Initial evaluation of commercial optical CT-based 3D gel dosimeter.

K T S Islam1, James F Dempsey, Manisha K Ranade, Marek J Maryanski, Daniel A Low.   

Abstract

We evaluated the OCTOPUS-ONE research laser CT scanner developed and manufactured by MGS Research, Inc. (Madison, CT). The scanner is designed for imaging 3D optical density distributions in BANG gels. The scanner operates in a translate-rotate configuration with a single scanning laser beam. The rotating cylindrical gel phantom is immersed in a refractive index matching solution and positioned at the center of a square tank made of plastic and glass. A stationary polarized He-Ne laser beam (633 nm) is reflected from a mirror moving parallel to the tank wall and scans the gel. Another mirror moves synchronously along the opposite side of the tank and collects the transmitted light and sends it to a single stationary silicon photodetector. A filtered backprojection algorithm is used to reconstruct projection data in a plane. The laser-mirrors-detector assembly is mounted on a horizontal platform that moves vertically for slice selection. We have tested the mechanical and optical setup, projection centering on the axis of rotation, linearity, and spatial resolution. We found the optical detector to respond linearly to transmitted light from control samples. The spatial resolution of the scanner was determined by employing a split field resolution technique. We obtained the horizontal and vertical full widths at half maxima of the laser beam intensity profiles as 0.6 and 0.8 mm, respectively. Dose calibration tests of the gel were performed using a nine-field (2 x 2 cm2 each) dose pattern irradiated at different dose levels. Finally, we compared gel-derived 2D planar dose distribution against radiochromic film measured dose distribution for both the nine-field and a uniform 5 x 5 cm2 field of 6 MV x rays. Very similar dose distributions were observed in gel and radiochromic film except in regions of steep dose gradient and highest dose. A dose normalization of 15.6% was required between the two dosimeters due to differences in overall radiation response. After normalization, analysis using the gamma evaluation showed that the radiochromic film and gel-measured dose distributions differed by a maximum gamma of 1.3 using 5% and 1.5 mm dose difference and distance-to-agreement criteria. The optical CT scanner has great potential as a 3D dosimeter, but a few refinements and further testing are necessary before its routine clinical use.

Entities:  

Mesh:

Substances:

Year:  2003        PMID: 12945982     DOI: 10.1118/1.1593636

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  15 in total

1.  Optical-CT gel-dosimetry. II: Optical artifacts and geometrical distortion.

Authors:  Mark Oldham; Leonard Kim
Journal:  Med Phys       Date:  2004-05       Impact factor: 4.071

2.  Optical-CT scanning of polymer gels.

Authors:  M Oldham
Journal:  J Phys Conf Ser       Date:  2004

3.  Fast, high-resolution 3D dosimetry utilizing a novel optical-CT scanner incorporating tertiary telecentric collimation.

Authors:  H S Sakhalkar; M Oldham
Journal:  Med Phys       Date:  2008-01       Impact factor: 4.071

4.  Quality assurance in 3D dosimetry by optical-CT.

Authors:  Pengyi Guo; John Adamovics; Mark Oldham
Journal:  J Phys Conf Ser       Date:  2006-12-01

5.  A dual-purpose CCD based micro-optical-CT scanning system.

Authors:  M Oldham; H Sakhalkar; P Guo
Journal:  J Phys Conf Ser       Date:  2006-12-01

6.  3D dosimetry by optical-CT scanning.

Authors:  Mark Oldham
Journal:  J Phys Conf Ser       Date:  2006

7.  Sensitivity calibration procedures in optical-CT scanning of BANG 3 polymer gel dosimeters.

Authors:  Y Xu; Cheng-Shie Wuu; Marek J Maryanski
Journal:  Med Phys       Date:  2010-02       Impact factor: 4.071

8.  Commissioning and benchmarking a 3D dosimetry system for clinical use.

Authors:  Andrew Thomas; Joseph Newton; John Adamovics; Mark Oldham
Journal:  Med Phys       Date:  2011-08       Impact factor: 4.071

9.  ScanSim: A tool for simulating optical-CT imaging.

Authors:  Mark Oldham
Journal:  J Phys Conf Ser       Date:  2010

Review 10.  Polymer gel dosimetry.

Authors:  C Baldock; Y De Deene; S Doran; G Ibbott; A Jirasek; M Lepage; K B McAuley; M Oldham; L J Schreiner
Journal:  Phys Med Biol       Date:  2010-02-11       Impact factor: 3.609

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.