtec protocol (8), no interferences such as hemolysis, lipemia, or bilirubin were detected for Enzymatic Creatinine Randox. Copyright 2003 Wiley-Liss, Inc.Entities: