[Left ventricular functional parameters using ECG-gated multidetector spiral CT in comparison with invasive ventriculography].

M Heuschmid1, A Küttner, S Schröder, B Trebar, C Burgstahler, A Mahnken, M Niethammer, T Trabold, A F Kopp, C D Claussen.   

Abstract

PURPOSE: To evaluate the accuracy of left ventricular function using a multidetector CT (MDCT) with retrospective ECG-gating and to compare the results with conventional ventriculography.
MATERIALS AND METHODS: In 26 patients, retrospectively ECG-gated MDCT of the heart (Volume Zoom, Siemens, Germany) as well as conventional coronary angiography including ventriculography was performed to exclude or follow coronary artery disease. CT examination (120 KV, 400 mAs) was carried out with 4 x 1 mm collimation (500 ms gantry rotation time). For CT angiography, 150 ml of contrast media was injected intravenously at a flow rate of 4 ml/s. All data sets of the functional cardiac parameters were reconstructed in end-systolic and end-diastolic phase. End-systolic volume (ESV), end-diastolic volume (EDV) and ejection fraction (EF) were determined from multiplanar reformations orthogonally through the cardiac short axis and analyzed using special evaluation software (ARGUS, Siemens). The results were compared with ESV, EDV and EF obtained from invasive ventriculography.
RESULTS: In all cases, a sufficient quality of the MDCT images was achieved. EDV (150.1 +/- 16.2 ml MDCT vs. 138.7 +/- 16.9 ml ventriculography; mean difference 11.4 +/- 12.7 ml; r = 0.51) had an acceptable correlation to conventional ventriculography, and ESV (58.1 +/- 14.6 ml vs. 50.2 +/- 13.4 ml; mean difference 7.9 +/- 8.8 ml; r = 0.81) and EF (60.9 +/- 13.6 % vs. 64.9 +/- 12.7 %; mean difference 4.0 +/- 6.2 %; r = 0.79) showed a good correlation. In comparison with invasive ventriculography, MDCT tended to overestimate significantly EDV (p = 0.008) and ESV (p = 0.003) and to underestimate EF (p = 0.001).
CONCLUSION: MDCT of the heart with retrospective ECG-gating enables efficient estimation of left ventricular function, providing important additional information of non-invasive cardiac imaging using MDCT. However, EDV and EVS were significantly overestimated and EF was underestimated in MDCT compared to ventriculography.

Entities:  

Mesh:

Year:  2003        PMID: 14556103     DOI: 10.1055/s-2003-42892

Source DB:  PubMed          Journal:  Rofo        ISSN: 1438-9010


  6 in total

1.  Multi-slice spiral computed tomography: useful for non-invasive coronary angiography?

Authors:  Stephen Schroeder; Jeroen J Bax
Journal:  Int J Cardiovasc Imaging       Date:  2004-02       Impact factor: 2.357

2.  Assessment of left ventricular myocardial function using 16-slice multidetector-row computed tomography: comparison with magnetic resonance imaging and echocardiography.

Authors:  Martin Heuschmid; Julia K Rothfuss; Stephen Schroeder; Michael Fenchel; Norbert Stauder; Christof Burgstahler; Andreas Franow; Ronald S Kuzo; Axel Kuettner; Stephan Miller; Claus D Claussen; Andreas F Kopp
Journal:  Eur Radiol       Date:  2005-10-08       Impact factor: 5.315

3.  Left ventricular function studied with MDCT.

Authors:  Kai Uwe Juergens; Roman Fischbach
Journal:  Eur Radiol       Date:  2005-08-20       Impact factor: 5.315

4.  Measurement of cardiac ventricular volumes using multidetector row computed tomography: comparison of two- and three-dimensional methods.

Authors:  M Montaudon; E Laffon; P Berger; O Corneloup; V Latrabe; F Laurent
Journal:  Eur Radiol       Date:  2006-04-12       Impact factor: 5.315

5.  Assessment of right ventricular function by 16-detector-row CT: comparison with magnetic resonance imaging.

Authors:  K Koch; F Oellig; K Oberholzer; P Bender; P Kunz; P Mildenberger; U Hake; K-F Kreitner; M Thelen
Journal:  Eur Radiol       Date:  2004-11-24       Impact factor: 5.315

6.  Multi-detector row computed tomography of the heart: does a multi-segment reconstruction algorithm improve left ventricular volume measurements?

Authors:  Kai Uwe Juergens; David Maintz; Matthias Grude; Jan M Boese; Britta Heimes; Eva Maria Fallenberg; Walter Heindel; Roman Fischbach
Journal:  Eur Radiol       Date:  2004-11-24       Impact factor: 5.315

  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.