Literature DB >> 14499778

Wide angle colonoscopy with a prototype instrument: impact on miss rates and efficiency as determined by back-to-back colonoscopies.

Douglas K Rex1, Vidyasree Chadalawada, Debra J Helper.   

Abstract

OBJECTIVE: Polyps are missed during conventional colonoscopy, even with meticulous technique. The aim of this study was to investigate whether a prototype wide angle colonoscope is associated with a reduced miss rate for polyps.
METHODS: Two studies were performed. In study 1, a total of 50 patients underwent back-to-back, same-day colonoscopy by a single examiner with the prototype wide angle colonoscope and with a standard colonoscope, with the order of scopes randomized. In study 1, an attempt was made to keep examination time with the two colonoscopes equal. In study 2, a total of 20 patients were examined, 10 by the same colonoscopist who performed study 1 and 10 by a second colonoscopist. In study 2, examiners tried to perform the examinations as quickly as accuracy would allow.
RESULTS: In study 1, the miss rate for all polyps was lower with the wide angle colonoscope (20% vs 31%; p = 0.046), although the mean examination time with the wide angle instrument was shorter (6.75 min vs 7.64 min; p = 0.0005). There was no significant difference in detection of adenomas. Polyps, including adenomas, were missed in the peripheral endoscopic field more frequently with the standard colonoscope. In study 2, wide angle colonoscopy was associated with reductions in examination time of 25% and 30% for the two examiners, respectively. Miss rates were the same for one colonoscopist but were higher for the other colonoscopist when the wide angle instrument was used.
CONCLUSION: A prototype wide angle colonoscope did not eliminate polyp miss rates. Wide angle colonoscopy has the potential to reduce examination time and improve visualization of the periphery of the endoscopic field of view, but improvements in resolution are needed.

Entities:  

Mesh:

Year:  2003        PMID: 14499778     DOI: 10.1111/j.1572-0241.2003.07662.x

Source DB:  PubMed          Journal:  Am J Gastroenterol        ISSN: 0002-9270            Impact factor:   10.864


  23 in total

1.  Prospective randomized controlled trial evaluating cap-assisted colonoscopy vs standard colonoscopy.

Authors:  Hoi-Poh Tee; Crispin Corte; Hamdan Al-Ghamdi; Emilia Prakoso; John Darke; Raman Chettiar; Wassim Rahman; Scott Davison; Sean-P Griffin; Warwick-S Selby; Arthur-J Kaffes
Journal:  World J Gastroenterol       Date:  2010-08-21       Impact factor: 5.742

2.  A new wide-angle arthroscopic system: a comparative study with a conventional 30° arthroscopic system.

Authors:  Kyunghwa Jung; Dong-Ju Kang; Aashay L Kekatpure; Arnold Adikrishna; Jaesung Hong; In-Ho Jeon
Journal:  Knee Surg Sports Traumatol Arthrosc       Date:  2016-01-08       Impact factor: 4.342

3.  Linear-array EUS improves detection of pancreatic lesions in high-risk individuals: a randomized tandem study.

Authors:  Eun Ji Shin; Mark Topazian; Michael G Goggins; Sapna Syngal; John R Saltzman; Jeffrey H Lee; James J Farrell; Marcia I Canto
Journal:  Gastrointest Endosc       Date:  2015-04-27       Impact factor: 9.427

4.  Current status of chromoendoscopy and narrow band imaging in colonoscopy.

Authors:  Jonathan P Nass; Sean E Connolly
Journal:  Clin Colon Rectal Surg       Date:  2010-02

5.  Polyp detection rates using magnification with narrow band imaging and white light.

Authors:  Nooman Gilani; Sally Stipho; James D Panetta; Sorin Petre; Michele A Young; Francisco C Ramirez
Journal:  World J Gastrointest Endosc       Date:  2015-05-16

6.  Short turn radius colonoscope in an anatomical model: retroflexed withdrawal and detection of hidden polyps.

Authors:  Sarah K McGill; Shivangi Kothari; Shai Friedland; Ann Chen; Walter G Park; Subhas Banerjee
Journal:  World J Gastroenterol       Date:  2015-01-14       Impact factor: 5.742

7.  The impact of narrow band imaging for colon polyp detection: a multicenter randomized controlled trial by tandem colonoscopy.

Authors:  Hiroaki Ikematsu; Yutaka Saito; Shinji Tanaka; Toshio Uraoka; Yasushi Sano; Takahiro Horimatsu; Takahisa Matsuda; Shiro Oka; Reiji Higashi; Hideki Ishikawa; Kazuhiro Kaneko
Journal:  J Gastroenterol       Date:  2012-03-24       Impact factor: 7.527

8.  Colonoscopy surveillance after polypectomy.

Authors:  Kunio Kasugai; Naotaka Ogasawara; Makoto Sasaki
Journal:  Clin J Gastroenterol       Date:  2011-10-16

9.  High-definition colonoscopy with i-Scan: better diagnosis for small polyps and flat adenomas.

Authors:  Pier Alberto Testoni; Chiara Notaristefano; Cristian Vailati; Milena Di Leo; Edi Viale
Journal:  World J Gastroenterol       Date:  2012-10-07       Impact factor: 5.742

10.  Gastrointestinal endoscopy nurse experience and polyp detection during screening colonoscopy.

Authors:  Evan S Dellon; Quinn K Lippmann; Robert S Sandler; Nicholas J Shaheen
Journal:  Clin Gastroenterol Hepatol       Date:  2008-09-17       Impact factor: 11.382

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.