Polyethylene wear in total hip prostheses: the influence of direction of linear wear on volumetric wear determined from radiographic data.

R Kosak1, V Antolic, V Pavlovcic, V Kralj-Iglic, I Milosev, G Vidmar, A Iglic.   

Abstract

PURPOSE: To develop a new mathematical model for calculating the volumetric wear of polyethylene cups from known values of the radius of the prosthesis head, the extent of linear wear and the direction of linear wear determined from standard antero-posterior radiographs.
METHOD: A new mathematical model was developed. The results of this new mathematical model were compared with the results obtained using the standard, frequently used mathematical model, which takes into consideration only the radius of the prosthesis head and the extent of linear wear of the polyethylene cups. The results of both mathematical models were further compared with the results obtained by direct measurement of volumetric wear using the fluid displacement method.
RESULTS: Comparison of the mathematical models shows that the average volumetric wear calculated using the new mathematical model is 8.5% smaller than the average volumetric wear determined by the fluid displacement method, while the average volumetric wear calculated by standard mathematical model is 17.5% higher. The results of the new mathematical model are, thus, notably less biased than those of the standard one.
CONCLUSION: In calculating the volumetric wear from antero-posterior radiographs, not only the radius of the prosthesis head and the extent of the linear wear but also the direction of the latter has to be considered.

Entities:  

Mesh:

Substances:

Year:  2003        PMID: 13680199     DOI: 10.1007/s00256-003-0685-2

Source DB:  PubMed          Journal:  Skeletal Radiol        ISSN: 0364-2348            Impact factor:   2.199


  24 in total

1.  Tissue response in relation to type of wear particles around failed hip arthroplasties.

Authors:  D W Howie
Journal:  J Arthroplasty       Date:  1990-12       Impact factor: 4.757

2.  Radiological evaluation of the interfaces after cemented total hip replacement. Interobserver and intraobserver agreement.

Authors:  A W McCaskie; A R Brown; J R Thompson; P J Gregg
Journal:  J Bone Joint Surg Br       Date:  1996-03

3.  Determination of polyethylene wear in total hip replacements with use of digital radiographs.

Authors:  J M Martell; S Berdia
Journal:  J Bone Joint Surg Am       Date:  1997-11       Impact factor: 5.284

4.  Statistical methods for assessing agreement between two methods of clinical measurement.

Authors:  J M Bland; D G Altman
Journal:  Lancet       Date:  1986-02-08       Impact factor: 79.321

5.  The Frank Stinchfield Award. Contribution of cable debris generation to accelerated polyethylene wear.

Authors:  J D Hop; J J Callaghan; J P Olejniczak; D R Pedersen; T D Brown; R C Johnston
Journal:  Clin Orthop Relat Res       Date:  1997-11       Impact factor: 4.176

6.  Cementless total joint arthroplasty prostheses with titanium-alloy articular surfaces. A human retrieval analysis.

Authors:  S Nasser; P A Campbell; D Kilgus; N Kossovsky; H C Amstutz
Journal:  Clin Orthop Relat Res       Date:  1990-12       Impact factor: 4.176

7.  Periprosthetic bone loss in total hip arthroplasty. Polyethylene wear debris and the concept of the effective joint space.

Authors:  T P Schmalzried; M Jasty; W H Harris
Journal:  J Bone Joint Surg Am       Date:  1992-07       Impact factor: 5.284

8.  Wear of polyethylene cups in total hip arthroplasty. A study of specimens retrieved post mortem.

Authors:  C J Sychterz; K H Moon; Y Hashimoto; K M Terefenko; C A Engh; T W Bauer
Journal:  J Bone Joint Surg Am       Date:  1996-08       Impact factor: 5.284

9.  15-21-year results of the Charnley low-friction arthroplasty.

Authors:  B M Wroblewski
Journal:  Clin Orthop Relat Res       Date:  1986-10       Impact factor: 4.176

10.  In vivo wear of polyethylene acetabular components.

Authors:  J M Kabo; J S Gebhard; G Loren; H C Amstutz
Journal:  J Bone Joint Surg Br       Date:  1993-03
View more
  8 in total

1.  Wear analysis of chamfered elongated acetabular cup liners.

Authors:  Hsiao-Che Lin; Tzuo-Liang Luo; Jian-Horng Chen
Journal:  Med Biol Eng Comput       Date:  2011-12-22       Impact factor: 2.602

2.  Wear patterns of, and wear volume formulae for, cylindrically elongated acetabular cup liners.

Authors:  James Shih-Shyn Wu; Shu-Ling Hsu; Jian-Horng Chen
Journal:  Med Biol Eng Comput       Date:  2010-04-28       Impact factor: 2.602

3.  Evaluating the accuracy of wear formulae for acetabular cup liners.

Authors:  James Shih-Shyn Wu; Shu-Ling Hsu; Jian-Horng Chen
Journal:  Med Biol Eng Comput       Date:  2009-10-10       Impact factor: 2.602

4.  Polyethylene wear is related to patient-specific contact stress in THA.

Authors:  Robert Košak; Veronika Kralj-Iglič; Aleš Iglič; Matej Daniel
Journal:  Clin Orthop Relat Res       Date:  2011-09-24       Impact factor: 4.176

5.  In vitro evaluation of stiffness graded artificial hip joint femur head in terms of joint stresses distributions and dimensions: finite element study.

Authors:  H Fouad
Journal:  J Mater Sci Mater Med       Date:  2011-04-20       Impact factor: 3.896

6.  THR Simulator--the software for generating radiographs of THR prosthesis.

Authors:  Tai-Yin Wu; Rong-Sen Yang; Chiou-Shann Fuh; Sheng-Mou Hou; Chen-Kun Liaw
Journal:  BMC Musculoskelet Disord       Date:  2009-01-16       Impact factor: 2.362

7.  A simple method of measuring the wear of explanted acetabular component inserts.

Authors:  L Krakow; A Klockow; E Roehner; S Brodt; H Eijer; J Bossert; G Matziolis
Journal:  Bone Joint Res       Date:  2017-09       Impact factor: 5.853

8.  Hip stress distribution - Predictor of dislocation in hip arthroplasties. A retrospective study of 149 arthroplasties.

Authors:  Matevž Tomaževič; Tina Kaiba; Urban Kurent; Rihard Trebše; Matej Cimerman; Veronika Kralj-Iglič
Journal:  PLoS One       Date:  2019-11-20       Impact factor: 3.240

  8 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.